Cementless hemispheric porous-coated sockets implanted with press-fit technique without screws: average ten-year follow-up.

Pacharapol Udomkiat1, Lawrence D Dorr, Zhinian Wan.   

Abstract

BACKGROUND: Press-fit implantation of a porous-coated hemispheric acetabular component without screws is an option for primary total hip replacement. The purpose of the present study was to evaluate the results of this technique after an average duration of follow-up of ten years to determine if there was any loss of fixation or increase in osteolysis over time.
METHODS: From June 1988 to November 1990, 132 primary total hip replacements were performed with a porous-coated socket that was implanted with use of a press-fit technique. Twenty-two hips were excluded because the patient had died or had been lost to follow-up, leaving 110 hips (103 patients) available for inclusion in the study after an average duration of follow-up of 10.2 +/- 1.0 years. The average age of the patients at the time of operation was 60.7 years (range, 23.7 to 86.2 years). Radiographs were evaluated with regard to initial gaps, radiolucent lines, migration, polyethylene wear, and osteolysis. Kaplan-Meier survivorship analysis was performed to calculate the rate of survival of the acetabular component.
RESULTS: One hip (0.9%) had revision of the socket because of aseptic loosening, and four hips (4%) had revision of a stable socket. With the numbers available, the presence of gaps on the initial postoperative radiographs was not associated with the occurrence of radiolucent lines (p = 0.039). Pelvic osteolysis was seen in four hips, with an average time to radiographic appearance of six years. Increased wear was directly related to an abduction angle of >40 degrees. The twelve-year survival rate was 99.1% with revision because of failure of fixation of the metal shell as the end point, 95.3% with revision for any reason as the end point, and 79.6% with exchange of the liner as the end point.
CONCLUSIONS: The fixation of this press-fit socket did not deteriorate over time and was associated with a low rate of osteolysis. The most common reasons for reoperation were wear and dissociation of the polyethylene insert.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12107321     DOI: 10.2106/00004623-200207000-00016

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  29 in total

1.  The incidence of acetabular osteolysis in young patients with conventional versus highly crosslinked polyethylene.

Authors:  Nathan A Mall; Ryan M Nunley; Jin Jun Zhu; William J Maloney; Robert L Barrack; John C Clohisy
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

2.  Cementless acetabular component inserted without screws - the effect of immediate weight-bearing.

Authors:  F Y Ng; Y Zhu; K Y Chiu
Journal:  Int Orthop       Date:  2006-08-19       Impact factor: 3.075

3.  [Minimally invasive posterior approach for total hip arthroplasty].

Authors:  B Fink; A Mittelstaedt
Journal:  Orthopade       Date:  2012-05       Impact factor: 1.087

4.  Press-fit cementless acetabular fixation with and without screws.

Authors:  Sheng-Hui Ni; Lei Guo; Tian-Long Jiang; Jie Zhao; Yan-Gang Zhao
Journal:  Int Orthop       Date:  2013-08-28       Impact factor: 3.075

5.  Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins.

Authors:  Wolfgang O Huber; Philip C Noble
Journal:  Int Orthop       Date:  2013-11-22       Impact factor: 3.075

6.  [Five-year survival rate of the Allofit titanium press-fit cup].

Authors:  K Schroeder; A Moehlenbruch; M Zimmermann-Stenzel; D Parsch
Journal:  Orthopade       Date:  2010-01       Impact factor: 1.087

7.  Correlation of computed tomography with histology in the assessment of periprosthetic defect healing.

Authors:  Stephen D Cook; Laura P Patron; Samantha L Salkeld; Kirk E Smith; Bruce Whiting; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2009-12       Impact factor: 4.176

8.  Basic science considerations in primary total hip replacement arthroplasty.

Authors:  Saqeb B Mirza; Douglas G Dunlop; Sukhmeet S Panesar; Syed G Naqvi; Shafat Gangoo; Saif Salih
Journal:  Open Orthop J       Date:  2010-05-11

9.  No differences in outcomes between cemented and uncemented acetabular components after 12-14 years: results from a randomized controlled trial comparing Duraloc with Charnley cups.

Authors:  Kristian Bjørgul; Wendy M Novicoff; S T Andersen; K Brevig; F Thu; M Wiig; O Ahlund
Journal:  J Orthop Traumatol       Date:  2010-03-03

10.  Long-term results of the threaded Weill cup in primary total hip arthroplasty: a 15-20-year follow-up study.

Authors:  Michael Clarius; Alexander W Jung; Patric Raiss; Marcus R Streit; Christian Merle; Peter R Aldinger
Journal:  Int Orthop       Date:  2009-07-24       Impact factor: 3.075

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.