Organ doses for reference adult male and female undergoing computed tomography estimated by Monte Carlo simulations.

Choonsik Lee1, Kwang Pyo Kim, Daniel Long, Ryan Fisher, Chris Tien, Steven L Simon, Andre Bouville, Wesley E Bolch.   

Abstract

PURPOSE: To develop a computed tomography (CT) organ dose estimation method designed to readily provide organ doses in a reference adult male and female for different scan ranges to investigate the degree to which existing commercial programs can reasonably match organ doses defined in these more anatomically realistic adult hybrid phantoms
METHODS: The x-ray fan beam in the SOMATOM Sensation 16 multidetector CT scanner was simulated within the Monte Carlo radiation transport code MCNPX2.6. The simulated CT scanner model was validated through comparison with experimentally measured lateral free-in-air dose profiles and computed tomography dose index (CTDI) values. The reference adult male and female hybrid phantoms were coupled with the established CT scanner model following arm removal to simulate clinical head and other body region scans. A set of organ dose matrices were calculated for a series of consecutive axial scans ranging from the top of the head to the bottom of the phantoms with a beam thickness of 10 mm and the tube potentials of 80, 100, and 120 kVp. The organ doses for head, chest, and abdomen/pelvis examinations were calculated based on the organ dose matrices and compared to those obtained from two commercial programs, CT-EXPO and CTDOSIMETRY. Organ dose calculations were repeated for an adult stylized phantom by using the same simulation method used for the adult hybrid phantom.
RESULTS: Comparisons of both lateral free-in-air dose profiles and CTDI values through experimental measurement with the Monte Carlo simulations showed good agreement to within 9%. Organ doses for head, chest, and abdomen/pelvis scans reported in the commercial programs exceeded those from the Monte Carlo calculations in both the hybrid and stylized phantoms in this study, sometimes by orders of magnitude.
CONCLUSIONS: The organ dose estimation method and dose matrices established in this study readily provides organ doses for a reference adult male and female for different CT scan ranges and technical parameters. Organ doses from existing commercial programs do not reasonably match organ doses calculated for the hybrid phantoms due to differences in phantom anatomy, as well as differences in organ dose scaling parameters. The organ dose matrices developed in this study will be extended to cover different technical parameters, CT scanner models, and various age groups.

Mesh:

Year:  2011        PMID: 21520832      PMCID: PMC3055697          DOI: 10.1118/1.3544658

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Application of draft European Commission reference levels to a regional CT dose survey.

Authors:  J Clarke; K Cranley; J Robinson; P H Smith; A Workman
Journal:  Br J Radiol       Date:  2000-01       Impact factor: 3.039

2.  A survey of patient dose and image quality for computed tomography scanners in Wales.

Authors:  P Hiles; S Brennen; S Scott; J Davies
Journal:  J Radiol Prot       Date:  2001-12       Impact factor: 1.394

3.  Variations in radiation dose between the same model of multislice CT scanner at different hospitals.

Authors:  C J Koller; J P Eatough; A Bettridge
Journal:  Br J Radiol       Date:  2003-11       Impact factor: 3.039

4.  The effect of unrealistic thyroid vertical position on thyroid dose in the MIRD phantom.

Authors:  Choonsik Lee; Jaiki Lee; Choonik Lee
Journal:  Med Phys       Date:  2004-07       Impact factor: 4.071

5.  A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms.

Authors:  J J DeMarco; C H Cagnon; D D Cody; D M Stevens; C H McCollough; J O'Daniel; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2005-08-11       Impact factor: 3.609

6.  A Monte Carlo-based method to estimate radiation dose from spiral CT: from phantom testing to patient-specific models.

Authors:  G Jarry; J J DeMarco; U Beifuss; C H Cagnon; M F McNitt-Gray
Journal:  Phys Med Biol       Date:  2003-08-21       Impact factor: 3.609

7.  A CT-based Monte Carlo simulation tool for dosimetry planning and analysis.

Authors:  J J DeMarco; T D Solberg; J B Smathers
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

8.  A comparison of radiation dose measured in CT dosimetry phantoms with calculations using EGS4 and voxel-based computational models.

Authors:  M Caon; G Bibbo; J Pattison
Journal:  Phys Med Biol       Date:  1997-01       Impact factor: 3.609

9.  An EGS4-ready tomographic computational model of a 14-year-old female torso for calculating organ doses from CT examinations.

Authors:  M Caon; G Bibbo; J Pattison
Journal:  Phys Med Biol       Date:  1999-09       Impact factor: 3.609

Review 10.  Voxel-based computational models of real human anatomy: a review.

Authors:  Martin Caon
Journal:  Radiat Environ Biophys       Date:  2004-01-17       Impact factor: 1.925

View more
  32 in total

1.  Reduction in radiation doses from paediatric CT scans in Great Britain.

Authors:  Choonsik Lee; Mark S Pearce; Jane A Salotti; Richard W Harbron; Mark P Little; Kieran McHugh; Claire-Louise Chapple; Amy Berrington de Gonzalez
Journal:  Br J Radiol       Date:  2016-02-11       Impact factor: 3.039

2.  Monte Carlo calculation of patient organ doses from computed tomography.

Authors:  Takeshi Oono; Fujio Araki; Shoya Tsuduki; Keiichi Kawasaki
Journal:  Radiol Phys Technol       Date:  2013-11-30

3.  Size-specific dose estimate (SSDE) provides a simple method to calculate organ dose for pediatric CT examinations.

Authors:  Bria M Moore; Samuel L Brady; Amy E Mirro; Robert A Kaufman
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

4.  Patient-based estimation of organ dose for a population of 58 adult patients across 13 protocol categories.

Authors:  Pooyan Sahbaee; W Paul Segars; Ehsan Samei
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

5.  Evaluation of dose conversion coefficients for an eight-year-old Iranian male phantom undergoing computed tomography.

Authors:  Parisa Akhlaghi; Hashem Miri Hakimabad; Laleh Rafat Motavalli
Journal:  Radiat Environ Biophys       Date:  2015-06-17       Impact factor: 1.925

6.  Monte Carlo simulations of adult and pediatric computed tomography exams: validation studies of organ doses with physical phantoms.

Authors:  Daniel J Long; Choonsik Lee; Christopher Tien; Ryan Fisher; Matthew R Hoerner; David Hintenlang; Wesley E Bolch
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

7.  BODY SIZE-SPECIFIC EFFECTIVE DOSE CONVERSION COEFFICIENTS FOR CT SCANS.

Authors:  Anna Romanyukha; Les Folio; Stephanie Lamart; Steven L Simon; Choonsik Lee
Journal:  Radiat Prot Dosimetry       Date:  2016-01-10       Impact factor: 0.972

8.  Evaluation of Effective Dose from CT Scans for Overweight and Obese Adult Patients Using the VirtualDose Software.

Authors:  Baohui Liang; Yiming Gao; Zhi Chen; X George Xu
Journal:  Radiat Prot Dosimetry       Date:  2017-04-25       Impact factor: 0.972

9.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

10.  A prospective study of medical diagnostic radiography and risk of thyroid cancer.

Authors:  Gila Neta; Preetha Rajaraman; Amy Berrington de Gonzalez; Michele M Doody; Bruce H Alexander; Dale Preston; Steven L Simon; Dunstana Melo; Jeremy Miller; D Michal Freedman; Martha S Linet; Alice J Sigurdson
Journal:  Am J Epidemiol       Date:  2013-03-25       Impact factor: 4.897

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.