Scanning-slit and specular microscopic pachymetry in comparison with ultrasonic determination of corneal thickness.

L Módis1, A Langenbucher, B Seitz.   

Abstract

PURPOSE: To determine the central corneal thickness values in healthy eyes with the recently developed Orbscan scanning-slit system, contact and noncontact specular microscopic pachymetry and compare the results to conventional ultrasonic pachymetry.
METHODS: In the following sequence, Orbscan, Topcon SP-2000P noncontact specular microscope, AL-1000 ultrasound, and Tomey contact specular microscope were used to record thickness values. Thirty-four healthy right corneas of 34 healthy subjects were investigated.
RESULTS: Orbscan pachymetry correlated significantly with ultrasound (r = 0.64, p < 0.001), contact (r = 0.45, p < 0.001), and noncontact specular microscopy (r = 0.72, p < 0.001). Likewise, the Topcon SP-2000P noncontact specular microscopy pachymetry disclosed similar statistical results compared with ultrasound (r = 0.88, p < 0.001), and contact specular microscopy pachymetry (r = 0.76, p < 0.001). The mean central corneal thickness results were significantly higher ( p < or = 0.01) than ultrasonic values (580 +/- 43 microm) using the contact specular microscope (640 +/- 43 microm) or Orbscan system (602 +/- 59 microm) but were significantly lower ( p < 0.001) using the noncontact specular microscope (547 +/- 49 microm).
CONCLUSIONS: The results indicate that the devices tested cannot be simply used interchangeably. For long-term patient follow-up, one specific instrument is recommended. Recently developed pachymetry machines are especially helpful when additional corneal data such as thickness profile, elevation maps, anterior chamber depth, and endothelial morphology are required.

Entities:  

Mesh:

Year:  2001        PMID: 11588422     DOI: 10.1097/00003226-200110000-00008

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  13 in total

1.  Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

Authors:  K Kawana; T Tokunaga; K Miyata; F Okamoto; T Kiuchi; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-04       Impact factor: 4.638

2.  Corneal wound healing after photorefractive keratectomy: a 3-year confocal microscopy study.

Authors:  Jay C Erie
Journal:  Trans Am Ophthalmol Soc       Date:  2003

3.  Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.

Authors:  Mohammad Naderan; Saeed Shoar; Morteza Naderan; Mohammad Amin Kamaleddin; Mohammad Taher Rajabi
Journal:  Int J Ophthalmol       Date:  2015-04-18       Impact factor: 1.779

4.  Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-11-22       Impact factor: 3.117

5.  Repeatability of central corneal thickness measurements measured with the Topcon SP2000P specular microscope.

Authors:  Kelechi C Ogbuehi; Turki M Almubrad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

6.  Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry.

Authors:  Jeevan S Ladi; Nitant A Shah
Journal:  Indian J Ophthalmol       Date:  2010 Sep-Oct       Impact factor: 1.848

7.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24

8.  Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter.

Authors:  Turki M Almubrad; Uchechukwu L Osuagwu; Ibrahim Alabbadi; Kelechi C Ogbuehi
Journal:  Clin Ophthalmol       Date:  2011-06-24

9.  Central Corneal Thickness Measurement by Ultrasound versus Orbscan II.

Authors:  Amir Faramarzi; Hossein Ziai
Journal:  J Ophthalmic Vis Res       Date:  2008-04

10.  Changes in central corneal thickness and refractive error after thin-flap laser in situ keratomileusis in Chinese eyes.

Authors:  Ming-Hui Zhao; Qiang Wu; Li-Li Jia; Ping Hu
Journal:  BMC Ophthalmol       Date:  2015-07-29       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.