Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Lifeng Yu1, Thomas J Vrieze1, Shuai Leng1, Joel G Fletcher1, Cynthia H McCollough1.   

Abstract

PURPOSE: The spatial resolution of iterative reconstruction (IR) in computed tomography (CT) is contrast- and noise-dependent because of the nonlinear regularization. Due to the severe noise contamination, it is challenging to perform precise spatial-resolution measurements at very low-contrast levels. The purpose of this study was to measure the spatial resolution of a commercially available IR method using ensemble-averaged images acquired from repeated scans.
METHODS: A low-contrast phantom containing three rods (7, 14, and 21 HU below background) was scanned on a 128-slice CT scanner at three dose levels (CTDIvol = 16, 8, and 4 mGy). Images were reconstructed using two filtered-backprojection (FBP) kernels (B40 and B20) and a commercial IR method (sinogram affirmed iterative reconstruction, SAFIRE, Siemens Healthcare) with two strength settings (I40-3 and I40-5). The same scan was repeated 100 times at each dose level. The modulation transfer function (MTF) was calculated based on the edge profile measured on the ensemble-averaged images.
RESULTS: The spatial resolution of the two FBP kernels, B40 and B20, remained relatively constant across contrast and dose levels. However, the spatial resolution of the two IR kernels degraded relative to FBP as contrast or dose level decreased. For a given dose level at 16 mGy, the MTF50% value normalized to the B40 kernel decreased from 98.4% at 21 HU to 88.5% at 7 HU for I40-3 and from 97.6% to 82.1% for I40-5. At 21 HU, the relative MTF50% value decreased from 98.4% at 16 mGy to 90.7% at 4 mGy for I40-3 and from 97.6% to 85.6% for I40-5.
CONCLUSIONS: A simple technique using ensemble averaging from repeated CT scans can be used to measure the spatial resolution of IR techniques in CT at very low contrast levels. The evaluated IR method degraded the spatial resolution at low contrast and high noise levels.

Mesh:

Year:  2015        PMID: 25979020      PMCID: PMC4401802          DOI: 10.1118/1.4916802

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Abdominal CT: comparison of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53 patients.

Authors:  Yoshiko Sagara; Amy K Hara; William Pavlicek; Alvin C Silva; Robert G Paden; Qing Wu
Journal:  AJR Am J Roentgenol       Date:  2010-09       Impact factor: 3.959

2.  Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique.

Authors:  Priyanka Prakash; Mannudeep K Kalra; Avinash K Kambadakone; Homer Pien; Jiang Hsieh; Michael A Blake; Dushyant V Sahani
Journal:  Invest Radiol       Date:  2010-04       Impact factor: 6.016

3.  A three-dimensional statistical approach to improved image quality for multislice helical CT.

Authors:  Jean-Baptiste Thibault; Ken D Sauer; Charles A Bouman; Jiang Hsieh
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

4.  Noise-resolution tradeoffs in x-ray CT imaging: a comparison of penalized alternating minimization and filtered backprojection algorithms.

Authors:  Joshua D Evans; David G Politte; Bruce R Whiting; Joseph A O'Sullivan; Jeffrey F Williamson
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

5.  A methodology for image quality evaluation of advanced CT systems.

Authors:  Joshua M Wilson; Olav I Christianson; Samuel Richard; Ehsan Samei
Journal:  Med Phys       Date:  2013-03       Impact factor: 4.071

6.  Pilot study of detection, radiologist confidence and image quality with sinogram-affirmed iterative reconstruction at half-routine dose level.

Authors:  Joel G Fletcher; William R Krueger; David M Hough; James E Huprich; Jeff L Fidler; Jia Wang; Maria M Shiung; W Scott Harmsen; Katharine L Grant; Cynthia H McCollough
Journal:  J Comput Assist Tomogr       Date:  2013 Mar-Apr       Impact factor: 1.826

7.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

8.  Statistical model based iterative reconstruction (MBIR) in clinical CT systems. Part II. Experimental assessment of spatial resolution performance.

Authors:  Ke Li; John Garrett; Yongshuai Ge; Guang-Hong Chen
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

9.  Correlation between human and model observer performance for discrimination task in CT.

Authors:  Yi Zhang; Shuai Leng; Lifeng Yu; Rickey E Carter; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2014-05-30       Impact factor: 3.609

10.  Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain.

Authors:  Shuai Leng; Lifeng Yu; Yi Zhang; Rickey Carter; Alicia Y Toledano; Cynthia H McCollough
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

View more
  21 in total

1.  Impact of number of repeated scans on model observer performance for a low-contrast detection task in computed tomography.

Authors:  Chi Ma; Lifeng Yu; Baiyu Chen; Christopher Favazza; Shuai Leng; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-05-26

2.  Predicting detection performance with model observers: Fourier domain or spatial domain?

Authors:  Baiyu Chen; Lifeng Yu; Shuai Leng; James Kofler; Christopher Favazza; Thomas Vrieze; Cynthia McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-30

3.  Tilted-wire method for measuring resolution properties of CT images under extremely low-contrast and high-noise conditions.

Authors:  Chiaki Tominaga; Hiroki Azumi; Mitsunori Goto; Masaaki Taura; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2018-02-23

4.  Rapid measurement of the low contrast detectability of CT scanners.

Authors:  Akinyinka Omigbodun; J Y Vaishnav; Scott S Hsieh
Journal:  Med Phys       Date:  2021-01-13       Impact factor: 4.071

5.  Spectral prior image constrained compressed sensing (spectral PICCS) for photon-counting computed tomography.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-08-23       Impact factor: 3.609

6.  Correlation between a 2D channelized Hotelling observer and human observers in a low-contrast detection task with multislice reading in CT.

Authors:  Lifeng Yu; Baiyu Chen; James M Kofler; Christopher P Favazza; Shuai Leng; Matthew A Kupinski; Cynthia H McCollough
Journal:  Med Phys       Date:  2017-07-13       Impact factor: 4.071

7.  An effective noise reduction method for multi-energy CT images that exploit spatio-spectral features.

Authors:  Zhoubo Li; Shuai Leng; Lifeng Yu; Armando Manduca; Cynthia H McCollough
Journal:  Med Phys       Date:  2017-04-12       Impact factor: 4.071

8.  Lesion insertion in the projection domain: Methods and initial results.

Authors:  Baiyu Chen; Shuai Leng; Lifeng Yu; Zhicong Yu; Chi Ma; Cynthia McCollough
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

9.  Predicting image properties in penalized-likelihood reconstructions of flat-panel CBCT.

Authors:  Wenying Wang; Grace J Gang; Jeffrey H Siewerdsen; J Webster Stayman
Journal:  Med Phys       Date:  2018-11-20       Impact factor: 4.071

10.  Can conclusions drawn from phantom-based image noise assessments be generalized to in vivo studies for the nonlinear model-based iterative reconstruction method?

Authors:  Daniel Gomez-Cardona; Ke Li; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.