Literature DB >> 12651815

Quantitative urine particle analysis: integrative approach for the optimal combination of automation with UF-100 and microscopic review with KOVA cell chamber.

Cornelia Ottiger1, Andreas R Huber.   

Abstract

BACKGROUND: Automated systems have enabled the counting of particles in urine to be standardized. Their superiority over traditional sediment analysis has been well documented, but they have not gained wide acceptance. The reasons for this are that sediment analysis has been performed and interpreted for decades. Additionally, pathologic casts and other unknown particles still must be confirmed under the microscope. Furthermore, comparison between the methods has revealed outliers and thus decreased confidence in automation.
METHODS: We used the standardized KOVA cell chamber system to count particles and compared the results with UF-100 flow cytometry as an alternative to traditional sediment analysis.
RESULTS: We compared 252 randomly selected urine samples and obtained a review rate of 33%. Microscopic verification was necessary because of the presence of casts, yeast, sperm, dysmorphic erythrocytes, and some misclassified erythrocytes or leukocytes that were detected by incongruent dipstick results and abnormal scattergrams. We obtained correlation coefficients of 0.966 for erythrocytes and 0.935 for leukocytes. Criteria for an algorithm to identify samples that needed microscopic review were derived from comparisons between the number of particles from UF-100, dipstick results, cell chamber counting, and sediment analysis.
CONCLUSIONS: Automated cell counting combined with microscopic counting with a standardized cell chamber system is useful. An objective algorithm for review criteria can be developed via systematic comparison of UF-100 flow cytometry and microscopy. Only urine samples that meet these criteria need to be confirmed microscopically.

Entities:  

Mesh:

Year:  2003        PMID: 12651815     DOI: 10.1373/49.4.617

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  8 in total

1.  New generation IQ-200 automated urine microscopy analyzer compared with KOVA cell chamber.

Authors:  Emel Altekin; Ozgur Kadiçesme; Pinar Akan; Tuncay Kume; Ozgul Vupa; Gul Ergor; Hakan Abacioglu
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

2.  Comparison of the Automated cobas u 701 Urine Microscopy and UF-1000i Flow Cytometry Systems and Manual Microscopy in the Examination of Urine Sediments.

Authors:  Wonmok Lee; Jung-Sook Ha; Nam-Hee Ryoo
Journal:  J Clin Lab Anal       Date:  2016-02-03       Impact factor: 2.352

3.  Evaluation of a standardized procedure for [corrected] microscopic cell counts [corrected] in body fluids.

Authors:  Jane F Emerson; Scott S Emerson
Journal:  J Clin Lab Anal       Date:  2005       Impact factor: 2.352

4.  Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments.

Authors:  S Mayo; D Acevedo; C Quiñones-Torrelo; I Canós; M Sancho
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

5.  Flow cytometry analysis using sysmex UF-1000i classifies uropathogens based on bacterial, leukocyte, and erythrocyte counts in urine specimens among patients with urinary tract infections.

Authors:  Tor Monsen; Patrik Rydén
Journal:  J Clin Microbiol       Date:  2014-12-03       Impact factor: 5.948

6.  Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy.

Authors:  Ebubekir Bakan; Nurinnisa Ozturk; Nurcan Kilic Baygutalp; Elif Polat; Kadriye Akpinar; Emrullah Dorman; Harun Polat; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

7.  Chronic Urinary Infection in Overactive Bladder Syndrome: A Prospective, Blinded Case Control Study.

Authors:  Zainab Khan; Gareth D Healey; Roberta Paravati; Nidhika Berry; Eugene Rees; Lavinia Margarit; Deyarina Gonzalez; Simon Emery; Robert Steven Conlan
Journal:  Front Cell Infect Microbiol       Date:  2021-09-30       Impact factor: 5.293

8.  The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysis.

Authors:  Fatma Demet İnce; Hamit Yaşar Ellidağ; Mehmet Koseoğlu; Neşe Şimşek; Hülya Yalçın; Mustafa Osman Zengin
Journal:  Pract Lab Med       Date:  2016-03-11