Bismuth shielding, organ-based tube current modulation, and global reduction of tube current for dose reduction to the eye at head CT.

Jia Wang1, Xinhui Duan, Jodie A Christner, Shuai Leng, Katharine L Grant, Cynthia H McCollough.   

Abstract

PURPOSE: To compare the dose and image quality of three methods for reducing the radiation dose to the eye at head computed tomography (CT): bismuth shielding, organ-based tube current modulation (TCM), and global reduction of the tube current.
MATERIALS AND METHODS: An anthropomorphic head phantom was scanned under six conditions: (a) without any dose reduction techniques (reference scanning); (b) with one bismuth eye shield; (c) with organ-based TCM; (d) with reduced tube current to yield the same dose reduction as one bismuth shield; (e) with two layers of bismuth shields; and (f) with organ-based TCM and one bismuth shield. Dose to the eye, image noise, and CT numbers in the brain region were measured and compared. The effect of increasing distance between the bismuth shield and eye lens was also investigated.
RESULTS: Relative to the reference scan, the dose to the eye was reduced by 26.4% with one bismuth shield, 30.4% with organ-based TCM, and 30.2% with a global reduction in tube current. A combination of organ-based TCM with one bismuth shield reduced the dose by 47.0%. Image noise in the brain region was slightly increased for all dose reduction methods. CT numbers were increased whenever the bismuth shield was used. Increasing the distance between the bismuth shield and the eye lens helped reduce CT number errors, but the increase in noise remained.
CONCLUSION: Organ-based TCM provided superior image quality to that with bismuth shielding while similarly reducing dose to the eye. Simply reducing tube current globally by about 30% provides the same dose reduction to the eye as bismuth shielding; however, CT number accuracy is maintained and dose is reduced to all parts of the head. © RSNA, 2012.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22190658     DOI: 10.1148/radiol.11110470

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  19 in total

1.  Data analysis of low dose multislice helical CT scan in orbital trauma.

Authors:  Ji-Wei Wang; Chong Tang; Bo-Rong Pan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Updated estimates of typical effective doses for common CT examinations in the UK following the 2011 national review.

Authors:  Paul C Shrimpton; Jan T M Jansen; John D Harrison
Journal:  Br J Radiol       Date:  2015-11-06       Impact factor: 3.039

3.  The effects of bismuth breast shields in conjunction with automatic tube current modulation in CT imaging.

Authors:  Sabah Servaes; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2013-05-23

4.  New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study.

Authors:  Jung-Su Kim; Soon-Mu Kwon; Jung-Min Kim; Sang-Wook Yoon
Journal:  Radiol Med       Date:  2017-03-24       Impact factor: 3.469

5.  [Dose optimization in CT examination of children].

Authors:  A Hojreh; H Prosch
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

6.  Impact of organ-specific dose reduction on the image quality of head and neck CT angiography.

Authors:  L Schimmöller; R S Lanzman; P Heusch; S Dietrich; F Miese; J Aissa; T A Heusner; G Antoch; P Kröpil
Journal:  Eur Radiol       Date:  2013-01-09       Impact factor: 5.315

7.  [Computed tomography in multiple trauma patients: technical aspects, work flow, and dose reduction].

Authors:  F A Fellner; J Krieger; N Lechner; D Flöry
Journal:  Radiologe       Date:  2014-09       Impact factor: 0.635

8.  Cerebral Ventricular Dimensions After Decompressive Craniectomy: A Comparison Between Bedside Sonographic Duplex Technique and Cranial Computed Tomography.

Authors:  Habib Bendella; Marc Maegele; Alexander Hartmann; Joachim Spreer; Nadine Rommel; Rolf Lefering; Makoto Nakamura; Alhadi Igressa
Journal:  Neurocrit Care       Date:  2017-06       Impact factor: 3.210

9.  Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions.

Authors:  Satoru Kawauchi; Koichi Chida; Yusuke Hamada; Wataro Tsuruta
Journal:  Radiol Phys Technol       Date:  2021-11-18

10.  Lens Dose Reduction by Patient Posture Modification During Neck CT.

Authors:  Elizabeth G Mosher; John A Butman; Les R Folio; Nadia M Biassou; Choonsik Lee
Journal:  AJR Am J Roentgenol       Date:  2018-03-16       Impact factor: 3.959

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.