Comparison of fully automated urine sediment analyzers H800-FUS100 and LabUMat-UriSed with manual microscopy.

Hatice Yüksel1, Elif Kiliç, Aysun Ekinci, Osman Evliyaoğlu.   

Abstract

BACKGROUND: Recent technical developments have focused on the full automation of urinalyses, however the manual microscopic analysis of urine sediment is considered the reference method. The aim of this study was to compare the performances of the LabUMat-UriSed and the H800-FUS100 with manual microscopy, and with each other.
METHODS: The urine sediments of 332 urine samples were examined by these two devices (LabUMat-UriSed, H800-FUS100) and manual microscopy.
RESULTS: The reproducibility of the analyzers, UriSed and Fus100 (4.1-28.5% and 4.7-21.2%, respectively), was better than that with manual microscopy (8.5-33.3%). The UriSed was more sensitive for leukocytes (82%), while the Fus-100 was more sensitive for erythrocyte cell counting (73%). There were moderate correlations between manual microscopy and the two devices, UriSed and Fus100, for erythrocyte (r = 0.496 and 0.498, respectively) and leukocyte (r = 0.597 and 0.599, respectively) cell counting however the correlation between the two devices was much better for erythrocyte (r = 0.643) and for leukocyte (r = 0.767) cell counting.
CONCLUSION: It can be concluded that these two devices showed similar performances. They were time-saving and standardized techniques, especially for reducing preanalytical errors such as the study time, centrifugation, and specimen volume for sedimentary analysis; however, the automated systems are still inadequate for classifying the cells that are present in pathological urine specimens.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  Automated urine analysis; Fus100; manual urine sediment analysis; urised

Mesh:

Year:  2013        PMID: 23852791      PMCID: PMC6807495          DOI: 10.1002/jcla.21604

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  13 in total

1.  Automation of urine sediment examination: a comparison of the Sysmex UF-100 automated flow cytometer with routine manual diagnosis (microscopy, test strips, and bacterial culture).

Authors:  K Hannemann-Pohl; S C Kampf
Journal:  Clin Chem Lab Med       Date:  1999-07       Impact factor: 3.694

2.  Safely reducing manual urine microscopy analyses by combining urine flow cytometer and strip results.

Authors:  S Roggeman; Z Zaman
Journal:  Am J Clin Pathol       Date:  2001-12       Impact factor: 2.493

3.  European guidelines for urinalysis: a collaborative document produced by European clinical microbiologists and clinical chemists under ECLM in collaboration with ESCMID.

Authors:  O Aspevall; H Hallander; V Gant; T Kouri
Journal:  Clin Microbiol Infect       Date:  2001-04       Impact factor: 8.067

4.  New generation IQ-200 automated urine microscopy analyzer compared with KOVA cell chamber.

Authors:  Emel Altekin; Ozgur Kadiçesme; Pinar Akan; Tuncay Kume; Ozgul Vupa; Gul Ergor; Hakan Abacioglu
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

5.  [Hematuria and acanthocyturia in patients with diabetes].

Authors:  Yuki Ono; Kazuhiro Sugimoto; Maki Goto; Masahiko Kimura; Takeshi Kasai; Masaru Shoji; Minoru Yasujima
Journal:  Rinsho Byori       Date:  2005-11

6.  Urine sediment examination: a comparison of automated urinalysis systems and manual microscopy.

Authors:  Tzu-I Chien; Jau-Tsuen Kao; Hui-Lan Liu; Po-Chang Lin; Jhih-Sian Hong; Han-Peng Hsieh; Miao-Ju Chien
Journal:  Clin Chim Acta       Date:  2007-05-26       Impact factor: 3.786

7.  Comparison of LabUMat-with-UriSed and iQ200 fully automatic urine sediment analysers with manual urine analysis.

Authors:  Okhan Kadir Akin; Muhittin A Serdar; Zeynep Cizmeci; Ozlem Genc; Suleyman Aydin
Journal:  Biotechnol Appl Biochem       Date:  2009-06       Impact factor: 2.431

8.  Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sediments.

Authors:  S Mayo; D Acevedo; C Quiñones-Torrelo; I Canós; M Sancho
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

9.  [Evaluation of the performances of the iQ(®)200 ELITE automated urine microscopy analyser and comparison with manual microscopy method].

Authors:  G Dewulf; D Harrois; E Mazars; C Cattoen; F Canis
Journal:  Pathol Biol (Paris)       Date:  2009-11-25

10.  The automation of sediment urinalysis using a new urine flow cytometer (UF-100)

Authors:  D Fenili; B Pirovano
Journal:  Clin Chem Lab Med       Date:  1998-12       Impact factor: 3.694

View more
  15 in total

1.  Performance Evaluation of Three URiSCAN Devices for Routine Urinalysis.

Authors:  Kiwoong Ko; Min-Jung Kwon; Seungho Ryu; Hee-Yeon Woo; Hyosoon Park
Journal:  J Clin Lab Anal       Date:  2015-08-24       Impact factor: 2.352

2.  Comparison of the Automated cobas u 701 Urine Microscopy and UF-1000i Flow Cytometry Systems and Manual Microscopy in the Examination of Urine Sediments.

Authors:  Wonmok Lee; Jung-Sook Ha; Nam-Hee Ryoo
Journal:  J Clin Lab Anal       Date:  2016-02-03       Impact factor: 2.352

3.  UriSed 3 and UX-2000 automated urine sediment analyzers vs manual microscopic method: A comparative performance analysis.

Authors:  Sathima Laiwejpithaya; Preechaya Wongkrajang; Kanit Reesukumal; Chonticha Bucha; Suriya Meepanya; Chanutchaya Pattanavin; Varanya Khejonnit; Achara Chuntarut
Journal:  J Clin Lab Anal       Date:  2017-05-02       Impact factor: 2.352

4.  The influence of vitamin C on the urine dipstick tests in the clinical specimens: a multicenter study.

Authors:  Wonmok Lee; Yukyung Kim; Soonhee Chang; A-Jin Lee; Chang-Ho Jeon
Journal:  J Clin Lab Anal       Date:  2016-10-24       Impact factor: 2.352

5.  High False Positives and False Negatives in Yeast Parameter in an Automated Urine Sediment Analyzer.

Authors:  Ozgur Aydin; Hamit Yasar Ellidag; Esin Eren; Necat Yilmaz
Journal:  J Med Biochem       Date:  2015-07-14       Impact factor: 3.402

6.  Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy.

Authors:  Ebubekir Bakan; Nurinnisa Ozturk; Nurcan Kilic Baygutalp; Elif Polat; Kadriye Akpinar; Emrullah Dorman; Harun Polat; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

Review 7.  Preanalytical requirements of urinalysis.

Authors:  Joris Delanghe; Marijn Speeckaert
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

8.  The effectiveness of BD Vacutainer® Plus Urinalysis Preservative Tubes in preservation of urine for chemical strip analysis and particle counting.

Authors:  Merve Kaymak Ekşioğlu; Özlem Çakır Madenci; Nihal Yücel; Abdullah Elçi; Bülent Turhan; Gani Orhan; Asuman Orçun
Journal:  Biochem Med (Zagreb)       Date:  2016       Impact factor: 2.313

9.  The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysis.

Authors:  Fatma Demet İnce; Hamit Yaşar Ellidağ; Mehmet Koseoğlu; Neşe Şimşek; Hülya Yalçın; Mustafa Osman Zengin
Journal:  Pract Lab Med       Date:  2016-03-11

10.  Preanalytics of urine sediment examination: effect of relative centrifugal force, tube type, volume of sample and supernatant removal.

Authors:  Amalija Bunjevac; Nora Nikolac Gabaj; Marijana Miler; Anita Horvat
Journal:  Biochem Med (Zagreb)       Date:  2018-01-10       Impact factor: 2.313

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.