The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip.

D J Langton1, S S Jameson, T J Joyce, J Webb, A V F Nargol.   

Abstract

Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (<or= 51 mm) femoral components were significantly higher than in those with the larger (>or= 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18757952     DOI: 10.1302/0301-620X.90B9.20785

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  77 in total

1.  Hip resurfacing revision rates: radiological audit of risk factors.

Authors:  N Ramisetty; K M Krishnan; P F Partington
Journal:  Ann R Coll Surg Engl       Date:  2011-05       Impact factor: 1.891

2.  Reduced articular surface of one-piece cups: a cause of runaway wear and early failure.

Authors:  William L Griffin; Christopher J Nanson; Bryan D Springer; Matthew A Davies; Thomas K Fehring
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

3.  Failure of a metal on metal hip prosthesis presenting as a destructive soft tissue mass due to ALVAL.

Authors:  Jonathan R B Hutt; Constant Busch; Rod A Hughes
Journal:  Rheumatol Int       Date:  2010-11-27       Impact factor: 2.631

4.  Serum metal ion levels after rotating-hinge knee arthroplasty: comparison between a standard device and a megaprosthesis.

Authors:  Joerg Friesenbichler; Werner Maurer-Ertl; Patrick Sadoghi; Thomas Lovse; Reinhard Windhager; Andreas Leithner
Journal:  Int Orthop       Date:  2011-07-14       Impact factor: 3.075

5.  Surgical variables influence metal ion levels after hip resurfacing.

Authors:  Nicholas M Desy; Stephane G Bergeron; Alain Petit; Olga L Huk; John Antoniou
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

6.  High cup angle and microseparation increase the wear of hip surface replacements.

Authors:  Ian J Leslie; Sophie Williams; Graham Isaac; Eileen Ingham; John Fisher
Journal:  Clin Orthop Relat Res       Date:  2009-04-11       Impact factor: 4.176

7.  Does CT-based navigation improve the long-term survival in ceramic-on-ceramic THA?

Authors:  Nobuhiko Sugano; Masaki Takao; Takashi Sakai; Takashi Nishii; Hidenobu Miki
Journal:  Clin Orthop Relat Res       Date:  2012-11       Impact factor: 4.176

8.  The learning curve for adopting hip resurfacing among hip specialists.

Authors:  Ryan M Nunley; Jinjun Zhu; Peter J Brooks; C Anderson Engh; Stephen J Raterman; John S Rogerson; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2010-02       Impact factor: 4.176

9.  A more reliable method to assess acetabular component position.

Authors:  John V Tiberi; Nicholas Pulos; Michael Kertzner; Thomas P Schmalzried; Thomas P Schmlazried
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

Review 10.  Management of metal-on-metal hip implant patients: Who, when and how to revise?

Authors:  Reshid Berber; John A Skinner; Alister J Hart
Journal:  World J Orthop       Date:  2016-05-18
View more