Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms.

Lifeng Yu1, Shuai Leng, Lingyun Chen, James M Kofler, Rickey E Carter, Cynthia H McCollough.   

Abstract

PURPOSE: Efficient optimization of CT protocols demands a quantitative approach to predicting human observer performance on specific tasks at various scan and reconstruction settings. The goal of this work was to investigate how well a channelized Hotelling observer (CHO) can predict human observer performance on 2-alternative forced choice (2AFC) lesion-detection tasks at various dose levels and two different reconstruction algorithms: a filtered-backprojection (FBP) and an iterative reconstruction (IR) method.
METHODS: A 35 × 26 cm(2) torso-shaped phantom filled with water was used to simulate an average-sized patient. Three rods with different diameters (small: 3 mm; medium: 5 mm; large: 9 mm) were placed in the center region of the phantom to simulate small, medium, and large lesions. The contrast relative to background was -15 HU at 120 kV. The phantom was scanned 100 times using automatic exposure control each at 60, 120, 240, 360, and 480 quality reference mAs on a 128-slice scanner. After removing the three rods, the water phantom was again scanned 100 times to provide signal-absent background images at the exact same locations. By extracting regions of interest around the three rods and on the signal-absent images, the authors generated 21 2AFC studies. Each 2AFC study had 100 trials, with each trial consisting of a signal-present image and a signal-absent image side-by-side in randomized order. In total, 2100 trials were presented to both the model and human observers. Four medical physicists acted as human observers. For the model observer, the authors used a CHO with Gabor channels, which involves six channel passbands, five orientations, and two phases, leading to a total of 60 channels. The performance predicted by the CHO was compared with that obtained by four medical physicists at each 2AFC study.
RESULTS: The human and model observers were highly correlated at each dose level for each lesion size for both FBP and IR. The Pearson's product-moment correlation coefficients were 0.986 [95% confidence interval (CI): 0.958-0.996] for FBP and 0.985 (95% CI: 0.863-0.998) for IR. Bland-Altman plots showed excellent agreement for all dose levels and lesions sizes with a mean absolute difference of 1.0% ± 1.1% for FBP and 2.1% ± 3.3% for IR.
CONCLUSIONS: Human observer performance on a 2AFC lesion detection task in CT with a uniform background can be accurately predicted by a CHO model observer at different radiation dose levels and for both FBP and IR methods.

Entities:  

Mesh:

Year:  2013        PMID: 23556902      PMCID: PMC3618092          DOI: 10.1118/1.4794498

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  49 in total

1.  Introduction to grayscale calibration and related aspects of medical imaging grade liquid crystal displays.

Authors:  Kenneth A Fetterly; Hartwig R Blume; Michael J Flynn; Ehsan Samei
Journal:  J Digit Imaging       Date:  2008-06       Impact factor: 4.056

2.  Channelized-ideal observer using Laguerre-Gauss channels in detection tasks involving non-Gaussian distributed lumpy backgrounds and a Gaussian signal.

Authors:  Subok Park; Harrison H Barrett; Eric Clarkson; Matthew A Kupinski; Kyle J Myers
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2007-12       Impact factor: 2.129

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

4.  Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images.

Authors:  Samuel Richard; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

5.  Adaptive SPECT.

Authors:  Harrison H Barrett; Lars R Furenlid; Melanie Freed; Jacob Y Hesterman; Matthew A Kupinski; Eric Clarkson; Meredith K Whitaker
Journal:  IEEE Trans Med Imaging       Date:  2008-06       Impact factor: 10.048

6.  Evaluation of z-axis resolution and image noise for nonconstant velocity spiral CT data reconstructed using a weighted 3D filtered backprojection (WFBP) reconstruction algorithm.

Authors:  Jodie A Christner; Karl Stierstorfer; Andrew N Primak; Christian D Eusemann; Thomas G Flohr; Cynthia H McCollough
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

7.  The efficiency of the human observer for lesion detection and localization in emission tomography.

Authors:  Bin Liu; Lili Zhou; Santosh Kulkarni; Gene Gindi
Journal:  Phys Med Biol       Date:  2009-04-08       Impact factor: 3.609

8.  CT with a computer-simulated dose reduction technique for detection of pediatric nephroureterolithiasis: comparison of standard and reduced radiation doses.

Authors:  Boaz Karmazyn; Donald P Frush; Kimberly E Applegate; Charles Maxfield; Mervyn D Cohen; Robert P Jones
Journal:  AJR Am J Roentgenol       Date:  2009-01       Impact factor: 3.959

9.  Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies.

Authors:  Sarabjeet Singh; Mannudeep K Kalra; Michael A Moore; Randheer Shailam; Bob Liu; Thomas L Toth; Ellen Grant; Sjirk J Westra
Journal:  Radiology       Date:  2009-05-12       Impact factor: 11.105

10.  Collimator optimization in SPECT based on a joint detection and localization task.

Authors:  Lili Zhou; Gene Gindi
Journal:  Phys Med Biol       Date:  2009-06-26       Impact factor: 3.609

View more
  54 in total

1.  Computational and human observer image quality evaluation of low dose, knowledge-based CT iterative reconstruction.

Authors:  Brendan L Eck; Rachid Fahmi; Kevin M Brown; Stanislav Zabic; Nilgoun Raihani; Jun Miao; David L Wilson
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  Anthropomorphic model observer performance in three-dimensional detection task for low-contrast computed tomography.

Authors:  Alexandre Ba; Miguel P Eckstein; Damien Racine; Julien G Ott; Francis Verdun; Sabine Kobbe-Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2015-12-29

3.  Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.

Authors:  Alexandre Ba; Craig K Abbey; Damien Racine; Anaïs Viry; Francis R Verdun; Sabine Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-20

4.  A Web-Based Software Platform for Efficient and Quantitative CT Image Quality Assessment and Protocol Optimization.

Authors:  Mingdong Fan; Theodore Thayib; Liqiang Ren; Scott Hsieh; Cynthia McCollough; David Holmes; Lifeng Yu
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

5.  Optimization of digital breast tomosynthesis (DBT) acquisition parameters for human observers: effect of reconstruction algorithms.

Authors:  Rongping Zeng; Aldo Badano; Kyle J Myers
Journal:  Phys Med Biol       Date:  2017-02-02       Impact factor: 3.609

6.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

7.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

Review 8.  Task-based measures of image quality and their relation to radiation dose and patient risk.

Authors:  Harrison H Barrett; Kyle J Myers; Christoph Hoeschen; Matthew A Kupinski; Mark P Little
Journal:  Phys Med Biol       Date:  2015-01-07       Impact factor: 3.609

9.  Determination of Optimal Image Type and Lowest Detectable Concentration for Iodine Detection on a Photon Counting Detector-Based Multi-Energy CT System.

Authors:  Wei Zhou; Rachel Schornak; Gregory Michalak; Jayse Weaver; Dilbar Abdurakhimova; Andrea Ferrero; Kenneth A Fetterly; Cynthia H McCollough; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03

10.  Synthetic breast phantoms from patient based eigenbreasts.

Authors:  Gregory M Sturgeon; Subok Park; William Paul Segars; Joseph Y Lo
Journal:  Med Phys       Date:  2017-10-19       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.