Common-mode differential-mode (CMDM) method for double-nuclear MR signal excitation and reception at ultrahigh fields.

Yong Pang1, Xiaoliang Zhang, Zhentian Xie, Chunsheng Wang, Daniel B Vigneron.   

Abstract

Double-tuned radio-frequency (RF) coils for heteronuclear mangentic resonance (MR) require sufficient electromagnetic isolation between the two resonators operating at two Larmor frequencies and independent tuning in order to attain highly efficient signal acquisition at each frequency. In this work, a novel method for double-tuned coil design at 7T based on the concept of common-mode differential-mode (CMDM) was developed and tested. Common mode (CM) and differential mode (DM) currents exist within two coupled parallel transmission lines, e.g., microstrip lines, yielding two different current distributions. The electromagnetic (EM) fields of the CM and DM are orthogonal to each other, and thus, the two modes are intrinsically EM decoupled. The modes can be tuned independently to desired frequencies, thus satisfying the requirement of dual-frequency MR applications. To demonstrate the feasibility and efficiency of the proposed CMDM technique, CMDM surface coils and volume coils using microstrip transmission line for (1)H and (13)C MRI/MRSI were designed, constructed, and tested at 7T. Bench test results showed that the isolations between the two frequency channels of the CMDM surface coil and volume coil were better than -30 and -25 dB, respectively. High quality MR phantom images were also obtained using the CMDM coils. The performance of the CMDM technique was validated through a comparison with the conventional two-pole design method at 7T. The proposed CMDM technique can be also implemented by using other coil techniques such as lumped element method, and can be applied to designing double-tuned parallel imaging coil arrays. Furthermore, if the two resonant modes of a CMDM coil were tuned to the same frequency, the CMDM coil becomes a quadrature coil due to the intrinsic orthogonal field distribution of CM and DM.

Entities:  

Mesh:

Year:  2011        PMID: 21693414      PMCID: PMC3277813          DOI: 10.1109/TMI.2011.2160192

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  58 in total

1.  Acquisition of human multislice MR images at 8 Tesla.

Authors:  A M Abduljalil; A Kangarlu; X Zhang; R E Burgess; P M Robitaille
Journal:  J Comput Assist Tomogr       Date:  1999 May-Jun       Impact factor: 1.826

2.  Radiofrequency shielding of surface coils at 4.0 T.

Authors:  K C Ong; H Wen; A S Chesnick; S Duewell; F A Jaffer; R S Balaban
Journal:  J Magn Reson Imaging       Date:  1995 Nov-Dec       Impact factor: 4.813

3.  Multiple tuning of birdcage resonators.

Authors:  S Amari; A M Uluğ; J Bornemann; P C van Zijl; P B Barker
Journal:  Magn Reson Med       Date:  1997-02       Impact factor: 4.668

4.  Determination of the rate of the glutamate/glutamine cycle in the human brain by in vivo 13C NMR.

Authors:  J Shen; K F Petersen; K L Behar; P Brown; T W Nixon; G F Mason; O A Petroff; G I Shulman; R G Shulman; D L Rothman
Journal:  Proc Natl Acad Sci U S A       Date:  1999-07-06       Impact factor: 11.205

5.  High frequency volume coils for clinical NMR imaging and spectroscopy.

Authors:  J T Vaughan; H P Hetherington; J O Otu; J W Pan; G M Pohost
Journal:  Magn Reson Med       Date:  1994-08       Impact factor: 4.668

6.  Dual-frequency, dual-quadrature, birdcage RF coil design with identical B1 pattern for sodium and proton imaging of the human brain at 1.5 T.

Authors:  G X Shen; F E Boada; K R Thulborn
Journal:  Magn Reson Med       Date:  1997-11       Impact factor: 4.668

7.  Noninvasive measurements of [1-(13)C]glycogen concentrations and metabolism in rat brain in vivo.

Authors:  I Y Choi; I Tkác; K Ugurbil; R Gruetter
Journal:  J Neurochem       Date:  1999-09       Impact factor: 5.372

8.  Detecting natural abundance carbon signal of NAA metabolite within 12-cm3 localized volume of human brain using 1H-[13C] NMR spectroscopy.

Authors:  W Chen; G Adriany; X H Zhu; R Gruetter; K Ugurbil
Journal:  Magn Reson Med       Date:  1998-08       Impact factor: 4.668

9.  3D 31P spectroscopic imaging of the human heart at 4.1 T.

Authors:  H P Hetherington; D J Luney; J T Vaughan; J W Pan; S L Ponder; O Tschendel; D B Twieg; G M Pohost
Journal:  Magn Reson Med       Date:  1995-03       Impact factor: 4.668

10.  Two configurations of the four-ring birdcage coil for 1H imaging and 1H-decoupled 31P spectroscopy of the human head.

Authors:  J Murphy-Boesch; R Srinivasan; L Carvajal; T R Brown
Journal:  J Magn Reson B       Date:  1994-02
View more
  16 in total

1.  A monopole/loop dual-tuned RF coil for ultrahigh field MRI.

Authors:  Xinqiang Yan; Rong Xue; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-08

2.  Quadrature transmit array design using single-feed circularly polarized patch antenna for parallel transmission in MR imaging.

Authors:  Yong Pang; Baiying Yu; Daniel B Vigneron; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

3.  Design and numerical evaluation of a volume coil array for parallel MR imaging at ultrahigh fields.

Authors:  Yong Pang; Ernest W H Wong; Baiying Yu; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

4.  Parallel imaging performance investigation of an 8-channel common-mode differential-mode (CMDM) planar array for 7T MRI.

Authors:  Xiaoqing Hu; Xiao Chen; Xin Liu; Hairong Zheng; Ye Li; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

5.  Flexible transceiver array for ultrahigh field human MR imaging.

Authors:  Bing Wu; Xiaoliang Zhang; Chunsheng Wang; Ye Li; Yong Pang; Jonathan Lu; Duan Xu; Sharmila Majumdar; Sarah J Nelson; Daniel B Vigneron
Journal:  Magn Reson Med       Date:  2012-01-13       Impact factor: 4.668

6.  A dual-tuned quadrature volume coil with mixed λ/2 and λ/4 microstrip resonators for multinuclear MRSI at 7 T.

Authors:  Yong Pang; Zhentian Xie; Duan Xu; Douglas A Kelley; Sarah J Nelson; Daniel B Vigneron; Xiaoliang Zhang
Journal:  Magn Reson Imaging       Date:  2011-11-03       Impact factor: 2.546

7.  Multi-channel microstrip transceiver arrays using harmonics for high field MR imaging in humans.

Authors:  Bing Wu; Chunsheng Wang; Jonathan Lu; Yong Pang; Sarah J Nelson; Daniel B Vigneron; Xiaoliang Zhang
Journal:  IEEE Trans Med Imaging       Date:  2011-08-30       Impact factor: 10.048

8.  Numerical Analysis of Human Sample Effect on RF Penetration and Liver MR Imaging at Ultrahigh Field.

Authors:  Yong Pang; Bing Wu; Chunsheng Wang; Daniel B Vigneron; Xiaoliang Zhang
Journal:  Concepts Magn Reson Part B Magn Reson Eng       Date:  2011-10       Impact factor: 1.176

9.  Precompensation for mutual coupling between array elements in parallel excitation.

Authors:  Yong Pang; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2011-12

10.  1 H-13 C independently tuned radiofrequency surface coil applied for in vivo hyperpolarized MRI.

Authors:  Peng Cao; Xiaoliang Zhang; Ilwoo Park; Chloe Najac; Sarah J Nelson; Sabrina Ronen; Peder E Z Larson
Journal:  Magn Reson Med       Date:  2015-11-24       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.