The effectiveness of additional lead-shielding drape and low pulse rate fluoroscopy in protecting staff from scatter radiation during cardiac resynchronization therapy (CRT).

Yoshiaki Morishima1,2, Koichi Chida3, Yoshiaki Katahira4.   

Abstract

PURPOSE: Cardiac resynchronization therapy (CRT) often requires a long fluoroscopic time and protection from scatter radiation. This study reports on scatter radiation levels during CRT, with and without additional shielding, and using standard or low pulse rate fluoroscopy.
MATERIALS AND METHODS: Additional lead-shielding drape (0.35-mm lead equivalent) was used on the left side of the table and pulsed fluoroscopy was performed at rates of 10 pulses/s (usual rate) and 7.5 pulses/s (low pulse rate). Fluoroscopy scatter radiation was measured for both pulse rates using an acrylic phantom with a radiation survey meter, both with and without the additional lead-shielding drape.
RESULTS: With the additional lead-shielding drape, the fluoroscopy scatter radiation was reduced by 74.3% at 10 pulses/s and 78.6% at 7.5 pulses/s. If the fluoroscopy was changed from 10 pulses/s to 7.5 pulses/s, the scattered radiation at the primary physician's position was reduced by 24.0%. The combined use of additional shielding drape and low pulse rate fluoroscopy reduced scatter radiation by over 80%.
CONCLUSION: Additional lead-shielding drape and low pulse rate fluoroscopy are effective in reducing the scattered radiation dose to physicians and nurses during CRT.

Entities:  

Keywords:  Additional lead-shielding drape; CRT procedure; Low pulse rate fluoroscopy; Radiation protection; Scatter radiation

Mesh:

Year:  2018        PMID: 30324249     DOI: 10.1007/s11604-018-0783-7

Source DB:  PubMed          Journal:  Jpn J Radiol        ISSN: 1867-1071            Impact factor:   2.374


  25 in total

1.  Radiation exposure of patient and physician during implantation and upgrade of cardiac resynchronization devices.

Authors:  Christian Butter; Thomas Schau; Juergen Meyhoefer; Klaus Neumann; Hans H Minden; Joerg Engelhardt
Journal:  Pacing Clin Electrophysiol       Date:  2010-05-04       Impact factor: 1.976

2.  Radiation exposure of operator performing interventional procedures using a flat panel angiography system: evaluation with photoluminescence glass dosimeters.

Authors:  Yoshinori Funama; Nozomu Nagasue; Kazuo Awai; Ichiro Sakamoto; Kiyotaka Kakei; Masamichi Shimamura; Yasuyuki Yamashita; Masataka Uetani
Journal:  Jpn J Radiol       Date:  2010-07-27       Impact factor: 2.374

3.  Reducing operator radiation exposure during cardiac resynchronization therapy.

Authors:  Marco Brambilla; Eraldo Occhetta; Martina Ronconi; Laura Plebani; Alessandro Carriero; Paolo Marino
Journal:  Europace       Date:  2010-10-06       Impact factor: 5.214

4.  Occupational radiation doses to the extremities and the eyes in interventional radiology and cardiology procedures.

Authors:  E P Efstathopoulos; I Pantos; M Andreou; A Gkatzis; E Carinou; C Koukorava; N L Kelekis; E Brountzos
Journal:  Br J Radiol       Date:  2011-01       Impact factor: 3.039

Review 5.  Radiation dose and radiation protection for patients and physicians during interventional procedure.

Authors:  Koichi Chida; Mamoru Kato; Yutaka Kagaya; Masayuki Zuguchi; Haruo Saito; Tadashi Ishibashi; Shoki Takahashi; Shogo Yamada; Yoshihiro Takai
Journal:  J Radiat Res       Date:  2010       Impact factor: 2.724

6.  Cardiac resynchronization therapy: a meta-analysis of randomized controlled trials.

Authors:  George Wells; Ratika Parkash; Jeffrey S Healey; Mario Talajic; J Malcolm Arnold; Shannon Sullivan; Joan Peterson; Elizabeth Yetisir; Patricia Theoret-Patrick; Marilynn Luce; Anthony S L Tang
Journal:  CMAJ       Date:  2011-01-31       Impact factor: 8.262

7.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

8.  Evaluation of additional lead shielding in protecting the physician from radiation during cardiac interventional procedures.

Authors:  Koichi Chida; Yoshiaki Morishima; Yoshiaki Katahira; Hiroo Chiba; Masayuki Zuguchi
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2005-12-20

9.  Radiation dose of interventional radiology system using a flat-panel detector.

Authors:  Koichi Chida; Yohei Inaba; Haruo Saito; Tadashi Ishibashi; Shoki Takahashi; Masahiro Kohzuki; Masayuki Zuguchi
Journal:  AJR Am J Roentgenol       Date:  2009-12       Impact factor: 3.959

Review 10.  Avoidance of radiation injuries from medical interventional procedures.

Authors:  J Valentin
Journal:  Ann ICRP       Date:  2000
View more
  3 in total

1.  Evaluation of novel X-ray protective eyewear in reducing the eye dose to interventional radiology physicians.

Authors:  Mime Endo; Yoshihiro Haga; Masahiro Sota; Akiko Tanaka; Kazuki Otomo; Yuuki Murabayashi; Mitsuya Abe; Yuji Kaga; Yohei Inaba; Msatoshi Suzuki; Taiichiro Meguro; Koichi Chida
Journal:  J Radiat Res       Date:  2021-05-12       Impact factor: 2.724

2.  New Radioprotective Device that can be Used for Fluoroscopic Exam: Possibility to Contribute to Staff Exposure Protection During VFSS.

Authors:  Yoshiaki Morishima; Koichi Chida; Osamu Ito
Journal:  Dysphagia       Date:  2022-02-15       Impact factor: 3.438

3.  Development of a New Radiation Shield for the Face and Neck of IVR Physicians.

Authors:  Toshimitsu Sato; Yoichi Eguchi; Chika Yamazaki; Takanobu Hino; Toshikazu Saida; Koichi Chida
Journal:  Bioengineering (Basel)       Date:  2022-07-29
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.