Technical complications of implant-causes and management: A comprehensive review.

Swati Gupta1, Hemant Gupta2, Amrit Tandan1.   

Abstract

Given the increasing popularity of dental implants, the number of failures due to late implant fracture is also expected to increase. Hence, the scope for prevention and management needs to be emphasized. The objective of this review article is to analyze the various causes of failure of dental implants due to implant fixture/abutment screw fractures and also to enumerate the management and the preventive options for these failures, thereby aiming to help the clinicians to properly plan the implant-supported prosthesis treatment by considering the important biomechanical aspects of this type of rehabilitation. The present review emphasizes the causes and management of technical complications and not the incidence of such complications.

Entities:  

Keywords:  Abutment screw fracture; implant failure; implant fixture fracture

Year:  2015        PMID: 26668445      PMCID: PMC4668729          DOI: 10.4103/0975-5950.168233

Source DB:  PubMed          Journal:  Natl J Maxillofac Surg        ISSN: 0975-5950


INTRODUCTION

Given the increasing popularity of dental implants, an increase in the number of failures due to late fractures, that is, post successful osseointegration, is expected. The failure of dental implants is due not only to biological factors, such as unsuccessful osseointegration or the presence of periimplantitis, butthey also result from technical complications thatinvolve implant body/fixture fracture, abutment screw fracture, abutment fracture, fractured prosthesis, etc. The failure of an implant as a single entity, irrespective of its components, may be classified as early or late. Early failures occur shortly after surgery and are characterized by the lack of osseointegration. In contrast, late failures correspond to those implants that have been regarded as successful for some time, and they occur after prostheticrestoration has been made. There are two main causes for late implant fracture: Mechanical problems, including fracturesMetal fatigue, due to biomechanical overloading, appears to be the most frequent cause[12] Loss of supporting tissue secondary to infection or periimplantitis— The prevalence of periimplantitis is estimated to be 4-15% among the surviving implant population (i.e., implants still in the mouth). The present review discusses the various technical/mechanical complications related to the restorative aspect of the dental implant-based restorations. However, to solve these complications, one needs to understand the mechanics of such components. This paper would entail the causes, mechanics, prevention, and management of such component-related complications.

Implant fixture fracture

Various researchers have listed the causes of implant fractures, but there is no universally accepted classification so far. There are two main causes of implant fracture: (a) Mechanical overloads leading to metal fatigue. When the resistance limit is exceeded, it results in a fracture. Overload can also be a consequence of the patient's physiological alterations (e.g., para functional activity). In effect, both centric and eccentric bruxism can lead to implant overload and metal fatigue. Other overload-related factors are of prosthetic origin, including inadequate occlusion, the presence of distal extensions or cantilevers in implant-supported prostheses, and a lack of prosthetic passive fit over the implants.[13456] (b) Peri implant vertical bone loss is attributable to both chronic peri implant inflammation and occlusal trauma. When vertical bone loss coincides with the apical limit of the screw, thereby joining trans epithelial abutment to implant, the risk of implant fracture increases drastically.[13478] The implant may also fracture due to defects in implant design and manufacturing.[9] Metal fractures are related to microstructures that influence the material properties. They are related to the localized identification of chemical phases and segregations, frequently associated with failures at interfaces or components, which in turn can cause the fractures.[10]

Incidence/location of implant fracture

The incidence of dental implant fracture is between 0.16-1.5%, as reported in the literature.[234910111213141516] One study reported that the majority of fractured implants are located in the regions of the molars and premolars,[3917] and that the fractures were evenly distributed between the maxilla and the mandible,[9] whereas another study reported more fractures in the maxilla than in the mandible.[17]

Diameter of implant

Implants with small diameters such as 4 mm[17] and 3.75 mm[910] tend to fracture more easily than those with large diameters, especially when placed in a posterior location. A study has reported that an implant with a diameter of 5.0 mm is three times stronger than one with a diameter of 3.75 mm, whereas a 6.0-mm implant is six times stronger than a 3.75-mm implant.[18]

Mechanism of implant fracture

The specific bone loss pattern seen in implant fracture cases has been described as a primary cause and a mechanism of fracture.[31920] Coronal bone resorption increases the bending stress of the implant because of the loss of supporting bone,[21] which usually extends to the level corresponding to the end of the abutment screw, there by reducing the resistance to bending in this region. When such reabsorption extends apically beyond the third implant thread, it reaches a structurally weak zone coinciding with the end of the prosthetic screw and the border of the hybrid surface. This contributes to the fatigue at a point of low resistance to torque. The warning signs of such fatigue include loosening, torsion, or fracture of the post screws, and ceramic fractures of the prosthesis.

Clinical manifestations

Patients may often report spontaneous bleeding and mobility. Exploration (manually or electronically) confirms increased mobility, increased pocket depth, and gingival indexes, and occasionally, also, plaque accumulation resulting from the patient's fear of the pain is triggered by brushing. Radiologically, separation of the fragments and bone loss may be seen. It has been reported that marginal bone reabsorption seems to be the most important risk factor indicating the start of an implant fracture, and may often extend beyond the actual fracture line.[9] Thus, an x-ray study is very useful. For diagnostic purposes, the fracture risk factors[22] have been grouped into three main categories [Table 1]. In the presence of more than three factors pertaining to one or more of these categories, the risk of fracture is high.
Table 1

Clinical findings frequently documented in the literature and related to implant fracture, grouped by categories

Clinical findings frequently documented in the literature and related to implant fracture, grouped by categories

MANAGEMENT

There are three management options in the event of an implant fracture:[923] Complete removal of the fractured implant using explantation trephines.[923] After the removal of the fractured implant, a new larger diameter implant can be installed at the same surgical bed or at another placeto achieve primary stability.[20] Their complete removal is considered to be the best solution to the problem Removal of the coronal portion of the fractured implant with the purpose of placing a new prosthetic post.[923] It is essential to radiologically confirm the absence of radio transparency and to electronically determine the mobility of the fragment. A prosthetic post should be considered only if there is still sufficient remaining internal thread retention[9] Removal of the coronal portion of the fractured implant, leaving the remaining apical part integrated in the bone.[923] and the missing tooth/teeth can be restored using conventional prosthodontic procedures like removable partial denture, fixed partial denture etc. A recent study described “apicoectomy” as a suitable technique for removing fractured implants and inserting new implants at the same clinical session. This technique was based on opening a hole in the bone in order to improve the visualization of the apical fragments of the fractured implants and to remove those fragments through this hole. Afterward, a new implant is placed in a conventional manner, and the hole is closed by using the same bone that is removed from the patient.[24]

Preventive measures

The following preventive measures have been suggested: Avoid placing hybrid-type implant fixtures Carefully control occlusal forces: Eliminate all posterior contacts in mandibular eccentric movements Perform staggered placement of implants: Avoid a straight-line configuration Avoid or minimize posterior cantilevers and buccolingual offsets, particularly in partially edentulous patients During the chronic loosening of gold or abutment screws or the fracture of components, critically reassess the prosthesis, retighten the abutment or usea gold screw, and check the fit Pronounced bruxers or clenchers who have experienced multiple implant fractures should be managed by the placement of additional implants fractures[17] Ensure perfect fit: Resolder and assess passive fit.[25]

Implant-abutment connection system

The misfit at the abutment-implant interface and the absence of a passive adaptation between the prosthesis and the abutment can lead to fracture and the most common technical complication in implants is a fracture of the abutment screw. The stability of the implant abutment connection is influenced by factors such as internal connection versus external connection, screw head design, screw geometry, materials, screw diameter, preload, joint separating force, settling effect, and overall screw mechanics. Factors affecting the implant-abutment connection system are as follows:

Internal connection versus external connection

The implant-abutment interface connection is generally described as an internal or external connection. Relative to external connections, the internal implant-abutment connections have been shown to present a higher stability and improved force distribution as a result of not only their ability to dissipate lateral loads deeply within the implant, there by shielding the abutment screw better from stress, but also due to the longer internal wall engagement that creates a stiff, unified body to resist joint opening (micro movement).[26272829] However, there is another aspect to stronger internal abutment connection after cyclic loading relative to external connections.[303132] While mechanical strength is gained, the thinner lateral fixture wall at the connecting part of the internal connections may lead to a higher absolute strain value at the cervical area.[31] Such stress shift has raised concerns regarding an increased risk of marginal bone resorption or fixture fracture.[31] Although there are studies that show that the fracture strength in implant systems with an external fixation is equal or superior to those with internal fixation.[3334] The literature supports a gold alloy screw with a flat head, internal hex or square, and a high tightening force (torque driver) as having the greatest ability to produce the best results.[35]

Screw head design

To maximize the preload and minimize the loss of input torque to friction, the head of the screw should be wider than the thread diameter. An abutment head most often should be flat. A tapered head design reduces the clamping effect and reduces the tensile force in the threads of the screw. The tapered screw head distorts and aligns non passive components and gives a non passive casting the appearance of proper fit. However, the superstructure is not deformed permanently and leads to stress in the system. A flat-head screw distributes forces more evenly within the threads and the head of the screw, and is also less likely to distort a non passive casting. As such, the abutment head should also be flat on top to increase the clamping force in the screw head and the tensile force in the threads.[36]

Screw material

The most commonly used retaining screws are either of gold or titanium. Gold screws are designed to be the most “flexible” portion of the implant assembly. Due to their higher modulus of elasticity than titanium, they permit an adequate micro movement to distribute force to the implant body.[37] Implant posts, that are retained with gold or gold coated screws, show a reduction of screw loosening and improved clamping force in comparison with titanium screws.[383940] Titanium alloy has four times the bending fracture resistance of grade 1 titanium. Therefore, abutment screws made of grade 1 titanium will deform and fracture more easily than the alloy. A higher torque magnitude can be used on the titanium alloy abutment screw and female component (implant body).[36] Titanium retaining screws are stronger than gold, but have a lower modulus of elasticity; metal fatigue will produce a gold screw fracture before the titanium retaining screw is affected.[37] The major disadvantage of titanium retaining screws is their tendency to cause galling, which results in excessive friction between the two mating surfaces there by causinga localized welding with a further roughening of the mating surfaces.[41] Galling occurs in the following manner: Titanium of the retainings crew slides in contact with the titanium of the implant body, the coefficient of friction increases whereby titanium molecules transfer from the mating surfaces.[42] This has been described as the adhesive wear mechanism[43] which causes slight damage to both the implant body and the retaining screw threads.

Screw diameter

The greater the diameter, the higher the preload that may be applied and the greater the clamping force on the screw joint. As a general rule, abutment screws loosen less often and can take a higher preload than coping screws.[44]

Preload

The screw joint has been described as two parts tightened together by a screw, such as an abutment and implant being held together by a screw. A screw is tightened by applying torque. The applied torque develops a force within the screw called the preload.[40] As a screw is tightened, it elongates, thereby producing tension. Elastic recovery of the screw pulls the two parts together, creating a clamping force.[45] Excessive forces cause slippage between the threads of the screw and the threads of the bone, resulting in the loss of the preload.[46] Adequate preload of the screw creates less micromotion of the implant-abutment screw interface and less screw loosening.[45474849] Pre loading the screw improves the fatigue resistance and the locking effect of the abutment/implant connection. The preload should fluctuate as little as possible to prevent loosening. Occlusal forces, which result in the loading of implant supported prostheses, will cause changes in the preload of the abutment screws that may result in loosening of the screws.[50]

Joint separating forces

Intra oral separating forces may include off-axis occlusal contacts, lateral excursive contacts, interproximal contacts between natural teeth and implant restorations, protrusive contacts, para functional forces, and non passive frameworks that attach to the implants. Opposing the clamping force is a joint-separating force, which attempts to separate the screw joint. Screw loosening occurs when the joint-separating forces acting on the screw joint are greater than the clamping forces holding the screw unit together.[40]

Settling effect

Settling occurs as the rough spot interfaces flatten under load/micro movements, since they are the only contacting surfaces when the initial tightening torque is applied.[51] It has been reported that 2-10% of the initial preload is lost as a result of settling.[52] This reduction effect is known as embedment relaxation.[53]

Process of screw loosening

The process of screw loosening has been described in two stages.[54] Initially, external forces such as mastication applied to the screw joint causes slippage, contributing to the release of preload of the screw. The second stage of loosening involves continual preload reduction below critical level, allowing threads to turn and the loss of intended screw joint junction.[55]

Incidence of screw loosening

A literature review of clinical complications of osseo integrated implants showed that screw loosening or screw fracture varied 2-45% of the implant restorations, with the highest amount in single crowns.[4] In one of the systematic reviews that evaluated the 5-year survival rates of implant-supported single crowns, it was observed that from the 26 clinical studies included, the cumulative incidence of abutment screw or abutment loosening was 7.3% after 5 years of clinical service in both external and internal connections.[56]

Location of screw loosening

Posterior abutment screws loosen at a higher percentage than anterior implants. This supports the project of eliminating unnecessary occlusal and off-axial forces on implant-supported restorations.[35] One of the studies have reporteda 38% loosening of single implant restorations in the posterior maxilla and mandible,[57] but such studies do not take into account various important factors.

Preventive management

It is recommended in clinical practice that in order to reduce the settling effect, the implant screws should be retightened 10 min after the initial torque application.[5058596061] Mechanical torque gauges should be used instead of hand drivers to ensure a consistent tightening of the implant components to recommended torque values.[51] The use of sealers to fill in the gaps between thes crews and implant threads and adhesives to in creasethe frictional resistance has been suggested to reduce screw loosening. (Ceka Bond, Preat, San Mateo, CA) is listed as an adhesive paste on the package insert.[50] One study of screw loosening supported the following clinical recommendations: (a) Ensure that implants are placed perpendicular to the occlusal plane; (b) frameworks should have minimal cantilever lengths; (c) use components with low tolerance levels for component misfit; and (d) use components with anti rotational features for single tooth restorations.[35]

Management of fractured abutment screw

This mainly involves two options: Either retrieve the fractured screw, or remove the old implant and insert a new implant in one sitting. The retrieval of the fractured abutment screw can be done by the following means: Artery forceps: If the fractured part is above the head of the implant, then it can be unscrewed with the help of artery forceps[62] Ultrasonic scaler: If the fractured part cannot be grasped with any instrument, then use ultrasonic vibrations to disengage the threads. The oscillations from an ultrasonic scaler can gradually reverse the screw out by placing a thin tip of an ultrasonic scaler directly on the top of the screw[63] Lubricant: Lubricate the damaged screw before unscrewing it. Add a few drops of eugenol, handpiece lubricant, or even mineral oil in the area, and then try to remove the screw with a probe or ultrasonic scaler in a counter clockwise direction Contra angle low speed: Using a 1/4 round bur in a contra angle low speed running in a reverse mode could help spin the screw out. The small round drill acts as a screwdriver and holds the head of the fragment[63] Repair or rescue kit: Used when the site of the abutment fracture is deep. The kit consists of drills, drill guides, and tapping instruments. Some kits repair the implant by unscrewing the fractured fragment, while others cut into splinters Be indigenous: If the fractured end is not too deep then prepare a 1-mm slot across the most occlusal portion of the broken screw fragment with the help of a diamond bur and handpiece. Hold the handpiece firmly to avoid having the bur inadvertently jump into the implant body. Use an appropriate-sized “mini” screwdriver available at local hardware stores to remove the screw. A clinical report describes utilizing the non retrievable fractured abutment screw by fabricating the post/core. For this, modify the screw hole and fabricate a post crown after radiographically checking the osseo integration status of the implant.[64]

CONCLUSION

In the present review, in order to present a complete picture, various clinical as well as in vitro studies on the mechanics of implant fixture fracture and abutment screw loosening/fracture of a single implant-abutment connection have been reviewed. The incidence of such complications is very small and the literature is still not conclusive concerning the choice of connection and/or crown-retaining system. However, understanding the mechanics of these failures will help the clinicians in formulating a more predictable treatment.
  56 in total

1.  Implant abutment screw rotations and preloads for four different screw materials and surfaces.

Authors:  W C Martin; R D Woody; B H Miller; A W Miller
Journal:  J Prosthet Dent       Date:  2001-07       Impact factor: 3.426

Review 2.  Clinical complications of osseointegrated implants.

Authors:  C J Goodacre; J Y Kan; K Rungcharassaeng
Journal:  J Prosthet Dent       Date:  1999-05       Impact factor: 3.426

Review 3.  Implant screw mechanics and the settling effect: overview.

Authors:  Sheldon Winkler; Karla Ring; Jamie D Ring; Kenneth G Boberick
Journal:  J Oral Implantol       Date:  2003       Impact factor: 1.779

4.  Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants.

Authors:  E A Patterson; R B Johns
Journal:  Int J Oral Maxillofac Implants       Date:  1992       Impact factor: 2.804

5.  Screw loosening for standard and wide diameter implants in partially edentulous cases: 3- to 7-year longitudinal data.

Authors:  Sang-Choon Cho; Paula-Naomi Small; Nicolas Elian; Dennis Tarnow
Journal:  Implant Dent       Date:  2004-09       Impact factor: 2.454

6.  Periimplantitis.

Authors:  María Angeles Sánchez-Gárces; Cosme Gay-Escoda
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2004

7.  Tissue-integrated prosthesis complications.

Authors:  D E Tolman; W R Laney
Journal:  Int J Oral Maxillofac Implants       Date:  1992       Impact factor: 2.804

8.  Oral implant treatment in posterior partially edentulous jaws: a 5-year follow-up report.

Authors:  T Jemt; U Lekholm
Journal:  Int J Oral Maxillofac Implants       Date:  1993       Impact factor: 2.804

9.  Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws.

Authors:  Youssef S Al Jabbari; Raymond Fournelle; Gerald Ziebert; Jeffrey Toth; Anthony M Iacopino
Journal:  J Prosthodont       Date:  2007-11-15       Impact factor: 2.752

Review 10.  Etiology, risk factors and management of implant fractures.

Authors:  Arturo Sánchez-Pérez; Ma José Moya-Villaescusa; Alfonso Jornet-Garcia; Santiago Gomez
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2010-05-01
View more
  16 in total

1.  Complete mechanical characterization of an external hexagonal implant connection: in vitro study, 3D FEM, and probabilistic fatigue.

Authors:  María Prados-Privado; Sérgio A Gehrke; Rosa Rojo; Juan Carlos Prados-Frutos
Journal:  Med Biol Eng Comput       Date:  2018-06-11       Impact factor: 2.602

2.  Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study.

Authors:  Yuanyuan Duan; Jorge A Gonzalez; Pratim A Kulkarni; William W Nagy; Jason A Griggs
Journal:  Dent Mater       Date:  2018-06-18       Impact factor: 5.304

Review 3.  Removal of osseointegrated dental implants: a systematic review of explantation techniques.

Authors:  Margaux Roy; Lucie Loutan; Giovanni Garavaglia; Dena Hashim
Journal:  Clin Oral Investig       Date:  2019-11-15       Impact factor: 3.573

4.  Oral bacterial colonization on dental implants restored with titanium or zirconia abutments: 6-month follow-up.

Authors:  Alice Ramos de Freitas; Thalisson Saymo de Oliveira Silva; Ricardo Faria Ribeiro; Rubens Ferreira de Albuquerque Junior; Vinícius Pedrazzi; Cássio do Nascimento
Journal:  Clin Oral Investig       Date:  2018-01-18       Impact factor: 3.573

5.  Accidental displacement of a dental implant into the sublingual space: A Case report.

Authors:  Paolo Cariati; José Fernández-Solís; Ana-Belén Marín-Fernández; Alfredo Valencia-Laseca; Fernando Monsalve-Iglesias
Journal:  J Clin Exp Dent       Date:  2016-10-01

6.  Restoration: Implant with Devastated Platform through Metal Post.

Authors:  Luna Salinas Tatiana; Del Valle Lovato Juan
Journal:  Case Rep Dent       Date:  2017-02-13

7.  Principles of biomechanics in oral implantology.

Authors:  Avram Manea; Simion Bran; Cristian Dinu; Horatiu Rotaru; Ioan Barbur; Bogdan Crisan; Gabriel Armencea; Florin Onisor; Madalina Lazar; Daniel Ostas; Mihaela Baciut; Sergiu Vacaras; Ileana Mitre; Liana Crisan; Ovidiu Muresan; Rares Roman; Grigore Baciut
Journal:  Med Pharm Rep       Date:  2019-12-15

Review 8.  Complications in implant dentistry.

Authors:  Ayesha Hanif; Saima Qureshi; Zeeshan Sheikh; Haroon Rashid
Journal:  Eur J Dent       Date:  2017 Jan-Mar

9.  Comparative effect of implant-abutment connections, abutment angulations, and screw lengths on preloaded abutment screw using three-dimensional finite element analysis: An in vitro study.

Authors:  Krishna Chaitanya Kanneganti; Dileep Nag Vinnakota; Srinivas Rao Pottem; Mahesh Pulagam
Journal:  J Indian Prosthodont Soc       Date:  2018 Apr-Jun

10.  A novel technique to retrieve cold-welded implant abutment: A case series.

Authors:  Mahnaz Arshad; Hossein Ali Mahgoli; Kamran Rasouli; Sina Refoua
Journal:  Clin Case Rep       Date:  2019-08-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.