Literature DB >> 22698392

A prospective study on the use of intraoperative computed tomography (iCT) for image-guided placement of thoracic pedicle screws.

Shree Kumar Dinesh1, Rajendra Tiruchelvarayan, Ivan Ng.   

Abstract

BACKGROUND: Placement of thoracic pedicle screws is a technically demanding procedure. The risk of thoracic pedicle breaches range from 6.5 to 41%. Current image guidance systems consist of computer based systems utilizing preoperative CT scans or 2D/3D intraoperative fluoroscopy.
OBJECTIVE: The aim of this prospective study was to evaluate the clinical feasibility and accuracy of a new intraoperative CT (iCT) based image guidance system for thoracic pedicle screw instrumentation.
METHODS: We prospectively studied the use of iCT for the first 43 consecutive cases for which thoracic pedicle screws were inserted as part of the instrumentation for spinal fusion between April 2008 and July 2011. In every case, a post-instrumentation intraoperative check CT was done before wound closure to assess accuracy of implant placement. Outcomes were analysed with regards to the incidence of pedicle wall violations detected on intraoperative check CT imaging, and the rate of immediate intraoperative revision of misplaced screws. Pedicle violations were graded according to an established classification system.
RESULTS: A total of 261 thoracic pedicle screws (T1-T12) were inserted in 43 patients (age range 13-83). Mean follow-up was 12 months. There were 7 (2.7%) pedicle violations detected on the intraoperative check CT. Out of the seven, three were grade I (< 2 mm), two were grade II (2-4 mm) and rest two were grade III (> 4 mm) violations. Only four of the screws (1.5%) that breached the pedicle wall by more than 2 mm were immediately revised before wound closure.
CONCLUSION: The iCT based spinal neuronavigation system allowed for highly safe and accurate placement (97.3%) of thoracic pedicle screws in our institution with no neurovascular injury reported.

Entities:  

Mesh:

Year:  2012        PMID: 22698392     DOI: 10.3109/02688697.2012.690917

Source DB:  PubMed          Journal:  Br J Neurosurg        ISSN: 0268-8697            Impact factor:   1.596


  13 in total

1.  Reliability of the Planned Pedicle Screw Trajectory versus the Actual Pedicle Screw Trajectory using Intra-operative 3D CT and Image Guidance.

Authors:  Catherine A Miller; Charles G Ledonio; Matthew A Hunt; Farhan Siddiq; David W Polly
Journal:  Int J Spine Surg       Date:  2016-10-24

2.  Computed tomographic morphometric analysis of cervical pedicles in a multi-ethnic Asian population and relevance to subaxial cervical pedicle screw fixation.

Authors:  Thangaraj Munusamy; Ady Thien; Mark Gerad Anthony; Ramesh Bakthavachalam; Shree Kumar Dinesh
Journal:  Eur Spine J       Date:  2014-08-26       Impact factor: 3.134

Review 3.  Techniques and accuracy of thoracolumbar pedicle screw placement.

Authors:  Varun Puvanesarajah; Jason A Liauw; Sheng-Fu Lo; Ioan A Lina; Timothy F Witham
Journal:  World J Orthop       Date:  2014-04-18

Review 4.  Methods to determine pedicle screw placement accuracy in spine surgery: a systematic review.

Authors:  Ahmed A Aoude; Maryse Fortin; Rainer Figueiredo; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Eur Spine J       Date:  2015-03-07       Impact factor: 3.134

5.  Accuracy Assessment of Freehand Pedicular Screw Placement.

Authors:  Afshin Ahmadzadeh Heshmati; Shahab Ilka
Journal:  Arch Bone Jt Surg       Date:  2020-07

6.  Accuracy and reliability of spinal navigation: An analysis of over 1000 pedicle screws.

Authors:  Dinesh Shree Kumar; Nishanth Ampar; Loo Wee Lim
Journal:  J Orthop       Date:  2019-10-24

7.  The potential risk of spinal cord injury from pedicle screw at the apex of adolescent idiopathic thoracic scoliosis: magnetic resonance imaging evaluation.

Authors:  Shoufeng Wang; Yong Qiu; Wenjun Liu; Benlong Shi; Bin Wang; Yang Yu; Zezhang Zhu; Bangping Qian; Feng Zhu; Xu Sun
Journal:  BMC Musculoskelet Disord       Date:  2015-10-20       Impact factor: 2.362

Review 8.  Differences between Manufacturers of Computed Tomography-Based Computer-Assisted Surgery Systems Do Exist: A Systematic Literature Review.

Authors:  Anas Nooh; Joushua Lubov; Ahmed Aoude; Sultan Aldebeyan; Peter Jarzem; Jean Ouellet; Michael H Weber
Journal:  Global Spine J       Date:  2017-02-01

9.  Surgical treatment of adult and pediatric C1/C2 subluxation with intraoperative computed tomography guidance.

Authors:  Ji Min Ling; Rajendra Tiruchelvarayan; Wan T Seow; Hua Bi Ng
Journal:  Surg Neurol Int       Date:  2013-03-22

10.  Screw Placement Accuracy and Outcomes Following O-Arm-Navigated Atlantoaxial Fusion: A Feasibility Study.

Authors:  Jacob D Smith; Megan M Jack; Nicholas R Harn; Judson R Bertsch; Paul M Arnold
Journal:  Global Spine J       Date:  2015-09-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.