Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry.

Jeevan S Ladi1, Nitant A Shah.   

Abstract

PURPOSE: To compare corneal pachymetry assessment by the Galilei dual Scheimpflug analyzer with that done by ultrasound (US) pachymetry.
MATERIALS AND METHODS: Forty six patients (92 eyes) were subjected to corneal pachymetry assessment by Galilei dual Scheimpflug analyzer and US. All the readings were taken by a single operator. Intraoperator repeatability for the Galilei was assessed by taking 10 readings in one eye each of 10 patients. To study the interoperator reproducibility for the Galilei, two observers took a single reading in both the eyes of 25 patients.
RESULTS: The mean central corneal thickness (CCT) measured by US was 541.83 +/- 30.56 microm standard deviation (SD) and that measured by Galilei was 541.27 +/- 30.07 microm (SD). There was no statistically significant difference between both the methods (P < 0.001). The coefficient of repeatability was 0.43% while the coefficient of reproducibility was 0.377% for the Galilei.
CONCLUSION: Objective, noncontact measurement of the CCT with the Galilei dual Scheimpflug analyzer was convenient, had excellent intraoperator repeatability and interoperator reproducibility, and findings were similar to those obtained with standard US pachymetry.

Entities:  

Mesh:

Year:  2010        PMID: 20689192      PMCID: PMC2992912          DOI: 10.4103/0301-4738.67045

Source DB:  PubMed          Journal:  Indian J Ophthalmol        ISSN: 0301-4738            Impact factor:   1.848


Accurate measurement of central corneal thickness (CCT) has assumed significance in ophthalmic practice. It is important in corneal refractive surgery, especially laser assisted in situ keratomileusis (LASIK) which is the most commonly performed corneal procedure currently.[1] The CCT allows determination of the amount of stromal ablation which can be safely carried out minimizing the risk of iatrogenic keratectasia.[1] CCT is also used in glaucoma practice to modify the intraocular pressure reading for accuracy.[2] Applanation ultrasound (US) pachymetry is currently the most often method used for the measurement of CCT. This method has been reported to have a high degree of intraoperator and interoperator reproducibility.[3] However, placement of the probe on the corneal center is subjective and operator-dependent errors due to off-center placement (leading to thicker measurements) are a possibility.[45] Errors can be caused by indentation leading to slightly thinner readings.[6] In addition, disadvantages like patient discomfort, epithelial damage and risk of infection exist. Hence, it is not surprising that noncontact techniques to measure CCT are gaining popularity. Partial coherence interferometry (PCI),[7] optical coherence tomography (OCT),[8] scanning slit tomography/pachymetry[9] are some of the optical techniques introduced to measure CCT. The Galilei dual Scheimpflug analyzer (Ziemer, Switzerland), uses two rotating Scheimpflug cameras combined with a Placido disk to image the anterior segment of the eye. It is a noncontact instrument which provides data on anterior and posterior corneal topography, complete corneal pachymetry, lens densitometry and two-dimensional and three-dimensional anterior segment imaging. The purpose of this study was to report our initial experience in measuring CCT with this new device and compare it with the gold standard US pachymetry.

Materials and Methods

Forty six patients (92 eyes) who presented to our center were prospectively studied. Patients with ocular disease (other than refractive error), contact lens wearers and those with a history of previous eye surgery were excluded. All patients were subjected to a comprehensive ophthalmic examination including vision, refraction, slit lamp examination, CCT measurement with two methods, followed by measurement of intraocular pressure and dilated fundus examination. CCT measurements were first taken on the Galilei analyzer. Three readings were taken for each eye. A gap of 1 minute was given after each reading and the alignment was freshly done each time. Following this, the cornea was anesthetized with topical 0.5% proparacaine and five readings were taken with US pachymetry (Echorule, Biomedix, Bangalore, India). All the readings on the Galilei and US pachymeter were taken by a single trained optometrist. Galilei measurements were obtained as per the manufacturers’ instructions. The patient was comfortably seated with chin fully placed on the chin rest and forehead against the strap. The patient looked at the target (red spot) and was allowed to blink. The device was brought in focus by aligning the measurement head of the Galilei. Alignment was considered correct when the red cross (cross hair) passed through the white spots and the single red line touched the corneal epithelium. The iris was seen in sharp focus on the screen. The patient was asked to blink once, open the eye wide and the reading was taken. The time taken for scanning was 2–3 seconds. US pachymetry readings were taken by aligning the probe as perpendicularly as possible on the central cornea. Five readings were taken. The highest and lowest values were excluded and the mean of the remaining three were used for the analysis. The repeatability of the Galilei analyzer was studied. Ten successive scans were obtained by a single operator in one eye each of 10 patients. An interval of 1 minute was given between two readings and alignment was freshly done each time. Interoperator reproducibility was studied in the following manner. Two operators took a single reading of the right and left eyes of 25 patients. An interval of 5 minutes was given between the two operators. All the data were analyzed using the SPSS computer program for Windows (Version 11.5, SPSS, Inc., Chicago, Illinois).

Comparison of CCT

The results were entered as mean ± standard deviation (SD). All the data were tested for normality using skewness (acceptable range of normality is between –1 and +1) and kurtosis (acceptable range of normality is between –1 and +1). The statistical agreement between the two methods was assessed using interclass correlation coefficient. Bland–Altman plot was used to test the agreement between the two measuring techniques. A P value of less than 0.05 was considered as statistically significant.

Study of repeatability and reproducibility

The data were entered as mean (SD). All the data were tested for normality before analysis. In order to assess the pairwise statistical difference between the averages of two observers, the paired t-test was adopted. For each patient, the coefficient of repeatability was defined as the SD of the difference from the mean of the repeat measurements, divided by the mean response. The coefficient of interobserver reproducibility was defined as the SD of difference between the pairs of the measurements obtained by two observers, divided by the average of the means of each pair of observation. We have used definitions for repeatability and reproducibility adopted by the British Standards Institution as recommended by Bland and Altman.[10] A P value less than 0.05 was considered as statistically significant.

Results

The results of repeatability study are shown in Table 1. The mean coefficient of repeatability for 10 readings in 10 patients was 0.43% which demonstrates good reliability.
Table 1

Coefficients of repeatability for 10 patients using the Galilei analyzer

Patient IDCoefficient of repeatability (%)
10.15
20.19
30.35
40.25
50.32
60.44
70.79
80.21
91.29
100.34
Mean0.43
Coefficients of repeatability for 10 patients using the Galilei analyzer The mean CCT value taken in 50 eyes using the Galilei for observer 1 was 527.8 ± 33.7 μm, while it was 527.9 ± 31.9 μm for observer 2. There was no statistically significant difference between them. (P = 0.739). The coefficient of interoperator reproducibility for the Galilei was 0.377%. This and the scatter plot depicted in Fig. 1 indicate good interobserver reproducibility.
Figure 1

Scatter plot for comparing CCT between two observers with no change line. X and Y axes show CCT values measured by observer 2 and observer 1, respectively, in micrometers (either eye, n = 50)

Scatter plot for comparing CCT between two observers with no change line. X and Y axes show CCT values measured by observer 2 and observer 1, respectively, in micrometers (either eye, n = 50) The mean CCT obtained by US pachymetry for 92 eyes was 541.83 ± 30.56 μm (skewness –0.417 and kurtosis –0.312). The mean CCT by Galilei for 92 eyes was 541.27 ±30.07 μm (skewness –0.3 and kurtosis –0.151). Mean difference between the two methods was 0.55 μm. The interclass correlation coefficient was 0.978. There was no statistically significant difference between the CCT readings taken by Galilei and US (P < 0.001). Correlation between fellow eyes has been adjusted in the statistical analysis using multivariate linear regression. Fig. 2 is a scatter plot with a no change line to compare the CCT measurement by the two methods. Fig. 3 is a Bland–Altman plot and shows good agreement between Galilei and US. The 95% limits of agreement were –11.93 to +13.03. The plot indicates that there is no systematic bias between the two methods (P = 0.612 by linear regression method between difference and average by two methods.)
Figure 2

Scatter plot for comparing CCT between two methods with no change line. X axis: measurement of CCT with Galilei in micrometers. Y axis: measurement of CCT with US in micrometers

Figure 3

Bland–Altman Plot comparing the two modalities of measurement, viz., comparison between Galilei and ultrasound pachymetry

Scatter plot for comparing CCT between two methods with no change line. X axis: measurement of CCT with Galilei in micrometers. Y axis: measurement of CCT with US in micrometers Bland–Altman Plot comparing the two modalities of measurement, viz., comparison between Galilei and ultrasound pachymetry

Discussion

Noncontact techniques like PCI and OCT have been shown to have high intraoperator repeatability and interoperator reproducibility in literature.[78] This is important for a new instrument to gain widespread acceptance. Anterior segment imaging using Scheimpflug analyzers was reported initially using the EAS-1000 (Nidek Co. Ltd.).[11] This model acquired a single image of the anterior segment. The Pentacam was an improvement over this, having a single rotating camera which acquires upto 50 images in a scan. The Galilei analyzer has two rotating Scheimpflug cameras, 180° apart, which record simultaneously. This helps to compensate for errors associated with scans at an oblique angle and increases the accuracy for measuring not only the central cornea but also the peripheral cornea (in spite of eye micromovements which are unavoidable). Previous studies have shown the coefficient of repeatability to be 0.84% for the Pentacam, 0.33% for the optical low-coherence reflectometer (OLCR) pachymeter, and 0.71% for standard US pachymetry.[12] The OCT has been shown to have a coefficient of 2%.[8] In our study, the coefficient of repeatability for the Galilei was 0.43%. Menassa et al.[13] have reported a remarkably low intraobserver and interobserver variation, this being the only report on the Galilei in the literature so far. The coefficient of interobserver reproducibility was 0.37% in our study. This coefficient has been reported to be 1.10% for the Pentacam and 0.59% for the OLCR,[12] while it was 0.18% for OCT.[8] Another study by Lackner et al.[14] has shown the reproducibility by Pentacam to be the highest between Pentacam, Orbscan and US. Menassa et al.[13] have reported better interobserver reproducibility with the Galilei compared to Orbscan II. Previous studies have reported that CCT measurements in healthy eyes by different instruments can be significantly different. In a study of 34 normal eyes, Modis et al.[9] have reported mean CCT to be 547 ± 49 μm by noncontact specular microscopy, 580 ± 43 μm by US, 602 ± 59 μm by Orbscan and 640 ± 43 μm by contact specular microscopy. Chakrabartiet al.[15] have reported Orbscan measurements to be 28 μm higher than those of US. Due to the considerable difference, Orbscan manufacturers recommended incorporation of an acoustic factor of 0.92. This was calculated to compensate for the effect of tear film whose thickness gets included while taking CCT measurements. Indeed after applying the correction factor, Wong et al.[16] have reported CCT to be 555.96 ± 32 μm with Orbscan and 555.11 ± 35 μm with US. Also, Suzukiet al.[17] have reported CCT to be 548.1 ± 33 μm with US and 546.9 ± 35.4 μm with Orbscan. Ho et al.[18] have recommended the use of a custom acoustic factor specific to the study site. They have used an acoustic factor of 0.89 for Orbscan II to obtain readings compatible with US. Similarly, Menassa et al.[13] have used a correction factor of 0.96 for Orbscan II. Lackner et al.[14] have reported an underestimation of 9.8 μm with the Pentacam as compared to US. Barkana et al.[12] have reported an underestimation of 6.09 μm with the Pentacam as compared to US, which was statistically significant. Menassa et al.[13] have reported a mean difference of 6.8 μm between CCT readings by Galilei (551.7 ± 36.6) and US (558.5 ± 38.4), which was statistically significant. In our study, comparison of CCT measurements between the Galilei and US has shown that they differed by a mean of only 0.55 μm. This difference is not statistically significant. The 95% limits of agreement showed that the difference in measurements between the two methods was between –11.93 and +13.03 μm. This is remarkably close to the range of ±11 μm for the diurnal pachymetric variation of CCT. Noncontact methods would include the tear film thickness (5–7 μm) in CCT measurement. However, as suggested by Ho et al.,[18] US may also overestimate CCT due to the effect of analgesic eyedrops used before pachymetry. These can cause epithelial edema leading to increased corneal thickness. Further studies would be required to find out which method ultimately gives us the true CCT. In conclusion, the Galilei analyzer can measure CCT with an excellent repeatability and reproducibility. This noncontact method of examination can thus be delegated to nonmedical personnel like technicians or optometrists. The pachymetry readings with Galilei showed good correlation with those of US pachymetry and there was no statistically significant difference between them. To the best of our knowledge, this is the first study using the Galilei dual Scheimpflug analyzer in Indian Asian eyes.
  18 in total

1.  Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes.

Authors:  H S Chakrabarti; J P Craig; A Brahma; T Y Malik; C N McGhee
Journal:  J Cataract Refract Surg       Date:  2001-11       Impact factor: 3.351

2.  Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness.

Authors:  L Módis; A Langenbucher; B Seitz
Journal:  Cornea       Date:  2001-10       Impact factor: 2.651

3.  Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography.

Authors:  Sarah Muscat; Nicola McKay; Stuart Parks; Ewan Kemp; David Keating
Journal:  Invest Ophthalmol Vis Sci       Date:  2002-06       Impact factor: 4.799

4.  Intra-observer and inter-observer repeatability of anterior eye segment analysis system (EAS-1000) in anterior chamber configuration.

Authors:  Andrew K C Lam; Rufina Chan; George C Woo; Peter C K Pang; Roger Chiu
Journal:  Ophthalmic Physiol Opt       Date:  2002-11       Impact factor: 3.117

5.  Correlational study of central corneal thickness measurements on Hong Kong Chinese using optical coherence tomography, Orbscan and ultrasound pachymetry.

Authors:  A C-M Wong; C-C Wong; N S-Y Yuen; S-P Hui
Journal:  Eye (Lond)       Date:  2002-11       Impact factor: 3.775

6.  Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery.

Authors:  Oliver Findl; Katharina Kriechbaum; Stefan Sacu; Barbara Kiss; Kaija Polak; Johannes Nepp; Gebtraud Schild; Georg Rainer; Saskia Maca; Vanessa Petternel; Birgit Lackner; Wolfgang Drexler
Journal:  J Cataract Refract Surg       Date:  2003-10       Impact factor: 3.351

7.  Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry.

Authors:  Shigenobu Suzuki; Tetsuro Oshika; Kotaro Oki; Isao Sakabe; Aiko Iwase; Shiro Amano; Makoto Araie
Journal:  J Cataract Refract Surg       Date:  2003-07       Impact factor: 3.351

8.  Comparison of ultrasound pachymetry and partial coherence interferometry in the measurement of central corneal thickness.

Authors:  Georg Rainer; Vanessa Petternel; Oliver Findl; Leopold Schmetterer; Christian Skorpik; Alexandra Luksch; Wolfgang Drexler
Journal:  J Cataract Refract Surg       Date:  2002-12       Impact factor: 3.351

Review 9.  Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.

Authors:  M J Doughty; M L Zaman
Journal:  Surv Ophthalmol       Date:  2000 Mar-Apr       Impact factor: 6.048

10.  Central corneal pachymetry in patients undergoing laser in situ keratomileusis.

Authors:  F W Price; D L Koller; M O Price
Journal:  Ophthalmology       Date:  1999-11       Impact factor: 12.079

View more
  13 in total

1.  Difference in normal corneal thickness and curvature between Mongolian and Han nationalities.

Authors:  Hai-Xia Zhao; Li Zhang; Wen-Ying Guan
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

2.  Repeatability and comparability of corneal thickness measurements obtained from Dual Scheimpflug Analyzer and from ultrasonic pachymetry.

Authors:  Zaina N Al-Mohtaseb; Li Wang; Mitchell P Weikert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-02-17       Impact factor: 3.117

3.  Measurement of central corneal thickness and pre-corneal tear film thickness of rabbits using the Scheimpflug system.

Authors:  Jing Dong; Qiang Wu; Xiao-Gang Wang
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

Review 4.  A review of imaging modalities for detecting early keratoconus.

Authors:  Xuemin Zhang; Saleha Z Munir; Syed A Sami Karim; Wuqaas M Munir
Journal:  Eye (Lond)       Date:  2020-07-16       Impact factor: 3.775

5.  Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter.

Authors:  Turki M Almubrad; Uchechukwu L Osuagwu; Ibrahim Alabbadi; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2011-06-24

6.  Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster.

Authors:  Roma P Patel; Rahul T Pandit
Journal:  J Ophthalmol       Date:  2012-01-19       Impact factor: 1.909

7.  Comparison of central corneal thickness measurements using different imaging devices and ultrasound pachymetry.

Authors:  Ertugrul Can; Hilal Eser-Ozturk; Mustafa Duran; Tugba Cetinkaya; Nursen Arıturk
Journal:  Indian J Ophthalmol       Date:  2019-04       Impact factor: 1.848

8.  Evaluation of anterior segment parameters and possible influencing factors in normal subjects using a dual Scheimpflug analyzer.

Authors:  Xiaogang Wang; Jing Dong; Qiang Wu
Journal:  PLoS One       Date:  2014-05-16       Impact factor: 3.240

9.  Comment on: Intraocular pressure and anetrior segment anatomy after phacoemulsification surgery.

Authors:  Yakup Aksoy; Taner Kar; Abdullah Kaya
Journal:  Indian J Ophthalmol       Date:  2016-01       Impact factor: 1.848

10.  Comparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography.

Authors:  Soyeon Jung; Hee Seung Chin; Na Rae Kim; Kang Won Lee; Ji Won Jung
Journal:  J Ophthalmol       Date:  2017-12-10       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.