The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments.

K D Hopper1, S H King, M E Lobell, T R TenHave, J S Weaver.   

Abstract

PURPOSE: To evaluate the ability of thin overlying bismuth radioprotective shielding to reduce the x-ray dose to radiosensitive superficial organs during diagnostic computed tomography (CT).
MATERIALS AND METHODS: A variety of patient and phantom studies were performed with four thicknesses of bismuth radioprotective latex over the breast. Dose savings were determined with thermoluminescent dosimeters. A prototype and then a final manufactured radioprotective brassiere was constructed and tested for radiation dose savings to the breast during diagnostic chest CT. Preliminary studies were also performed to evaluate shielding of the thyroid, orbit, and testes.
RESULTS: The use of bismuth radioprotective latex saved an average 57% of the radiation dose to the breast from thoracic CT, decreasing the radiation level from an average 2.2 rad (0.022 Gy) to 1.0 rad (0.010 Gy) (P < .001). Preliminary tests of shielding other superficial radiosensitive organs frequently included at diagnostic CT (eyes, thyroid gland, and testes) were performed with the same thickness of overlying bismuth radioprotective latex, with similar results. Radiation to the thyroid gland was reduced by 60% (from 0.0573 to 0.0229 Gy) and radiation to the eye and testes was reduced by 40% (from 0.0256 to 0.0154 Gy) and 51% (from 0.0463 to 0.0229 Gy), respectively.
CONCLUSION: The use of in-plane overlying bismuth radioprotective latex manufactured into form-fitting garments did not affect the diagnostic CT image but reduced the amount of radiation to radiosensitive superficial structures.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9393547     DOI: 10.1148/radiology.205.3.9393547

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  33 in total

1.  Radiation doses in computed tomography. The increasing doses of radiation need to be controlled.

Authors:  M M Rehani; M Berry
Journal:  BMJ       Date:  2000-03-04

2.  Efficacy of breast shielding during CT of the head.

Authors:  Z Brnić; B Vekić; A Hebrang; P Anić
Journal:  Eur Radiol       Date:  2003-06-25       Impact factor: 5.315

3.  Pulmonary embolism: is multislice CT the method of choice? Against.

Authors:  Carl Schuemichen
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-01       Impact factor: 9.236

4.  The use of breast shielding for dose reduction in pediatric CT: arguments against the proposition.

Authors:  Jacob Geleijns; Jia Wang; Cynthia McCollough
Journal:  Pediatr Radiol       Date:  2010-08-21

5.  Bismuth shielding in CT: support for use in children.

Authors:  Sangroh Kim; Donald P Frush; Terry T Yoshizumi
Journal:  Pediatr Radiol       Date:  2010-08-24

6.  Outcomes of negative multidetector computed tomography with pulmonary angiography in pregnant women suspected of pulmonary embolism.

Authors:  Ghada Bourjeily; Hanan Khalil; Christina Raker; Susan Martin; Pauline Auger; Michel Chalhoub; Lucia Larson; Margaret Miller
Journal:  Lung       Date:  2011-10-18       Impact factor: 2.584

7.  Effect of bismuth breast shielding on radiation dose and image quality in coronary CT angiography.

Authors:  Andrew J Einstein; Carl D Elliston; Daniel W Groves; Bin Cheng; Steven D Wolff; Gregory D N Pearson; M Robert Peters; Lynne L Johnson; Sabahat Bokhari; Gary W Johnson; Ketan Bhatia; Theodore Pozniakoff; David J Brenner
Journal:  J Nucl Cardiol       Date:  2011-11-09       Impact factor: 5.952

8.  Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality.

Authors:  J Geleijns; M Salvadó Artells; W J H Veldkamp; M López Tortosa; A Calzado Cantera
Journal:  Eur Radiol       Date:  2006-04-08       Impact factor: 5.315

Review 9.  Radiation safety.

Authors:  Donald P Frush
Journal:  Pediatr Radiol       Date:  2009-06

10.  Dose reduction in multislice CT by means of bismuth shields: results of in vivo measurements and computed evaluation.

Authors:  P Catuzzo; S Aimonetto; G Fanelli; P Marchisio; T Meloni; L Mistretta; M Pasquino; E Richetta; S Tofani
Journal:  Radiol Med       Date:  2009-12-14       Impact factor: 3.469

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.