D J Langton1, S S Jameson, T J Joyce, J Webb, A V F Nargol.
Abstract
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (<or= 51 mm) femoral components were significantly higher than in those with the larger (>or= 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18757952 DOI: 10.1302/0301-620X.90B9.20785
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X