Literature DB >> 12648044

Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation.

Loreto T Rose1, Con N Moshegov.   

Abstract

PURPOSE: A comparison of axial length estimates using applanation A-scan ultrasound and the Zeiss IOLMaster was conducted. The accuracy in predicting postoperative refraction determined by each method was also compared.
METHODS: A cross-sectional study was performed on 51 eyes in 46 patients presenting to clinical practice for cataract surgery assessment. Preoperative measurement of axial length was performed with applanation ultrasound and the IOLMaster. The IOLMaster measurements were used to determine the intraocular lens power based on the SRK/T formula. Postoperative refractive outcomes were obtained and spherical equivalent calculated and compared to the predicted refractive error with each biometric method.
RESULTS: On average the axial lengths measured by the IOLMaster were longer by 0.15 mm compared to ultrasound biometry (P < 0.01). Using the IOLMaster over applanation ultrasound biometry significantly improved the postoperative refractive outcome from 0.65 D to 0.42 D (P = 0.011).
CONCLUSIONS: The IOLMaster provides an accurate axial length measurement and results in accurate intraocular lens power calculation based on the SRK/T formula. It is quick and easy to use and provides a non-contact technique with no risk of infection or corneal abrasion.

Entities:  

Mesh:

Year:  2003        PMID: 12648044     DOI: 10.1046/j.1442-9071.2003.00617.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  24 in total

1.  Misleading axial length measurements with the IOLMaster due to a dense posterior vitreous surface membrane and a macular edema in a diabetic patient.

Authors:  Christoph Fidel Mayer; Navid Ardjomand; Werner Wackernagel; Michaela Velikay-Parel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-11-03       Impact factor: 3.117

2.  Factors affecting refractive outcome after cataract surgery in patients with a history of acute primary angle closure.

Authors:  Hyo Seok Lee; Jung Won Park; Sang Woo Park
Journal:  Jpn J Ophthalmol       Date:  2013-10-29       Impact factor: 2.447

3.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

4.  Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer.

Authors:  Shi-Ming Cheng; Wen-Tao Yan; Jia-Sheng Zhang; Tian-Tian Li; Xi Li; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-04-30       Impact factor: 3.535

5.  Agreement of IOL power and axial length obtained by IOLMaster 500 vs IOLMaster 500 with Sonolink connection.

Authors:  Sabong Srivannaboon; Chareenun Chirapapaisan; Patchara Nantasri; Mathinee Chongchareon; Pratuangsri Chonpimai
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-12       Impact factor: 3.117

6.  Comparison of preoperative and postoperative ocular biometry in eyes with phakic intraocular lens implantations.

Authors:  Joo Youn Shin; Jae Bum Lee; Kyoung Yul Seo; Eung Kweon Kim; Tae-im Kim
Journal:  Yonsei Med J       Date:  2013-09       Impact factor: 2.759

7.  Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths.

Authors:  Jia-Kang Wang; Shu-Wen Chang
Journal:  Int J Ophthalmol       Date:  2013-04-18       Impact factor: 1.779

Review 8.  Recurring themes during cataract assessment and surgery.

Authors:  Shruti Chandra; Sobha Sivaprasad; Paul G Ursell; Khayam Naderi; David O'Brart; Amar Alwitry; Mayank A Nanavaty
Journal:  Eye (Lond)       Date:  2021-04-29       Impact factor: 4.456

9.  The short-term effects of exercise on intraocular pressure, choroidal thickness and axial length.

Authors:  Jie Hong; Hui Zhang; Debbie S Kuo; Huaizhou Wang; Yanjiao Huo; Diya Yang; Ningli Wang
Journal:  PLoS One       Date:  2014-08-29       Impact factor: 3.240

10.  Accuracy of Intraocular Lens Power Calculation Formulas for Highly Myopic Eyes.

Authors:  Yichi Zhang; Xiao Ying Liang; Shu Liu; Jacky W Y Lee; Srinivasan Bhaskar; Dennis S C Lam
Journal:  J Ophthalmol       Date:  2016-03-29       Impact factor: 1.909