| Literature DB >> 31321615 |
Arnold B Vardiman1, David J Wallace2, Neil R Crawford3, Jessica R Riggleman3, Leigh A Ahrendtsen3, Charles G Ledonio3.
Abstract
In the emerging field of robot-assisted spine surgery, the radiographic evaluation of pedicle screw accuracy in clinical application is an area of high interest. This study describes the pedicle screw accuracy of the first 56 consecutive cases in which navigated robotic assistance was used in a private practice clinical setting. A retrospective, Institutional Review Board-exempt review of the first 56 navigated robot-assisted spine surgery cases was performed. Pedicle screw malposition, reposition, and return to operating room (OR) rates were collected. A CT-based Gertzbein and Robbins system (GRS) was used to classify pedicle screw accuracy. In the first 56 robotic cases, 356 total pedicle screws were placed. Eight screws were placed without the robot due to surgeon discretion. Of the 348 pedicle screws inserted by navigated robotic guidance, only 2.6% (9/348) were repositioned, resulting in a 97.4% (339/348) successful screw placement rate. The average age was 64, and 48% were female. Average body mass index was 31 kg/m2. Based on the GRS CT-based grading, 97.7% (340/348) were graded A or B, 1.7% (6/348) screws were graded C, and only 0.6% (2/348) of screws were graded D. Two complications, explantation of interbody and vacuum-assisted wound closure, were reported as requiring a return to the OR, but these were not related to robotic guidance or pedicle screws. This study demonstrated a high level of accuracy (97.7%) in the first 56 cases using navigated, robot-assisted surgery based on the GRS. There were two non-screw-related complications requiring return to the operating room.Entities:
Keywords: Minimally invasive surgery; Pedicle screw placement accuracy; Robotic-assisted spine surgery; Robotic-navigated
Year: 2019 PMID: 31321615 PMCID: PMC7237509 DOI: 10.1007/s11701-019-00994-3
Source DB: PubMed Journal: J Robot Surg ISSN: 1863-2483
Fig. 1Screw insertion with the robotic positioning system
Gertzbein and Robbins classification system of pedicle screw accuracy
| Grade | Breach distance (mm) |
|---|---|
| A | 0 |
| B | < 2 |
| C | < 4 |
| D | < 6 |
| E | > 6 |
Baseline characteristics
| Parameter | Overall |
|---|---|
| Number of patients | 56 |
| Gender | |
| Female, | 27 (48.2%) |
| Male, | 29 (51.8%) |
| Age, mean (SD, range) | 64.2 (11) (31–87) |
| BMI, mean (SD, range) | 31 (5) (21–44) |
| Mean number of screws/case | 6 |
| Mean number of vertebrae | 3.2 |
Procedure characteristics
| Parameter | Overall |
|---|---|
| Number of levels with screws placed, | |
| 1 | 4 (7.1%) |
| 2 | 13 (23.2%) |
| 3 | 22 (39.3%) |
| 4 | 8 (14.3%) |
| 5 | 8 (14.3%) |
| 6 | 1 (1.8%) |
| Disposition, | |
| Home | 19 (33.9%) |
| Rehab | 18 (32.1%) |
| Home health | 12 (21.4%) |
| Skilled nursing facility | 7 (12.5%) |
Fig. 2A bar graph depicts the breakdown of vertebral levels among 56 spinal surgery cases including 174 levels. The most common levels were L4 and L5
Pedicle screw placement accuracy grades according to the Gertzbein and Robbins classification system
| Number of screws | |||||
|---|---|---|---|---|---|
| Grade | A | B | C | D | E |
| L1 | 5 | 1 | 0 | 0 | 0 |
| L2 | 21 | 3 | 2 | 0 | 0 |
| L3 | 63 | 8 | 0 | 1 | 0 |
| L4 | 83 | 11 | 0 | 0 | 0 |
| L5 | 86 | 3 | 2 | 1 | 0 |
| S1 | 53 | 3 | 2 | 0 | 0 |
| Total | 311 | 29 | 6 | 2 | 0 |