Abstract
PURPOSE: To investigate whether corneal thickness parameters measured by optical coherence tomography (OCT), such as central corneal thickness (CCT), central corneal stromal thickness (CCST), and central corneal epithelial thickness (CCET), influence the intraocular pressure (IOP) difference measured by Goldmann applanation tonometry (GAT) and non-contact tonometry (NCT).Entities:
Mesh:
Year: 2016 PMID: 26998838 PMCID: PMC4801388 DOI: 10.1371/journal.pone.0151868
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1(A) Example of CCT measurement using Cirrus-OCT. (B) CCET was defined as the distance from the air–cornea junction to the Bowman’s layer line (white line). (C) CCT was defined as the distance from the air–cornea junction to the cornea–aqueous humor junction (white line).
Descriptive statics of involved patients.
| Mean | Standard deviation | Minimum | Maximum | |
|---|---|---|---|---|
| Age | 48.2 | 11.0 | 19 | 70 |
| BCVA | 0.976 | 0.056 | 0.8 | 1.0 |
| Cylinder | -0.552 | 0.567 | -1.500 | 1.150 |
| Spherical Equivalent | -1.561 | 2.103 | -5.875 | 1.825 |
| GAT IOP | 16.7 | 3.0 | 10 | 23 |
| NCT IOP | 18.1 | 3.8 | 9 | 27 |
| GAT IOP—NCT IOP | 1.5 | 1.7 | -2 | 5 |
| CCT | 559.6 | 36.0 | 490 | 630 |
| CCET | 57.9 | 5.6 | 46 | 71 |
| CCST | 501.7 | 33.8 | 440 | 571 |
| CCSP | 0.896 | 0.009 | 0.874 | 0.915 |
BCVA, best corrected visual acuity; GAT, Goldmann applanation tonometry; IOP, intraocular pressure; NCT, Non-contact tonometry; CCT, Central corneal thickness; CCET, Central corneal epithelial thickness; CCST, Central corneal stromal thickness; CCSP, Proportion of stromal thickness in central corneal thickness.
Correlations between IOP parameters and central corneal thickness parameters (Pearson correlation coefficient R and P value in parenthesis).
| CCT | CCET | CCST | CCSP | |
|---|---|---|---|---|
| GAT IOP | 0.648(<0.001) | 0.057(0.696) | 0.679(<0.001) | 0.379(0.007) |
| NCT IOP | 0.676(<0.001) | 0.242(0.090) | 0.678(<0.001) | 0.199(0.165) |
| GAT IOP—NCT IOP | 0.333(0.018) | 0.435(0.002) | 0.281(0.051) | -0.239(0.095) |
GAT, Goldmann applanation tonometry; IOP, intraocular pressure; NCT, Non-contact tonometry; CCT, Central corneal thickness; CCET, Central corneal epithelial thickness; CCST, Central corneal stromal thickness; CCSP, Proportion of stromal thickness in central corneal thickness.
Fig 2(A) Correlation between CCET and IOP measured by GAT. (B) Correlation between CCET and IOP measured by NCT. (C) Correlation between CCST and GAT IOP. (D) Correlation between CCST and NCT IOP.
Fig 3(A) IOP difference between CCT and IOP measured by GAT and NCT. (B) Correlation between GAT IOP and NCT IOP difference and CCET. (C) Correlation between GAT IOP and NCT IOP difference and CCST.