Comparison of the new biometer OA-1000 with IOLMaster and Tomey AL-3000.

Susanne Christiane Goebels1, Berthold Seitz, Achim Langenbucher.   

Abstract

BACKGROUND: The OA-1000 (Tomey, Japan) is a new optical biometer, which allows measurements of axial length (AL), anterior chamber depth (ACD) and corneal thickness (CT) due to partial coherence interferometry (PCI) technology. The aim of this study was to compare the OA-1000 results with those obtained with the IOLMaster and contact applanation A-scan ultrasonography.
METHODS: We examined 133 eyes of 75 patients with age related cataract. Mean age was 72.0 ± 9.5 years. Biometry measurements of AL and ACD were performed with the Tomey OA-1000 based on PCI, the IOLMaster based on PCI (AL) and slit projection (ACD), and the Tomey AL-3000 based on contact applanation A-scan ultrasonography.
RESULTS: Mean AL using the IOLMaster was 23.21 ± 1.08 mm, using the AL-3000 was 22.79 ± 1.04 mm, using the OA-1000 it was 22.97 ± 1.1 mm. Mean ACD using the IOLMaster was 2.99 ± 0.41 mm, using the OA-1000 3.40 ± 0.46 mm, using the Tomey AL-3000 it was 2.93 ± 0.43 mm. Mean difference between the AL and ACD measured with the OA-1000 and the IOLMaster was 0.22 ± 0.047 mm and 0.40 ± 0.33 mm, between OA-1000 and the AL-3000 it was 0.19 ± 0.23 mm and 0.47 ± 0.33 mm, and between IOLMaster and AL-3000 it was 0.42 ± 0.23 and 0.09 ± 0.36 mm. For AL the correlation coefficient R between IOLMaster and OA-1000 was 0.999, between IOLMaster and AL-3000 it was 0.976, between AL-3000 and OA-1000 it was 0.977. For ACD R between IOLMaster and OA-1000 was 0.735, between IOLMaster and AL-3000 it was 0.822, between AL-3000 and OA-1000 it was 0.716 (all p < 0.001).
CONCLUSIONS: Compared with other used clinical instruments the OA-1000 generates accurate results. Although differences were found, the OA-1000 provided results that correlated well with the values of IOLMaster and AL-3000.

Entities:  

Mesh:

Year:  2013        PMID: 23841799     DOI: 10.3109/02713683.2013.788722

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  6 in total

1.  Comparison of a new optical biometer and a standard biometer in cataract patients.

Authors:  Pipat Kongsap
Journal:  Eye Vis (Lond)       Date:  2016-10-17

2.  Comparison of optical biometry versus ultrasound biometry in cases with borderline signal-to-noise ratio.

Authors:  Tao Ming Thomas Chia; Minh T Nguyen; Hoon C Jung
Journal:  Clin Ophthalmol       Date:  2018-09-10

3.  OCT Biometry (B-OCT): A New Method for Measuring Ocular Axial Dimensions.

Authors:  Bartosz L Sikorski; Pawel Suchon
Journal:  J Ophthalmol       Date:  2019-08-14       Impact factor: 1.909

4.  Reproducibility of the optical Biometer OA-1000 (Tomey).

Authors:  Susanne Christiane Goebels; Berthold Seitz; Achim Langenbucher
Journal:  Biomed Res Int       Date:  2014-04-10       Impact factor: 3.411

5.  Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes.

Authors:  Yanjun Hua; Wei Qiu; Qiuyi Xiao; Qiang Wu
Journal:  PLoS One       Date:  2018-02-27       Impact factor: 3.240

6.  Development of Corneal Astigmatism (CA) according to Axial Length/Corneal Radius (AL/CR) Ratio in a One-Year Follow-Up of Children in Beijing, China.

Authors:  Fenglei Wang; Lin Xiao; Xuxia Meng; Ling Wang; Dabo Wang
Journal:  J Ophthalmol       Date:  2018-08-30       Impact factor: 1.909

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.