Literature DB >> 11702645

Accuracy of intraocular lens power calculations using the Zeiss IOL master. A prospective study.

E Verhulst1, J C Vrijghem.   

Abstract

PURPOSE: Partial Coherence Interferometry (PCI) is a fast, noncontact method to calculate lens implant power for cataract surgery. It has been reported as a potentially more accurate method than ultrasound biometry. Prospective study of the refractive outcomes of a consecutive series of patients undergoing phacoemulsification surgery with preoperative biometry by both ultrasound A-scan and PCI.
METHODS: A series of 50 eyes of 35 patients underwent small-incision phacoemulsification cataract surgery and lens implantation by one single surgeon. All patients had preoperative biometry performed by both ultrasound using the Sonomed and IOL Master optical biometry. The IOL Master results were included in the SRK II formula to calculate the lens implant power. Postoperative refractive assessment was performed 4 weeks after surgery.
RESULTS: The mean difference in axial length between ultrasound and optical biometry was 0.2 mm. The IOL Master measures a longer axial length. The mean keratometric power using the Javal instrument was 43.4 D and for the Zeiss IOL Master it was 42.9 D. At the week 4 postsurgery assessment, the overall refractive outcome was in the range of +/- 1D. Five patients were unable to undergo PCI biometry due to the density of cataract.
CONCLUSIONS: Intraocular lens power calculations using the Zeiss IOL Master are easy to perform and result in excellent refractive outcomes. A-scan biometry is still needed in case of mature cataract.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11702645

Source DB:  PubMed          Journal:  Bull Soc Belge Ophtalmol        ISSN: 0081-0746


  15 in total

1.  Agreement analysis of LENSTAR with other techniques of biometry.

Authors:  S Jasvinder; T F Khang; K K S Sarinder; V P Loo; V Subrayan
Journal:  Eye (Lond)       Date:  2011-03-11       Impact factor: 3.775

2.  [Comparison of IOL-Master 500 vs. Lenstar LS900 concerning the calculation of target refraction: A retrospective analysis].

Authors:  M Stattin; C Zehetner; N E Bechrakis; L Speicher
Journal:  Ophthalmologe       Date:  2015-05       Impact factor: 1.059

3.  Prediction of refractive outcome after cataract surgery using partial coherence interferometry: comparison of SRK/T and Haigis formulae.

Authors:  Rohit Sharma; P Maharajan; S Kotta; S Maharajan
Journal:  Int Ophthalmol       Date:  2014-03-30       Impact factor: 2.031

4.  Usefulness of anterior chamber depth measurement for efficacy assessment of steroid pulse therapy in patients with Vogt-Koyanagi-Harada disease.

Authors:  Tomohiro Otsuki; Kimiya Shimizu; Akihito Igarashi; Kazutaka Kamiya
Journal:  Jpn J Ophthalmol       Date:  2010-11-05       Impact factor: 2.447

5.  Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.

Authors:  Ahmet Kirgiz; Kurşat Atalay; Havva Kaldirim; Kubra Serefoglu Cabuk; Mehmet Orcun Akdemir; Muhittin Taskapili
Journal:  Int Ophthalmol       Date:  2016-09-01       Impact factor: 2.031

6.  [Comparison of two optical biometric devices for intraocular lens calculation].

Authors:  M-S Uckmann; M Stattin; C Zehetner; S Neururer; L Speicher
Journal:  Ophthalmologe       Date:  2019-03       Impact factor: 1.059

7.  Correlation between contrast sensitivity and higher-order aberration based on pupil diameter after cataract surgery.

Authors:  Takefumi Yamaguchi; Kazuno Negishi; Kazuhiko Ohnuma; Kazuo Tsubota
Journal:  Clin Ophthalmol       Date:  2011-12-07

8.  IOLMaster versus Manual Keratometry after Photorefractive Keratectomy.

Authors:  Hasan Razmju; Leila Rezaei; Kobra Nasrollahi; Hamid Fesharaki; Hossein Attarzadeh; Farhad Janbaz Footami
Journal:  J Ophthalmic Vis Res       Date:  2011-07

9.  Intraocular lens power estimation in combined phacoemulsification and pars plana vitrectomy in eyes with epiretinal membranes: a case-control study.

Authors:  Min Kim; Hyoung Eun Kim; Dong Hyun Lee; Hyoung Jun Koh; Sung Chul Lee; Sung Soo Kim
Journal:  Yonsei Med J       Date:  2015-05       Impact factor: 2.759

10.  The distribution of axial length, anterior chamber depth, lens thickness, and vitreous chamber depth in an adult population of Shahroud, Iran.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Mohammad Miraftab; Mohammad Hassan Emamian; Mohammad Shariati; Tahereh Abdolahinia; Akbar Fotouhi
Journal:  BMC Ophthalmol       Date:  2012-09-18       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.