Yoshiaki Morishima1,2, Koichi Chida2, Takayoshi Meguro3.
Abstract
Endoscopic retrograde cholangiopancreatography (ERCP) is often complex and involves long fluoroscopic times, with significant radiation exposure to medical staff. We investigated protective effects of an additional attached lead shielding device. The lead shielding device covered with the X-ray tube table (0.125 mm lead equivalent) during ERCP procedures. Fluoroscopy scatter radiation, with or without the lead shielding device, was measured using an acrylic phantom and a radiation survey meter. Measurements (25 points) were made at 50 cm intervals, at both 90 and 150 cm above the floor. We created radiation maps, with and without the additional lead shielding device. Moreover, we monitored annual staff exposure to radiation, before and after inclusion of the shielding device. Without additional shielding, exposure doses at the physician's position, 90 and 150 cm above the floor, were 1940 and 4040 (μSv/h) respectively. In contrast, with the shielding device, corresponding exposures were 270 and 450 (μSv/h) at 90 and 150 cm, respectively. Scattered radiation was decreased by 86.1% at 90 cm or 88.9% at 150 cm. However, with additional lead shielding in the middle, rather than hung over the operating table, scattered radiation was decreased by only ~10%. The staff's annual dose equivalents (DEs) were 12.2-29.8 mSv/year without and 3.8-8.4 mSv/year with lead shielding. With lead shielding, dose equivalent values for the staff were decreased by 41.0-76.5%. Thus, with additional lead shielding, properly used, scattered radiation would be decreased by ~90%, thus decreasing exposure doses to medical staff during ERCPs.Entities:
Mesh:
Year: 2018 PMID: 29409055 PMCID: PMC5951079 DOI: 10.1093/jrr/rrx039
Source DB: PubMed Journal: J Radiat Res ISSN: 0449-3060 Impact factor: 2.724
Figure 1.The measurement arrangement plan of a phantom study. (Measurements were made at the positions of the black dots (each separated from the other by 50 cm.)) Positions of staff relative to the patient and X-ray tube during the ERCP procedure: P = Physician (endoscopist), Q = 1st nurse, R = 2nd nurse, S = assistant physician.
Figure 2.Without shielding (A) and with shielding to the operating table (B). The front and back of the additional lead shielding are 85 × 66 cm and the two sides are each 85 × 50 cm.The upper part of the made of mesh (without the lead shield).
Scattering doses (μSv/h) at the physicians’ (P, S) and nurses’ (Q, R) positions, and the maximum points, without (−) and with (+) lead shielding
| 150 cm | 90 cm | |||
|---|---|---|---|---|
| (−) | (+) | (−) | (+) | |
| P | 4040 | 450 | 1940 | 270 |
| Q | 890 | 135 | 495 | 185 |
| R | 2270 | 140 | 1170 | 134 |
| S | 1400 | 200 | 784 | 125 |
| maximum point | 4820 | 525 | 2310 | 342 |
P = physician, Q = 1st nurse, R = 2nd nurse, S = assistant physician.
Figure 3.(A) Radiation map (without shielding 90cm in height above the floor). (B) Radiation map (with shielding 90 cm in height above the floor).
Figure 4.(A) Radiation map (without shielding 150 cm in height above the floor). (B) Radiation map (with shielding 150 cm in height above the floor).
Figure 5.Additional lead shielding to the operating table not hung down (in the middle).
Scattering radiation at the physician’s position, without shielding or with the additional lead shielding device either installed properly (with) or not hung down to the operating table (in the middle)
| Height above the floor | 150 cm | 90 cm | ||
|---|---|---|---|---|
| μSv/h | Reducing rate (%a) | μSv/h | Reducing rate (%a) | |
| Without | 4040 | 1940 | ||
| With | 450 | 88.9 | 270 | 86.1 |
| In the middle | 3700 | 8.4 | 1720 | 11.3 |
aScatter radiation dose reduction rate (%) = [(without − (with or middle))/without] × 100.
Physician exposure doses (mSv/year; per 1 year), before and after installation of the additional lead shielding device
| Before (July 2012–June 2013) | After (July 2013–June 2014) | ||
|---|---|---|---|
| Physician | ED | 6.3 | 4.5 |
| DE | 12.9 | 4.6 | |
| Ns1 | ED | 8.5 | 1.0 |
| DE | 12.2 | 7.2 | |
| Ns2 | ED | 4.4 | 1.5 |
| DE | 29.8 | 7.0 | |
| Ns3 | ED | 4.9 | 0.8 |
| DE | 17.0 | 8.4 | |
| Ns4 | ED | 1.6 | 0.4 |
| DE | 12.9 | 3.8 |