Literature DB >> 7590997

Elongation and preload stress in dental implant abutment screws.

J E Haack1, R L Sakaguchi, T Sun, J P Coffey.   

Abstract

A common problem associated with dental implant restorations is loosening of screws that retain the prosthesis to the implant. A method was developed to determine initial preload on UCLA-type abutment screws by measuring elongation after applying known tightening torques with a digital torque gauge. Loosening torque was also measured after tightening to 32 N-cm torque for gold alloy abutment screws and 20 N-cm for titanium abutment screws. Gold alloy and titanium abutment screws were each used to secure a gold UCLA hexed abutment to a titanium implant. Stresses and forces were calculated from the elongation measurements for three regions of each screw. Elongation of the screws after applying the manufacturer's recommended tightening torques were within the elastic range. Induced stresses were 57.5% and 56% of the yield strengths for gold alloy and titanium, respectively. Tightening of screws beyond recommended levels may be possible without producing plastic deformation. At manufacturer's recommended torques, mean preload was 468.2 (+/- 57.9) N using gold alloy screws and 381.5 (+/- 72.9) N with titanium screws.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7590997

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  29 in total

1.  A Simple and Cost Effective Method used for Removal of a Fractured Implant Abutment Screw: A Case Report.

Authors:  Parth Satwalekar; K Subash Chander; B Anantha Reddy; N Sandeep; N Sandeep; Tanushree Satwalekar
Journal:  J Int Oral Health       Date:  2013-10-26

2.  The combined effect of dismantling for steam sterilization and aging on the accuracy of spring-style mechanical torque devices.

Authors:  Minoo Mahshid; Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayyaz; Mahdi Kadkhodazadeh
Journal:  J Periodontal Implant Sci       Date:  2013-10-29       Impact factor: 2.614

3.  Removal torque evaluation of three different abutment screws for single implant restorations after mechanical cyclic loading.

Authors:  T Paepoemsin; P A Reichart; P Chaijareenont; F P Strietzel; P Khongkhunthian
Journal:  Oral Implantol (Rome)       Date:  2016-11-16

4.  Effect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis.

Authors:  Seok-Gyu Kim; Chae-Heon Chung; Mee-Kyoung Son
Journal:  J Adv Prosthodont       Date:  2015-06-23       Impact factor: 1.904

5.  Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems.

Authors:  Jae-Young Jo; Dong-Seok Yang; Jung-Bo Huh; Jae-Chan Heo; Mi-Jung Yun; Chang-Mo Jeong
Journal:  J Adv Prosthodont       Date:  2014-12-17       Impact factor: 1.904

6.  Abrasion of abutment screw coated with TiN.

Authors:  Seok-Won Jung; Mee-Kyoung Son; Chae-Heon Chung; Hee-Jung Kim
Journal:  J Adv Prosthodont       Date:  2009-07-31       Impact factor: 1.904

7.  Preload and torque removal evaluation of three different abutment screws for single standing implant restorations.

Authors:  Rafael Augusto Stüker; Eduardo Rolim Teixeira; João Carlos Pinheiro Beck; Nilza Pereira da Costa
Journal:  J Appl Oral Sci       Date:  2008 Jan-Feb       Impact factor: 2.698

8.  Peri-implantitis and the prosthodontist.

Authors:  A Dawood; B Marti Marti; S Tanner
Journal:  Br Dent J       Date:  2017-09-08       Impact factor: 1.626

9.  Factors associated with abutment screw loosening in single implant supported crowns: A cross-sectional study.

Authors:  S M Londhe; E Mahesh Gowda; V B Mandlik; M P Shashidhar
Journal:  Med J Armed Forces India       Date:  2018-10-10

10.  Influence of the implant abutment types and the dynamic loading on initial screw loosening.

Authors:  Eun-Sook Kim; Soo-Yeon Shin
Journal:  J Adv Prosthodont       Date:  2013-02-28       Impact factor: 1.904

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.