AIMS: Cementless acetabular components rely on press-fit fixation for initial stability. In certain cases, initial stability is more difficult to obtain (such as during revision). No current study evaluates how a surgeon's impaction technique (mallet mass, mallet velocity, and number of strikes) may affect component fixation. This study seeks to answer the following research questions: 1) how does impaction technique affect a) bone strain generation and deterioration (and hence implant stability) and b) seating in different density bones?; and 2) can an impaction technique be recommended to minimize risk of implant loosening while ensuring seating of the acetabular component? METHODS: A custom drop tower was used to simulate surgical strikes seating acetabular components into synthetic bone. Strike velocity and drop mass were varied. Synthetic bone strain was measured using strain gauges and stability was assessed via push-out tests. Polar gap was measured using optical trackers. RESULTS: A phenomenon of strain deterioration was identified if an excessive number of strikes was used to seat a component. This effect was most pronounced in low-density bone at high strike velocities. Polar gap was reduced with increasing strike mass and velocity. CONCLUSION: A high mallet mass with low strike velocity resulted in satisfactory implant stability and polar gap, while minimizing the risk of losing stability due to over-striking. Extreme caution not to over-strike must be exercised when using high velocity strikes in low-density bone for any mallet mass.Cite this article: Bone Joint Res 2020;9(7):386-393.
AIMS: Cementless acetabular components rely on press-fit fixation for initial stability. In certain cases, initial stability is more difficult to obtain (such as during revision). No current study evaluates how a surgeon's impaction technique (mallet mass, mallet velocity, and number of strikes) may affect component fixation. This study seeks to answer the following research questions: 1) how does impaction technique affect a) bone strain generation and deterioration (and hence implant stability) and b) seating in different density bones?; and 2) can an impaction technique be recommended to minimize risk of implant loosening while ensuring seating of the acetabular component? METHODS: A custom drop tower was used to simulate surgical strikes seating acetabular components into synthetic bone. Strike velocity and drop mass were varied. Synthetic bone strain was measured using strain gauges and stability was assessed via push-out tests. Polar gap was measured using optical trackers. RESULTS: A phenomenon of strain deterioration was identified if an excessive number of strikes was used to seat a component. This effect was most pronounced in low-density bone at high strike velocities. Polar gap was reduced with increasing strike mass and velocity. CONCLUSION: A high mallet mass with low strike velocity resulted in satisfactory implant stability and polar gap, while minimizing the risk of losing stability due to over-striking. Extreme caution not to over-strike must be exercised when using high velocity strikes in low-density bone for any mallet mass.Cite this article: Bone Joint Res 2020;9(7):386-393.
This in vitro study investigated the effect of impaction technique (including mallet mass, mallet velocity, and number of strikes) upon acetabular component stability.This study identified a trend of stability deterioration with excessive impaction strikes.The use of a high mass mallet with low strike velocity may allow adequate fixation in low-density bone while minimizing the possibility of stability deterioration.The effects of each element of impaction technique (mallet mass, strike velocity, and number of strikes) upon clinically relevant outcomes have been reported. To our knowledge, no comparable study has been published.This study used a synthetic bone model. The magnitudes of the effects reported may vary in live bone tissue. In addition, only time-zero effects can be reported.
Introduction
The majority of hip implants used in total hip arthroplasty (THA) in Europe, Australia, and the USA are now cementless[1-4] and show reduced mortality risk compared to cemented implants.[5,6] Cementless procedures have been shown to have favourable outcomes even in revision cases, despite decreased bone stock, and are often recommended over cemented alternatives.[7-9] Cementless implants rely on a mechanical press-fit to achieve initial stability. If successful, this can result in lifelong secure implant fixation.[10]Press-fit stability is achieved by impacting an implant into an undersized cavity in the bone and generating hoop strain, creating friction at the implant/bone interface. With adequate strain the implant is stable and long-term bone ingrowth is promoted.[11,12] However, if there is too little strain the implant may be unstable, and small movements can occur during loading, preventing bony ingrowth.[13,14] Implant stability is particularly important during revision cases, where bone stock and quality may be poorer. In these cases secure implant stability is more difficult to achieve,[9,15] and the likelihood of fracture may increase.[16,17] Early instability and loosening has been reported as the main cause of cementless implant failure following revision.[7,18,19] In order to maximize the benefits of using cementless implants, particularly in more difficult cases, stability must be achieved.[20]Much previous research has focused on how initial stability is influenced by implant/bony factors: component design,[21,22] implant surface,[23] component stiffness,[24] bone density,[25] and bone/component interference.[26-30] However there is a paucity of literature discussing surgical technique in seating cementless implants. Impaction technique (mallet mass, mallet velocity, and number of impaction strikes) played a crucial role in seating over 34,000 cementless components in the UK in 2017. Previous work has investigated the role of impaction energy or implantation force, both of which depend upon mallet mass and velocity, implant stability, and seating.[31] However the effect of number of impaction strikes remains unclear. As it can be difficult for the surgeon to accurately infer when a component is seated,[32] the effect of under- or over-impacting an acetabular component requires investigation.This study uses an in vitro model to answer two questions: 1) how does impaction technique affect a) bone strain generation and deterioration (and hence implant stability) and b) seating in different density bones?; and 2) can an impaction technique be recommended to minimize risk of implant loosening while ensuring seating of the acetabular component?
Methods
Setup
An in vitro drop tower was used to impact acetabular components into synthetic bone (Figure 1). By varying the height of the drop, the rig is able to mimic surgical mallet impaction strikes in a repeatable manner. In addition the drop rig mimics in vivo soft tissue boundary conditions. Briefly, the synthetic bone is mounted to a mass/spring/dashpot that provides the same displacement and acceleration rate under impact as the human acetabulum during surgery. This ensures the same damping of soft tissue seen in vivo is represented in vitro. The details of the test setup and human cadaver validation have been described previously.[31,33]
Fig. 1
Experimental impact setup.
Experimental impact setup.Expanded polyurethane foam (Model #1522 to 02 and #1522 to 04; Sawbones, Pacific Laboratories, Malmö, Sweden) was used as a repeatable synthetic bone substitute for the acetabulum: the high-density (30 PCF) foam replicated good-quality bone with a strong cortical rim while the low-density (15 PCF) foam replicated poorer quality bone, such as the exposed cancellous, sub-chondral bone seen in revision cases.[34-38] The bones were computer numerical control (CNC) milled to create a 53 mm diameter acetabular cavity, with a previously validated anatomically representative geometry, including two ‘notches’. These were used to simulate the pinching effect of the ischeal and ilial columns, applying similar forces to acetabular components as in vivo testing.[39-42] The hemisphere cavity was centred 1 mm below the surface of the block.The acetabular components used were designed to have a rim-to-rim stiffness similar to commercially available devices,[1,43] manufactured with a 3 mm shell thickness in Ti6Al4V alloy. Each component was manufactured with an extruded spigot, projected into the component from the pole, allowing components to be attached to a straight introducer for impaction. A finite element analysis was used to ensure the resulting components fell within the range of stiffness reported for commercial acetabular components.[38,43] A rough outer surface (Rz > 500 μm) was applied to the components, representing a typical implant fixation surface.[44] Then, 55 mm and 54 mm component sizes were chosen for the 15 PCF and 30 PCF models resulting in 2 mm and 1 mm diametric interference respectively, which represents typical clinical practice.[11,27,45,46]
Testing
Data were acquired for three strike velocities and three strike masses. Low, medium, and high velocities of 1.5 m/s, 2.75 m/s, and 4.0 m/s respectively were chosen based upon existing literature and previous testing.[33,47-49] A video demonstration intended to aid surgeon replication of these velocities is available in the supplementary material (Supplementary Video i). Low, medium, and high strike masses of 0.6 kg, 1.2 kg, and 1.8 kg were implemented by changing the mass of the drop platen (Figure 1) and were chosen to cover a range of commonly used surgical mallets. Five repeats were completed for each of the nine mass/velocity combinations, in each of the two densities of synthetic bone. Strain and displacement data were recorded for ten strikes for each setup. In total 180 tests were completed.Click here for additional data file.
Implant stability: strain measurement
Strain gauges (350 ohm, SGT-3F/350TY11; Omega, Manchester, UK) were used to measure hoop strain on the surface of the bone. Two gauges per bone were glued radially around the cavity periphery, 2 mm from the cavity edge (Figure 2). This location has previously been shown to experience the highest strain.[21,31] Strain gauge data were acquired for one second following each strike using a USB datalogger (NI9327; National Instruments, Budapest, Hungary) in quarter bridge configuration. Raw data were captured using Labview (National Instruments, 2014) and analyzed using MATLAB (Mathworks, Natick, Massachusetts, USA).
Fig. 2
Synthetic bone cavity geometry and strain gauge placement.
Synthetic bone cavity geometry and strain gauge placement.
Implant stability: strain deterioration
Pilot testing indicated a characteristic pattern to the strain with an increasing number of strikes (Figure 3): initially strain increased strike by strike (Figure 3a), which represents an increase in the stability of the implant.[31] This was followed by an inflexion point (peak strain) and a subsequent deterioration of strain with each strike (Figure 3b). This deterioration in strain with subsequent strikes was quantified and analyzed as it poses a clinical risk: over-impacting the component may reduce implant stability. A metric of strain reduction, hereby known as strain deterioration, is defined by the following equation, with terms defined graphically in Figure 3.
Fig. 3
Representative bone microstrain measurements during seating. An increase of strain is observed (A), followed by an inflexion point and subsequent deterioration (B).
Representative bone microstrain measurements during seating. An increase of strain is observed (A), followed by an inflexion point and subsequent deterioration (B).A strain deterioration of zero would represent conditions where an excessive number of strikes beyond some optimum does not reduce bone strain at all. A larger strain deterioration magnitude represents conditions where excessive strikes cause a larger reduction in bone strain, and therefore reduction in component stability.
Implant stability: pushout
Pushout fixation was measured as an adjunct to bone strain as both are shown to relate to stability and correlate well.[31] Pushout fixation was measured for two conditions. The first was after ten strikes for all mass, velocity, and bone density setups. The second was for optimum number of strikes, when the implant could be considered ‘seated’. This was when the implant had progressed by no more than 0.1 mm during the previous strike (see ‘Seating: polar gap’ for measurement method). A material testing machine (Instron model 5565; Instron, High Wycombe, UK) pushed components out at 0.5 mm per minute using a ø6 mm diameter steel rod through a 7 mm diameter entry hole milled into the synthetic bone at the pole of the component. Peak load was recorded. The pushout after ten strikes was always less than the ‘seated’ number of strikes, due to the strain deterioration illustrated in Figure 3. A pushout deterioration was then defined in the following equation:
Seating: polar gap
Two sets of infrared reflective markers were attached to the component introducer and impact platen (Figure 1), allowing their relative positions to be measured by an optical tracker (fusionTrack 500; Atracsys, Puidoux, Switzerland). Following ten strike seating the final polar gap of each specimen was measured using a depth gauge micrometer. Superposition of final polar gap and relative component movement during seating was used to calculate polar gap following each strike. Using these data the ‘optimum number of strikes’ for each impaction setup could be recorded.
Statistical analysis
Low-density bone data were tested for normality using a Shapiro-Wilk test in SPSS v24 (IBM, Armonk, New York, USA). Data were then analyzed with a two-way repeated measures analysis of variance with strain deterioration as the dependent variable and mallet mass (low, medium, high) and strike velocity (low, medium, high) as independent variables. The significance level was set to p < 0.05. Post-hoc paired t-test with Bonferroni correction was applied when differences across tests were found. Adjusted p-values, multiplied by the appropriate Bonferroni correction factor in SPSS, have been reported rather than reducing the significance level. A linear regression analysis was also performed to determine if a relationship exists between strain and pushout deterioration. Polar gap data were analyzed with paired t-tests. These analyses were then repeated for the high-density bone data.
Results
For both bone densities, at all velocities and all mass strikes, implant impaction was characterized by a period of increasing bone strain, followed by an inflection point and then by strain deterioration (Figure 4). This severity of the peak and strain deterioration is observed to varying degrees depending upon the mass/velocity strike combination. In the most extreme case with low-density bone, high strike mass, and high strike velocity, the peak strain was generated after one strike; the second strike then halved the strain, and the third strike reduced the strain to zero (Figure 4c).
Fig. 4
Mean microstrain measured (with SD) post-strike for low-density bone at a) low, b) medium, and c) high velocities. Strains are characterized by increasing strain, an inflexion point, and strain reduction.
Mean microstrain measured (with SD) post-strike for low-density bone at a) low, b) medium, and c) high velocities. Strains are characterized by increasing strain, an inflexion point, and strain reduction.For low-density bone, the effects of mallet mass upon strain deterioration depended upon the strike velocity (interaction p = 0.007) (Figure 5). At low velocity, all the strain deterioration values were low (-8 to -17) and the effect of strike mass was low (maximum difference -9; p = 0.070). Conversely, at high strike velocities the strain deterioration was much higher (-42 to -78) and the effects of strike mass were much greater (maximum difference -36; p < 0.005). This indicates that low velocities were less sensitive to over-impaction and that increasing strike mass at low velocity in low-density bone does not increase the strain deterioration if too many strikes are used. For high-density bone strain deterioration magnitudes were low (-2 to -19; Supplementary Figure a), indicating that strain deterioration poses a greater clinical risk for low-density bone.
Fig. 5
Mean strain deterioration (with SD) for each impaction technique. Strain deterioration was greatest for lower-density bone, and for medium- or high-velocity strikes.
Mean strain deterioration (with SD) for each impaction technique. Strain deterioration was greatest for lower-density bone, and for medium- or high-velocity strikes.The number of strikes until the ‘optimally seated’ condition and pushout values are provided in Table I. For the group that were pushout tested when ‘optimally seated’, fixation increased with both strike mass and velocity. However, in the ten-strike group an increase in fixation with increasing mallet mass was only observed in the low velocity group. In the medium and high velocity group no increase in fixation was observed with mallet mass (Table I). Pushout deterioration is plotted as a function of strain deterioration (Figure 6; R2 = 0.83, p < 0.05). These data confirmed that strain deterioration was an appropriate measure for risk of component loosening, as fixation strength reduced linearly with strain deterioration.
Table I.
Number of strikes required for ‘seated’ condition and mean pushout values for each strike combination.
Strike velocity
Low
Medium
High
Strike mass
Low
Med
High
Low
Med
High
Low
Med
High
Number of strikes to seat
10*
10*
6
7
4
2
3
2
2
Mean pushout value (seated), N (SD)
123 (12)
248 (16)
442 (38)
289 (46)
356 (77)
496 (22)
368 (10)
442 (79)
630 (65)
Mean pushout value (10 strikes), N (SD)
123 (12)
248 (16)
362 (18)
172 (19)
127 (48)
90 (18)
45 (14)
60 (14)
49 (6)
< 0.1 mm displacement at tenth strike.
Fig. 6
Relationship between strain and pushout deterioration. Larger magnitude strain deterioration correlates with a reduction in pushout. Strain deterioration should be avoided to ensure optimum fixation.
Number of strikes required for ‘seated’ condition and mean pushout values for each strike combination.< 0.1 mm displacement at tenth strike.Relationship between strain and pushout deterioration. Larger magnitude strain deterioration correlates with a reduction in pushout. Strain deterioration should be avoided to ensure optimum fixation.Polar gap reduced with both increasing mallet mass and velocity in both low- and high-density bone in all but the high velocity, high density group (Figure 7). A difference in polar gap between low and high mallet mass was detected (0.335 mm; p = 0.014) but not between low/medium and medium/high masses. Combining the findings of Figures 5 and 7 across the low- and high-density bone materials, the only mass/velocity combination that provided a low strain deterioration and full implant seating (polar gap ≤ 1 mm) was the high mallet mass at low velocity.
Fig. 7
Mean polar gaps at seating (with SD) for each impaction technique. Polar gap is reduced with both increasing impact mass and velocity.
Mean polar gaps at seating (with SD) for each impaction technique. Polar gap is reduced with both increasing impact mass and velocity.
Discussion
The most important finding of this study was that the strain in bone deteriorated after a certain number of impaction strikes, and that impaction technique had a large effect on this deterioration. A deterioration in strain was measured in all cases, but was most severe in the case of lower density bone (representing revision or poor-quality bone cases). This deterioration correlated well with a reduction in implant stability and so must be avoided. The strike velocity, which can be interpreted as how hard the surgeon hits the introducer, was a key risk factor for this deterioration – at high velocity even the lightest 0.6 kg mallet caused a severe deterioration in strain with excessive impaction strikes. At low velocity, for all mallet masses the deterioration in strain still occurred but was far less severe than at medium or high velocity. Therefore, a lower strike velocity may be preferable. It should also be considered that polar gaps of more than 2 mm have been linked to an increase in early component migration,[50] while gaps of less than 2 mm have been deemed clinically acceptable[46] and the gaps of 1 mm or less have been observed to be filled without intervention.[51] A high mallet mass, coupled with a low strike velocity, minimizes strain deterioration while achieving < 1 mm polar gap and 66% of stability attained by the technique with the highest risk of strain deterioration.This study demonstrates that high velocity mallet strikes achieve good implant stability, but only if the correct number of strikes are used. However, it has previously been shown that it is difficult for surgeons to accurately pinpoint when an implant is most effectively seated (i.e. the inflection points in Figure 3).[32] Therefore if high-velocity strikes are used to seat components in low-density bone, extreme care should be taken to avoid an excessive number of strikes.A synthetic bone model was used for this study. It is acknowledged that an isotropic material does not represent the anatomical complexity of the pelvis. However, cadaveric bone strength, stiffness, density, acetabular socket size, and rim anatomy all vary with specimen. A cadaveric repeated measures study design is also not possible due to the damage seating a component does to a specimen, preventing repeat testing. A synthetic model was chosen over cadaveric tissues to allow isolation of several variables and to perform meaningful analysis on the results. Every effort has been made to best represent the clinical scenario. Previous literature demonstrates the value of these models to draw important conclusions.[41,44,52] Synthetic bone densities were carefully chosen using existing literature in order to most closely represent high-density (mostly cortical in primary THA) and low-density (more sub-chondral/defected) bone.[20,34-36,38] While the model does not mimic the overall anatomy of the pelvis, the ‘notch’ model has been rigorously validated against cadaveric tissue in literature, providing mechanical properties,[53,54] component deformation,[39] component frictional resistance,[20] and stability[55] within the variability of mechanical samples, even when damaged.[56] Therefore the key ‘pinching’ mechanism of the acetabulum is represented.[39] The mechanism of the most important finding of this study (a deterioration of strain with excessive strikes) is likely due to the elastic nature of bone and the compliance of the hip during impact – two properties represented and validated for the model used.[33] A second limitation is that pushout loading is not a physiological method of loading. However, it does test peripheral fixation effectively, and has been shown to produce similar results to other fixation testing methods.[21,35,57]It is proposed that two previously reported phenomena contribute to the loss of stability measured when using too many high-velocity strikes. The first is damage to the bone during each strike.[58] As the component is seated the bone surface layer is overly compacted/abraded by the component surface, reducing effective interference fit between component and bone. The increased yield strength of the higher-density bone, resisting damage, may contribute to its increased tolerance of high energy impacts (before strain deterioration occurs). Similarly, in low-impact energy scenarios (such as in the low velocity group) bone may not be sufficiently damaged to contribute to deterioration. Higher mallet mass may be preferable as mallet momentum is increased (aiding implant seating) without largely increasing energy (therefore limiting bone damage). In addition to bone damage, a ‘bounce back’ of an acetabular component out of the bone cavity has been identified in a previous study that employs dynamic testing techniques.[39] Mid-strike, elastic energy is stored in the acetabular cavity and surrounding soft tissues. As the force of the hammer strike is released the component is subject to an ‘extraction’ force and can displace out of the bone cavity. Hothi et al[48] observed this phenomenon during higher-velocity strikes, but only in dynamic models. It has been previously shown that the acetabulum undergoes significant acceleration during impact,[33] which could produce an extraction force when experienced by introducer/component couplings with appreciable mass. Supplementary Video ii highlights the oscillation of the component/introducer construct within the acetabular cavity. Such mechanisms of elastic strain and acetabulum acceleration are known to occur in the in vivo environment, and therefore the mechanism of strain and stability deterioration demonstrated in this study warrants further investigation.In conclusion, inadequate implant stability in revision THA or primary cases with poor bone stock remains a challenge for orthopaedic surgeons, with second revisions often required if adequate stability is not achieved. This study assesses how surgical technique may improve fixation. The in vitro model presented has indicated that surgeons should take great care if high-velocity strikes are used (even with a low mallet mass) to seat components in low density, as stability may be reduced if excessive impact strikes are used. The use of low strike velocity (1.5 m/s) and high mallet mass (1.8 kg) may mitigate a risk of loss of implant stability when impacting components into low-density bone, while ensuring the implant is seated well enough to promote effective bone ingrowth.