Literature DB >> 29921465

Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study.

Yuanyuan Duan1, Jorge A Gonzalez2, Pratim A Kulkarni1, William W Nagy3, Jason A Griggs4.   

Abstract

OBJECTIVE: To validate the fatigue lifetime of a reduced-diameter dental implant system predicted by three-dimensional finite element analysis (FEA) by testing physical implant specimens using an accelerated lifetime testing (ALT) strategy with the apparatus specified by ISO 14801.
METHODS: A commercially-available reduced-diameter titanium dental implant system (Straumann Standard Plus NN) was digitized using a micro-CT scanner. Axial slices were processed using an interactive medical image processing software (Mimics) to create 3D models. FEA analysis was performed in ABAQUS, and fatigue lifetime was predicted using fe-safe® software. The same implant specimens (n=15) were tested at a frequency of 2Hz on load frames using apparatus specified by ISO 14801 and ALT. Multiple step-stress load profiles with various aggressiveness were used to improve testing efficiency. Fatigue lifetime statistics of physical specimens were estimated in a reliability analysis software (ALTA PRO). Fractured specimens were examined using SEM with fractographic technique to determine the failure mode.
RESULTS: FEA predicted lifetime was within the 95% confidence interval of lifetime estimated by experimental results, which suggested that FEA prediction was accurate for this implant system. The highest probability of failure was located at the root of the implant body screw thread adjacent to the simulated bone level, which also agreed with the failure origin in physical specimens. SIGNIFICANCE: Fatigue lifetime predictions based on finite element modeling could yield similar results in lieu of physical testing, allowing the use of virtual testing in the early stages of future research projects on implant fatigue.
Copyright © 2018 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Accelerated lifetime testing; Brown-Miller criteria; Cyclic fatigue; Dental implants; Finite element modeling; Fractography; Plastic strain; Simulated bone; Titanium

Mesh:

Substances:

Year:  2018        PMID: 29921465      PMCID: PMC7302927          DOI: 10.1016/j.dental.2018.06.002

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  34 in total

1.  Three-dimensional finite element analysis of implant stability in the atrophic posterior maxilla: a mathematical study of the sinus floor augmentation.

Authors:  Gabor Tepper; Robert Haas; Werner Zechner; Wolfgang Krach; Georg Watzek
Journal:  Clin Oral Implants Res       Date:  2002-12       Impact factor: 5.977

2.  Influence of connector design on fracture probability of ceramic fixed-partial dentures.

Authors:  W Oh; N Götzen; K J Anusavice
Journal:  J Dent Res       Date:  2002-09       Impact factor: 6.116

3.  Design improvement and dynamic finite element analysis of novel ITI dental implant under dynamic chewing loads.

Authors:  Yung-Chang Cheng; Deng-Huei Lin; Cho-Pei Jiang; Shyh-Yuan Lee
Journal:  Biomed Mater Eng       Date:  2015       Impact factor: 1.300

4.  Finite element analysis of three designs of an implant-supported molar crown.

Authors:  Allahyar Geramy; Steven M Morgano
Journal:  J Prosthet Dent       Date:  2004-11       Impact factor: 3.426

5.  Bone strain and interfacial sliding analyses of platform switching and implant diameter on an immediately loaded implant: experimental and three-dimensional finite element analyses.

Authors:  Jui-Ting Hsu; Lih-Jyh Fuh; Dan-Jae Lin; Yen-Wen Shen; Heng-Li Huang
Journal:  J Periodontol       Date:  2009-07       Impact factor: 6.993

6.  Effect of elasticity on stress distribution in CAD/CAM dental crowns: Glass ceramic vs. polymer-matrix composite.

Authors:  Yuanyuan Duan; Jason A Griggs
Journal:  J Dent       Date:  2015-01-24       Impact factor: 4.379

7.  Young's modulus of trabecular and cortical bone material: ultrasonic and microtensile measurements.

Authors:  J Y Rho; R B Ashman; C H Turner
Journal:  J Biomech       Date:  1993-02       Impact factor: 2.712

8.  Analysis of incidence and associated factors with fractured implants: a retrospective study.

Authors:  S E Eckert; S J Meraw; E Cal; R K Ow
Journal:  Int J Oral Maxillofac Implants       Date:  2000 Sep-Oct       Impact factor: 2.804

Review 9.  Mechanical complications of dental implants.

Authors:  M S Schwarz
Journal:  Clin Oral Implants Res       Date:  2000       Impact factor: 5.977

10.  Mechanical properties and corrosion resistance of Ti-6Al-7Nb alloy dental castings.

Authors:  E Kobayashi; T J Wang; H Doi; T Yoneyama; H Hamanaka
Journal:  J Mater Sci Mater Med       Date:  1998-10       Impact factor: 3.896

View more
  6 in total

1.  Biomechanical evaluations of the long-term stability of dental implant using finite element modeling method: a systematic review.

Authors:  Seyed Aref Hosseini-Faradonbeh; Hamid Reza Katoozian
Journal:  J Adv Prosthodont       Date:  2022-06-27       Impact factor: 1.989

2.  Effect of loading frequency on cyclic fatigue lifetime of a standard-diameter implant with an internal abutment connection.

Authors:  Yuanyuan Duan; Jason A Griggs
Journal:  Dent Mater       Date:  2018-09-13       Impact factor: 5.304

3.  New Dental Implant with 3D Shock Absorbers and Tooth-Like Mobility-Prototype Development, Finite Element Analysis (FEA), and Mechanical Testing.

Authors:  Avram Manea; Grigore Baciut; Mihaela Baciut; Dumitru Pop; Dan Sorin Comsa; Ovidiu Buiga; Veronica Trombitas; Horatiu Colosi; Ileana Mitre; Roxana Bordea; Marius Manole; Manuela Lenghel; Simion Bran; Florin Onisor
Journal:  Materials (Basel)       Date:  2019-10-21       Impact factor: 3.623

4.  Comparative analysis of stress distribution in one-piece and two-piece implants with narrow and extra-narrow diameters: A finite element study.

Authors:  Fabricia Teixeira Barbosa; Luiz Carlos Silveira Zanatta; Edélcio de Souza Rendohl; Sergio Alexandre Gehrke
Journal:  PLoS One       Date:  2021-02-04       Impact factor: 3.240

5.  A Novel Design Method of Gradient Porous Structure for Stabilized and Lightweight Mandibular Prosthesis.

Authors:  Renshun Liu; Yuxiong Su; Weifa Yang; Kai Wu; Ruxu Du; Yong Zhong
Journal:  Bioengineering (Basel)       Date:  2022-08-30

6.  Load-Bearing Capacity of Zirconia Crowns Screwed to Multi-Unit Abutments with and without a Titanium Base: An In Vitro Pilot Study.

Authors:  Hadas Heller; Adi Arieli; Ilan Beitlitum; Raphael Pilo; Shifra Levartovsky
Journal:  Materials (Basel)       Date:  2019-09-20       Impact factor: 3.623

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.