Xiaoqing Hu1, Xiao Chen1, Xin Liu1, Hairong Zheng1, Ye Li1, Xiaoliang Zhang1.
Abstract
An 8-channel planar phased array was proposed based on the common-mode differential-mode (CMDM) structure for ultrahigh field MRI. The parallel imaging performance of the 8-channel CMDM planar array was numerically investigated based on electromagnetic simulations and Cartesian sensitivity encoding (SENSE) reconstruction. The signal-to-noise ratio (SNR) of multichannel images combined using root-sum-of-squares (rSoS) and covariance weighted root-sum-of-squares (Cov-rSoS) at various reduction factors were compared between 8-channel CMDM array and 4-channel CM and DM array. The results of the study indicated the 8-channel CMDM array excelled the 4-channel CM and DM in SNR. The g-factor maps and artifact power were calculated to evaluate parallel imaging performance of the proposed 8-channel CMDM array. The artifact power of 8-channel CMDM array was reduced dramatically compared with the 4-channel CM and DM arrays demonstrating the parallel imaging feasibility of the CMDM array.Entities:
Keywords: Magnetic resonance image (MRI); common-mode differential-mode (CMDM); multichannel phased array; parallel MRI; signal-to-noise ratio (SNR)
Year: 2014 PMID: 24649433 PMCID: PMC3947983 DOI: 10.3978/j.issn.2223-4292.2014.02.05