Literature DB >> 16289687

Position of the acetabular cup -- accuracy of radiographic calculation compared to CT-based measurement.

Thomas Kalteis1, Martin Handel, Thomas Herold, Lars Perlick, Christian Paetzel, Joachim Grifka.   

Abstract

OBJECTIVE: A variety of mathematical and trigonometric methods has been described for determining the position of the acetabular cups from conventional radiographs. However, these formulae are subject to unduly large inaccuracies. The aim of the study was to compare the reliability and the accuracy of radiological and CT-based determination of acetabular cup position.
MATERIALS AND METHODS: The positions of acetabular cups of 31 patients were calculated in conventional plain radiographs of the hip using the method described by Widmer. Further, in all patients computed tomograms of the pelvis were performed and the cup position was measured with the aid of a CT-based computer-assisted navigation software. As reference values inclination and anteversion of the cups were calculated in 3D reconstructions of the pelvis with the aid of an image processing software.
RESULTS: The radiological measurement as well as the CT-based method showed good intra- and inter-observer reliability and no significant difference in the calculation of the inclination (p=0.409). However, CT-based determination of anteversion was significantly more exact than radiological measurement (p<0.001). The calculation of the cup anteversion from the X-rays showed serious deviations from the reference method and a substantial error range (X-ray: mean deviation +1.74 degrees, range -16.6 degrees to +29.8 degrees , S.D. +/-9.32 degrees; CT-based: mean deviation -0.74 degrees, range -6.6 degrees to +5.3 degrees, S.D. +/-2.87).
CONCLUSION: For any clinical problem or for clinical studies in which acetabular positions of acetabular cups have to be exactly determined, CT-based measuring methods are obviously the method of choice. Evaluations based only on conventional plain X-rays and calculation of the acetabular cup position using the formula described by Widmer must be regarded as unreliable, particularly, because of problems in measuring the anteversion.

Entities:  

Mesh:

Year:  2005        PMID: 16289687     DOI: 10.1016/j.ejrad.2005.10.003

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  52 in total

1.  Measuring acetabular component version after THA: CT or plain radiograph?

Authors:  Benjamin McArthur; Michael Cross; Christina Geatrakas; David Mayman; Bernard Ghelman
Journal:  Clin Orthop Relat Res       Date:  2012-03-01       Impact factor: 4.176

2.  A new method of measuring acetabular cup anteversion on simulated radiographs.

Authors:  Vikas Bachhal; Nipun Jindal; Gaurav Saini; Radheshyam Sament; Vishal Kumar; Devendra Chouhan; Mandeep Dhillon
Journal:  Int Orthop       Date:  2012-05-31       Impact factor: 3.075

3.  Accuracy and reliability of different methods to evaluate the acetabular cup version from plain radiographs.

Authors:  Maximilian Haenle; Wolfram Mittelmeier; Rodrigo Barbano; Klaus Wörtler; Roger Scholz; Rainer Bader
Journal:  Surg Radiol Anat       Date:  2010-06-25       Impact factor: 1.246

4.  Intraoperative radiographs for placing acetabular components in hip resurfacing arthroplasty.

Authors:  Thomas P Gross; Fei Liu; Lee Webb
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

5.  How does pelvic rotation or tilt affect radiographic measurement of acetabular component inclination angle during THA?

Authors:  Chris R Mellano; Andrew I Spitzer
Journal:  J Orthop       Date:  2015-06-23

6.  Assessment of cup position from plain radiographs: impact of pelvic tilting.

Authors:  M Haenle; A Heitner; W Mittelmeier; R Barbano; R Scholz; E Steinhauser; R Bader
Journal:  Surg Radiol Anat       Date:  2007-01-10       Impact factor: 1.246

7.  High incidence of squeaking in THAs with alumina ceramic-on-ceramic bearings.

Authors:  J C Keurentjes; R M Kuipers; D J Wever; B W Schreurs
Journal:  Clin Orthop Relat Res       Date:  2008-02-26       Impact factor: 4.176

8.  Navigated non-image-based positioning of the acetabulum during total hip replacement.

Authors:  Jean-Yves Jenny; Cyril Boeri; Jean-Claude Dosch; Marius Uscatu; Eugen Ciobanu
Journal:  Int Orthop       Date:  2007-11-13       Impact factor: 3.075

9.  Validation of a simple radiographic method to determine variations in pelvic and acetabular cup sagittal plane alignment after total hip arthroplasty.

Authors:  Won Yong Shon; Siddhartha Gupta; Sandeep Biswal; Chang Yong Hur; Nirmal Jajodia; Suk Joo Hong; Jae Sung Myung
Journal:  Skeletal Radiol       Date:  2008-08-07       Impact factor: 2.199

Review 10.  Imaging and navigation measurement of acetabular component position in THA.

Authors:  Zhinian Wan; Aamer Malik; Branislav Jaramaz; Lisa Chao; Lawrence D Dorr
Journal:  Clin Orthop Relat Res       Date:  2008-11-01       Impact factor: 4.176