Fatma Demet İnce1, Hamit Yaşar Ellidağ2, Mehmet Koseoğlu2, Neşe Şimşek1, Hülya Yalçın1, Mustafa Osman Zengin1.
Abstract
OBJECTIVES: Urinalysis is one of the most commonly performed tests in the clinical laboratory. However, manual microscopic sediment examination is labor-intensive, time-consuming, and lacks standardization in high-volume laboratories. In this study, the concordance of analyses between manual microscopic examination and two different automatic urine sediment analyzers has been evaluated. DESIGN AND METHODS: 209 urine samples were analyzed by the Iris iQ200 ELITE (İris Diagnostics, USA), Dirui FUS-200 (DIRUI Industrial Co., China) automatic urine sediment analyzers and by manual microscopic examination. The degree of concordance (Kappa coefficient) and the rates within the same grading were evaluated.Entities:
Keywords: Autoanalysis; Microscopy; Urinalysis
Year: 2016 PMID: 28856199 PMCID: PMC5574505 DOI: 10.1016/j.plabm.2016.03.002
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Comparison (numbers of samples in each category) between manual microscopy, FUS-200 and Iris iQ200. The shaded numbers indicate samples with the same grade agreement.
Comparison (numbers of samples in each category) between the two automated analyzers. The shaded numbers indicate samples with the same grade agreement.
Agreement of urinalysis results between the FUS-200, Iris iQ200 and manual microscopy represented as the rate(%) with the same grade.
| 86.1 | 74.1 | 82.7 | |
| 92.3 | 77.9 | 79.9 | |
| 90.4 | 81.8 | 86.1 |
Degree of concordance between methods for clinically positive results.
| 0.68 (0.54–0.82) | <0.001 | 0.77 (0.65–0.90) | <0.001 | 0.81 (0.70–0.92) | <0.001 | |
| 0.65 (0.55–0.76) | <0.001 | 0.73 (0.63–0.82) | <0.001 | 0.81 (0.72–0.89) | <0.001 | |
| 0.59 (0.46–0.72) | <0.001 | 0.50 (0.36–0.64) | <0.001 | 0.62 (0.48–0.76) | <0.001 | |
| 0.47 (0.33–0.62) | <0.001 | 0.52 (0.38–0.67) | <0.001 | 0.82 (0.70–0.93) | <0.001 | |
| 0.54 (0.37–0.71) | <0.001 | 0.57 (0.40–0.74) | <0.001 | 0.69 (0.49–0.88) | <0.001 | |
| 0.17 (−0.14–0.47) | 0.001 | 0.19 (−0.13–0.52) | <0.001 | 0.67 (0.05–1.00) | <0.001 | |
| 0.13 (−0.06–0.32) | 0.051 | 0.10 (−0.13–0.32) | 0.135 | 0.04 (−0.10–0.18) | 0.45 | |
p>0.05, Kappa is not significant. There is no agreement between methods.
Moderate agreement.
Good agreement.
Very good agreement.
Sensitivity, specificity and predictive values for automatic analyzers compared to manual microscopy.
| Erythrocyte | 72.7 | 94.9 | 72.7 | 94.9 |
| Leukocyte | 68.1 | 95.7 | 92.8 | 78.6 |
| Epithelial cell | 61.1 | 93.7 | 76.7 | 87.3 |
| Erythrocyte | 75.8 | 97.7 | 86.2 | 95.6 |
| Leukocyte | 77.7 | 93.9 | 91.2 | 83.7 |
| Epithelial cell | 42.6 | 92 | 92 | 83.2 |