Yong Pang1, Ernest W H Wong1, Baiying Yu1, Xiaoliang Zhang1.
Abstract
In this work, we propose and investigate a volume coil array design method using different types of birdcage coils for MR imaging. Unlike the conventional radiofrequency (RF) coil arrays of which the array elements are surface coils, the proposed volume coil array consists of a set of independent volume coils including a conventional birdcage coil, a transverse birdcage coil, and a helix birdcage coil. The magnetic fluxes of these three birdcage coils are intrinsically cancelled, yielding a highly decoupled volume coil array. In contrast to conventional non-array type volume coils, the volume coil array would be beneficial in improving MR signal-to-noise ratio (SNR) and also gain the capability of implementing parallel imaging. The volume coil array is evaluated at the ultrahigh field of 7T using FDTD numerical simulations, and the g-factor map at different acceleration rates was also calculated to investigate its parallel imaging performance.Keywords: Volume coil array; birdcage coil; g-factor map; helix birdcage coil; parallel imaging; signal-to-noise ratio (SNR); transverse birdcage coil
Year: 2014 PMID: 24649435 PMCID: PMC3947986 DOI: 10.3978/j.issn.2223-4292.2014.02.07