| Literature DB >> 25525312 |
Parisa Akhlaghi1, Hashem Miri-Hakimabad1, Laleh Rafat-Motavalli1.
Abstract
In computed tomography (CT), some superficial organs which have increased sensitivity to radiation, receive doses that are significant enough to be matter of concern. Therefore, in this study, the effects of using shields on the amount of dose reduction and image quality was investigated for pediatric imaging. Absorbed doses of breasts, eyes, thyroid and testes of a series of pediatric phantoms without and with different thickness of bismuth and lead were calculated by Monte Carlo simulation. Appropriate thicknesses of shields were chosen based on their weights, X-ray spectrum, and the amount of dose reduction. In addition, the effect of lead shield on image quality of a simple phantom was assessed quantitatively using region of interest (ROI) measurements. Considering the maximum reduction in absorbed doses and X-ray spectrum, using a lead shield with a maximum thickness of 0.4 mm would be appropriate for testes and thyroid and two other organs (which are exposed directly) should be protected with thinner shields. Moreover, the image quality assessment showed that lead was associated with significant increases in both noise and CT attenuation values, especially in the anterior of the phantom. Overall, the results suggested that shielding is a useful optimization tool in CT.Entities:
Keywords: Image quality; Monte Carlo simulation; lead and bismuth; superficial organs' doses
Year: 2014 PMID: 25525312 PMCID: PMC4258732 DOI: 10.4103/0971-6203.144490
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1Calibration phantom without (a), and with (b) lead shield and spacer
Figure 2Reduction in absorbed dose by using different thickness of lead shields at tube voltages of 80 kVp for breasts (a), eye lenses (b), thyroid (c) and testes (d), calculated by Monte Carlo simulation
EDs in (mSv/mAs) calculated by tissue-weighted sum of the equivalent doses (mGy/mAs) for chest examinations on five anthropomorphic phantoms at tube voltage of 80 kVp without and with five different thicknesses of bismuth and lead shields (EDs were calculated based on ICRP103 weighting factors)
Figure 3X-ray spectrum following passage through different thickness of lead at tube voltage of 80 kVp
Figure 4Unshielded (a) and shielded (b) CT images of one slice of the calibration phantom (In both images the grayscale ranges from 0–1211)
Figure 5Mean HUs units recorded in each ROI at tube voltage of 80 kVp and tube loading of 192 mAs
Figure 6Mean noise (SD of HU) recorded in each ROI at tube voltage of 80 kVp and tube loading of 192 mAs