David C Markel1, Nivedita Hora, Michele Grimm.
Abstract
Cementless acetabular components require good initial fixation to allow bony in-growth. The initial press-fit stability, important for designs that do not rely on supplemental fixation, was examined for three designs with different porous coating systems: beads, fiber mesh, and plasma spray. The ability to withstand tangential loads was determined (maximum rim loads: 122-1730 N). The plasma-sprayed acetabular cups withstood the greatest tangential load within the specified range of motion of 150 microm (P<0.01). Differences in surface preparation of titanium acetabular cups may significantly affect the initial stability of the implants in rim loading.Mesh:
Substances:
Year: 2002 PMID: 12078880 PMCID: PMC3620875 DOI: 10.1007/s00264-001-0314-4
Source DB: PubMed Journal: Int Orthop ISSN: 0341-2695 Impact factor: 3.075