Repeatability and reproducibility of corneal thickness measurements by optical coherence tomography.

Sarah Muscat1, Nicola McKay, Stuart Parks, Ewan Kemp, David Keating.   

Abstract

PURPOSE: To assess the repeatability and interoperator and intersession reproducibility of central corneal thickness (CCT) measurements made by a commercially available optical coherence tomography (OCT) system.
METHODS: Intersession and interoperator reproducibility and repeatability were defined according to the guidelines of the British Standards Institution and examined in a control group of 14 normal subjects. An in-house computer program was used to evaluate central corneal thickness from these scans.
RESULTS: The coefficient of interoperator reproducibility was 0.18%, whereas that for intersession reproducibility was 1.11%. Wilcoxon analysis (5% level of significance) showed that there was no statistically significant difference between scans acquired during different sessions or by different operators. Coefficients of repeatability were all less than 3%. The average CCT was 526 +/- 28 microm (SD) and the range of normal CCT between 5th and 95th percentiles was 498 to 576 microm.
CONCLUSIONS: Although the commercially available OCT scanner was designed for retinal imaging, with a few minor modifications, the system may be used to image the anterior segment. Previous studies have shown that OCT measurements correlate well with those from conventional techniques, and it has the added advantage of being a noncontact technique. This study further demonstrates that the OCT measurements show a high degree of repeatability and reproducibility. Thus, OCT is emerging as a promising tool for evaluation of CCT in the clinical setting.

Entities:  

Mesh:

Year:  2002        PMID: 12036980

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  29 in total

1.  Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness.

Authors:  S Miglior; E Albe; M Guareschi; G Mandelli; S Gomarasca; N Orzalesi
Journal:  Br J Ophthalmol       Date:  2004-02       Impact factor: 4.638

2.  Comparisons of morphologic characteristics between thin-flap LASIK and SBK.

Authors:  Yi Sun; Ying-Ping Deng; Lin Wang; Yong-Zhi Huang; Le-Mei Qiu
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

3.  Optical coherence tomography of the Ex-PRESS miniature glaucoma implant.

Authors:  F D Verbraak; D M de Bruin; M Sulak; L A M S de Jong; M Aalders; D J Faber; T G van Leeuwen
Journal:  Lasers Med Sci       Date:  2005-04-28       Impact factor: 3.161

4.  Repeated measurements of dynamic tear distribution on the ocular surface after instillation of artificial tears.

Authors:  Jianhua Wang; James Aquavella; Jayachandra Palakuru; Suk Chung
Journal:  Invest Ophthalmol Vis Sci       Date:  2006-08       Impact factor: 4.799

5.  Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry.

Authors:  Hanna Y Kim; Donald L Budenz; Pak S Lee; William J Feuer; Keith Barton
Journal:  Am J Ophthalmol       Date:  2007-12-11       Impact factor: 5.258

6.  Comparison of central corneal thickness measurements obtained by community optometrists to those obtained in secondary care.

Authors:  Laura Hughes; Cheng Yi Loo; Brian Hogan; Huai Ling Tan; Ling Shan Tang; Andrew J Tatham
Journal:  Eye (Lond)       Date:  2018-07-23       Impact factor: 3.775

7.  Accutome PachPen handheld ultrasonic pachymeter: intraobserver repeatability and interobserver reproducibility by personnel of different training grades.

Authors:  Mohammadreza Peyman; Lai Yong Tai; Keat Ween Khaw; Choung Min Ng; Maung Maung Win; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-09-05       Impact factor: 2.031

8.  Corneal pachymetry mapping with high-speed optical coherence tomography.

Authors:  Yan Li; Raj Shekhar; David Huang
Journal:  Ophthalmology       Date:  2006-05       Impact factor: 12.079

9.  Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetry.

Authors:  Jeevan S Ladi; Nitant A Shah
Journal:  Indian J Ophthalmol       Date:  2010 Sep-Oct       Impact factor: 1.848

10.  Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.

Authors:  Dan Z Reinstein; Timothy J Archer; Marine Gobbe; Ronald H Silverman; D Jackson Coleman
Journal:  J Refract Surg       Date:  2009-11-16       Impact factor: 3.573

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.