[Evaluation of iQ200 automated urine microscopy analyzer].

Joowon Park1, Jongwan Kim.   

Abstract

BACKGROUND: Microscopic examination of urine sediment is one of the most commonly performed tests in the clinical laboratory. However, manual microscopic sediment examination is labor-intensive, time-consuming and imprecise. In this study, we evaluated the analytical performance and clinical usefulness of a recently introduced image-based automated urinalysis system, Iris iQ200 (Iris Diagnostics, USA).
METHODS: We assessed the iQ200 for linearity, precision and carryover rate using patient's samples and quality control materials. On 337 urine samples, urine sediment analyses performed by the iQ200 were compared with manual microscopy results.
RESULTS: The iQ200 showed a good linearity (r2>0.99) for all cellular components analyzed. Within-run and total CVs on urine specimens and quality control samples were less than 10% except for within-run CV for the samples with low concentration of the squamous epithelial cells. The carryover rates were 0.21% for RBCs and 1.92% for WBCs. The agreement rates within one grade between the iQ200 and manual microscopy for RBCs, WBCs, and squamous epithelial cells were 93.8%, 94.2% and 96.9%, respectively.
CONCLUSIONS: Since the iQ200 showed a reliable analytical performance and good concordance with manual microscopy, it could be useful in the clinical practice as a screening procedure.

Entities:  

Mesh:

Year:  2008        PMID: 18728375     DOI: 10.3343/kjlm.2008.28.4.267

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  3 in total

1.  Comparison of the Automated cobas u 701 Urine Microscopy and UF-1000i Flow Cytometry Systems and Manual Microscopy in the Examination of Urine Sediments.

Authors:  Wonmok Lee; Jung-Sook Ha; Nam-Hee Ryoo
Journal:  J Clin Lab Anal       Date:  2016-02-03       Impact factor: 2.352

Review 2.  Progress in Automated Urinalysis.

Authors:  Matthijs Oyaert; Joris Delanghe
Journal:  Ann Lab Med       Date:  2019-01       Impact factor: 3.464

3.  The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysis.

Authors:  Fatma Demet İnce; Hamit Yaşar Ellidağ; Mehmet Koseoğlu; Neşe Şimşek; Hülya Yalçın; Mustafa Osman Zengin
Journal:  Pract Lab Med       Date:  2016-03-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.