PURPOSE: The orientation of the acetabular component in metal-on-metal hip resurfacing arthroplasty affects wear rate and hence failure. This study aimed to establish if interpretation of pelvic radiographs with TraumaCad software can provide a reliable alternative to CT in measuring the acetabular inclination and version.Entities:
Mesh:
Year: 2013 PMID: 23577620 PMCID: PMC3637092 DOI: 10.1186/1749-799X-8-8
Source DB: PubMed Journal: J Orthop Surg Res ISSN: 1749-799X Impact factor: 2.359
Figure 1Short axis (S) of projected ellipse and total length (TL) of projected cup cross-section provide the S/TL ratio from which the version is calculated.
Figure 2Technique for measuring version and inclination of the acetabular component using CT reconstructions.
Correlation between CT and TraumaCad
| | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Inclination | PCC* | 0.904 | 0.917 | 0.949 | 0.961 | 0.969 | 0.953 | 0.962 | 0.947 |
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||
| Version | PCC* | 0.909 | 0.865 | 0.892 | 0.907 | 0.886 | 0.891 | 0.880 | 0.913 |
| <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||
* Pearson’s Correlation Coefficient.
Figure 3Relationship between measurements of inclination from CT and TraumaCad. The data plots represent each observer’s first measurements with a line of best fit.
Figure 4Relationship between measurements of anteversion from CT and TraumaCad. The data plots represent each observer’s first measurements with a line of best fit.
Intra-observer reliability of TraumaCad measurements
| | ||||||||
|---|---|---|---|---|---|---|---|---|
| Inclination | 0.992 | <0.001 | 0.997 | <0.001 | 0.992 | <0.001 | 0.991 | <0.001 |
| Version | 0.977 | <0.001 | 0.989 | <0.001 | 0.992 | <0.001 | 0.901 | <0.001 |
*intra-class correlation co-efficient (>0.7 = good correlation).
Inter-observer reliability of TraumaCad measurements
| | ||||
|---|---|---|---|---|
| Inclination | 0.971 | <0.001 | 0.976 | <0.001 |
| Version | 0.954 | <0.001 | 0.962 | <0.001 |
*intra-class correlation co-efficient (>0.7 = good correlation).
Figure 5Divergence of the X-Ray beam contributes to underestimation of version. The distance between the femoral heads is assumed to be 25 cm and the X-Ray source is 100 cm from the patient.