Evaluation of z-axis resolution and image noise for nonconstant velocity spiral CT data reconstructed using a weighted 3D filtered backprojection (WFBP) reconstruction algorithm.

Jodie A Christner1, Karl Stierstorfer, Andrew N Primak, Christian D Eusemann, Thomas G Flohr, Cynthia H McCollough.   

Abstract

PURPOSE: To determine the constancy of z-axis spatial resolution, CT number, image noise, and the potential for image artifacts for nonconstant velocity spiral CT data reconstructed using a flexibly weighted 3D filtered backprojection (WFBP) reconstruction algorithm.
METHODS: A WFBP reconstruction algorithm was used to reconstruct stationary (axial, pitch=0), constant velocity spiral (pitch = 0.35-1.5) and nonconstant velocity spiral CT data acquired using a 128 x 0.6 mm acquisition mode (38.4 mm total detector length, z-flying focal spot technique), and a gantry rotation time of 0.30 s. Nonconstant velocity scans used the system's periodic spiral mode, where the table moved in and out of the gantry in a cyclical manner. For all scan types, the volume CTDI was 10 mGy. Measurements of CT number, image noise, and the slice sensitivity profile were made for all scan types as a function of the nominal slice width, table velocity, and position within the scan field of view. A thorax phantom was scanned using all modes and reconstructed transverse and coronal plane images were compared.
RESULTS: Negligible differences in slice thickness, CT number, noise, or artifacts were found between scan modes for data taken at two positions within the scan field of view. For nominal slices of 1.0-3.0 mm, FWHM values of the slice sensitivity profiles were essentially independent of the scan type. For periodic spiral scans, FWHM values measured at the center of the scan range were indistinguishable from those taken 5 mm from one end of the scan range. All CT numbers were within +/- 5 HU, and CT number and noise values were similar for all scan modes assessed. A slight increase in noise and artifact level was observed 5 mm from the start of the scan on the first pass of the periodic spiral. On subsequent passes, noise and artifact level in the transverse and coronal plane images were the same for all scan modes.
CONCLUSIONS: Nonconstant velocity periodic spiral scans can achieve z-axis spatial resolution, CT number accuracy, image noise and artifact level equivalent to those for stationary (axial), and constant velocity spiral scans. Thus, periodic spiral scans are expected to allow assessment of four-dimensional CT data for scan lengths greater than the detector width without sacrificing image quality.

Mesh:

Year:  2010        PMID: 20229899     DOI: 10.1118/1.3271110

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Comprehensive assessment of the slice sensitivity profiles in breast tomosynthesis and breast CT.

Authors:  Anita Nosratieh; Kai Yang; Shadi Aminololama-Shakeri; John M Boone
Journal:  Med Phys       Date:  2012-12       Impact factor: 4.071

2.  CT perfusion technique for assessment of early kidney allograft dysfunction: preliminary results.

Authors:  A Helck; M Wessely; M Notohamiprodjo; U Schönermarck; E Klotz; M Fischereder; F Schön; K Nikolaou; D A Clevert; M Reiser; C Becker
Journal:  Eur Radiol       Date:  2013-05-10       Impact factor: 5.315

3.  Porcine ex vivo liver phantom for dynamic contrast-enhanced computed tomography: development and initial results.

Authors:  Scott M Thompson; Juan C Ramirez-Giraldo; Bruce Knudsen; Joseph P Grande; Jodie A Christner; Man Xu; David A Woodrum; Cynthia H McCollough; Matthew R Callstrom
Journal:  Invest Radiol       Date:  2011-09       Impact factor: 6.016

4.  Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms.

Authors:  Lifeng Yu; Shuai Leng; Lingyun Chen; James M Kofler; Rickey E Carter; Cynthia H McCollough
Journal:  Med Phys       Date:  2013-04       Impact factor: 4.071

5.  Human Imaging With Photon Counting-Based Computed Tomography at Clinical Dose Levels: Contrast-to-Noise Ratio and Cadaver Studies.

Authors:  Ralf Gutjahr; Ahmed F Halaweish; Zhicong Yu; Shuai Leng; Lifeng Yu; Zhoubo Li; Steven M Jorgensen; Erik L Ritman; Steffen Kappler; Cynthia H McCollough
Journal:  Invest Radiol       Date:  2016-07       Impact factor: 6.016

6.  A comprehensive assessment of physical image quality of five different scanners for head CT imaging as clinically used at a single hospital centre-A phantom study.

Authors:  Patrizio Barca; Fabio Paolicchi; Giacomo Aringhieri; Federica Palmas; Daniela Marfisi; Maria Evelina Fantacci; Davide Caramella; Marco Giannelli
Journal:  PLoS One       Date:  2021-01-14       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.