Comparison of LabUMat-with-UriSed and iQ200 fully automatic urine sediment analysers with manual urine analysis.

Okhan Kadir Akin1, Muhittin A Serdar, Zeynep Cizmeci, Ozlem Genc, Suleyman Aydin.   

Abstract

Urine analysis is one of the most common tests for assessing urinary-tract and kidney diseases. In recent years there have been new developments in the automation of this test. The objective of the present study was to compare the performances of two urine sediment analysers, namely LabUMat with UriSed (77 Elektronika Kft, Budapest, Hungary) and iQ200 (Iris Diagnostics, Chatsworth, CA, U.S.A.), with the KOVA method for manual urine measurement by evaluating the results in terms of similar parameters (cells or particles per lower-power field or high-power field). The results obtained using the UriSed and iQ200 analysers were more reproducible (7.1-30.2 and 14.9-35.4% respectively) than those obtained using the manual technique (17.9-44.4%). Significant correlations were established among the three techniques in the evaluation of leucocytes, erythrocytes and epithelial cells. Although the UriSed, iQ200 and visual-microscopic measurements were in agreement, confirmation of the results from automated methods by manual urine analyses is significantly useful, especially for pathological cases that were close to the limits of the techniques.

Entities:  

Mesh:

Year:  2009        PMID: 19021504     DOI: 10.1042/BA20080188

Source DB:  PubMed          Journal:  Biotechnol Appl Biochem        ISSN: 0885-4513            Impact factor:   2.431


  11 in total

1.  Comparison of fully automated urine sediment analyzers H800-FUS100 and LabUMat-UriSed with manual microscopy.

Authors:  Hatice Yüksel; Elif Kiliç; Aysun Ekinci; Osman Evliyaoğlu
Journal:  J Clin Lab Anal       Date:  2013-07       Impact factor: 2.352

2.  Comparative Performance Analysis of Urised 3 and DIRUI FUS-200 Automated Urine Sediment Analyzers and Manual Microscopic Method.

Authors:  Emre Yalcinkaya; Hayriye Erman; Eray Kirac; Afife Serifoglu; Alperen Aksoy; Ferruh K Isman; Mustafa B Cekmen
Journal:  Medeni Med J       Date:  2019-09-27

3.  High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer.

Authors:  Ozgur Aydin; Hamit Yasar Ellidag; Esin Eren; Necat Yilmaz
Journal:  J Med Biochem       Date:  2015-07-14       Impact factor: 3.402

4.  Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy.

Authors:  Ebubekir Bakan; Nurinnisa Ozturk; Nurcan Kilic Baygutalp; Elif Polat; Kadriye Akpinar; Emrullah Dorman; Harun Polat; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

5.  Evaluation of the analytical performances of Cobas 6500 and Sysmex UN series automated urinalysis systems with manual microscopic particle counting.

Authors:  Ebubekir Bakan; Zafer Bayraktutan; Nurcan Kilic Baygutalp; Mehmet Ali Gul; Fatma Zuhal Umudum; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

6.  Advances in Automated Urinalysis Systems, Flow Cytometry and Digitized Microscopy.

Authors:  Sun Young Cho; Mina Hur
Journal:  Ann Lab Med       Date:  2019-01       Impact factor: 3.464

Review 7.  Progress in Automated Urinalysis.

Authors:  Matthijs Oyaert; Joris Delanghe
Journal:  Ann Lab Med       Date:  2019-01       Impact factor: 3.464

8.  Comparison of the performance of the IDEXX SediVue Dx® with manual microscopy for the detection of cells and 2 crystal types in canine and feline urine.

Authors:  Annalisa M Hernandez; Graham E A Bilbrough; Dennis B DeNicola; Celine Myrick; Suzanne Edwards; Jeremy M Hammond; Alex N Myers; Johanna C Heseltine; Karen Russell; Marco Giraldi; Mary B Nabity
Journal:  J Vet Intern Med       Date:  2018-12-03       Impact factor: 3.333

9.  A comparison of automated urine analyzers cobas 6500, UN 3000-111b and iRICELL 3000 with manual microscopic urinalysis.

Authors:  Piraya Tantisaranon; Kanyarat Dumkengkhachornwong; Peechana Aiadsakun; Areerat Hnoonual
Journal:  Pract Lab Med       Date:  2021-01-18

10.  The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysis.

Authors:  Fatma Demet İnce; Hamit Yaşar Ellidağ; Mehmet Koseoğlu; Neşe Şimşek; Hülya Yalçın; Mustafa Osman Zengin
Journal:  Pract Lab Med       Date:  2016-03-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.