Radiation dose reduction to the breast in thoracic CT: comparison of bismuth shielding, organ-based tube current modulation, and use of a globally decreased tube current.

Jia Wang1, Xinhui Duan, Jodie A Christner, Shuai Leng, Lifeng Yu, Cynthia H McCollough.   

Abstract

PURPOSE: The purpose of this work was to evaluate dose performance and image quality in thoracic CT using three techniques to reduce dose to the breast: bismuth shielding, organ-based tube current modulation (TCM) and global tube current reduction.
METHODS: Semi-anthropomorphic thorax phantoms of four different sizes (15, 30, 35, and 40 cm lateral width) were used for dose measurement and image quality assessment. Four scans were performed on each phantom using 100 or 120 kV with a clinical CT scanner: (1) reference scan; (2) scan with bismuth breast shield of an appropriate thickness; (3) scan with organ-based TCM; and (4) scan with a global reduction in tube current chosen to match the dose reduction from bismuth shielding. Dose to the breast was measured with an ion chamber on the surface of the phantom. Image quality was evaluated by measuring the mean and standard deviation of CT numbers within the lung and heart regions.
RESULTS: Compared to the reference scan, dose to the breast region was decreased by about 21% for the 15-cm phantom with a pediatric (2-ply) shield and by about 37% for the 30, 35, and 40-cm phantoms with adult (4-ply) shields. Organ-based TCM decreased the dose by 12% for the 15-cm phantom, and 34-39% for the 30, 35, and 40-cm phantoms. Global lowering of the tube current reduced breast dose by 23% for the 15-cm phantom and 39% for the 30, 35, and 40-cm phantoms. In phantoms of all four sizes, image noise was increased in both the lung and heart regions with bismuth shielding. No significant increase in noise was observed with organ-based TCM. Decreasing tube current globally led to similar noise increases as bismuth shielding. Streak and beam hardening artifacts, and a resulting artifactual increase in CT numbers, were observed for scans with bismuth shields, but not for organ-based TCM or global tube current reduction.
CONCLUSIONS: Organ-based TCM produces dose reduction to the breast similar to that achieved with bismuth shielding for both pediatric and adult phantoms. However, organ-based TCM does not affect image noise or CT number accuracy, both of which are adversely affected by bismuth shielding. Alternatively, globally decreasing the tube current can produce the same dose reduction to the breast as bismuth shielding, with a similar noise increase, yet without the streak artifacts and CT number errors caused by the bismuth shields. Moreover, globally decreasing the tube current reduces the dose to all tissues scanned, not simply to the breast.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22047373     DOI: 10.1118/1.3651489

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  24 in total

1.  The effects of bismuth breast shields in conjunction with automatic tube current modulation in CT imaging.

Authors:  Sabah Servaes; Xiaowei Zhu
Journal:  Pediatr Radiol       Date:  2013-05-23

2.  Limiting CT radiation dose in children with craniosynostosis: phantom study using model-based iterative reconstruction.

Authors:  Touko Kaasalainen; Kirsi Palmu; Anniina Lampinen; Vappu Reijonen; Junnu Leikola; Riku Kivisaari; Mika Kortesniemi
Journal:  Pediatr Radiol       Date:  2015-05-05

3.  Partial arc beam filtration: a novel approach to reducing CT breast radiation dose.

Authors:  Kelsey B Mathieu; Dianna D Cody
Journal:  AJR Am J Roentgenol       Date:  2013-03       Impact factor: 3.959

4.  Breast dose reduction with organ-based, wide-angle tube current modulated CT.

Authors:  Wanyi Fu; Gregory M Sturgeon; Greeshma Agasthya; William Paul Segars; Anuj J Kapadia; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-04

5.  Radiation dose and image quality of CT fluoroscopy with partial exposure mode.

Authors:  Keisuke Takiguchi; Atsushi Urikura; Tsukasa Yoshida; Yoshihiro Nakaya; Masahiro Endo; Takeshi Aramaki
Journal:  Diagn Interv Radiol       Date:  2020-07       Impact factor: 2.630

6.  [Dose optimization in CT examination of children].

Authors:  A Hojreh; H Prosch
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

7.  [Current strategies for dosage reduction in computed tomography].

Authors:  M S May; W Wuest; M M Lell; M Uder; W A Kalender; B Schmidt
Journal:  Radiologe       Date:  2012-10       Impact factor: 0.635

8.  Differences in behavior of tube current modulation techniques for thoracic CT examinations between male and female anthropomorphic phantoms.

Authors:  Kosuke Matsubara; Pei-Jan Paul Lin; Atsushi Fukuda; Kichiro Koshida
Journal:  Radiol Phys Technol       Date:  2014-05-27

9.  [Computed tomography in multiple trauma patients: technical aspects, work flow, and dose reduction].

Authors:  F A Fellner; J Krieger; N Lechner; D Flöry
Journal:  Radiologe       Date:  2014-09       Impact factor: 0.635

Review 10.  Imaging of acute pulmonary embolism.

Authors:  Maria Komissarova; Suzanne Chong; Kirk Frey; Baskaran Sundaram
Journal:  Emerg Radiol       Date:  2012-11-14
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.