A piecewise-focused high DQE detector for MV imaging.

Josh Star-Lack1, Daniel Shedlock1, Dennis Swahn2, Dave Humber1, Adam Wang1, Hayley Hirsh1, George Zentai1, Daren Sawkey1, Isaac Kruger1, Mingshan Sun1, Eric Abel1, Gary Virshup1, Mihye Shin3, Rebecca Fahrig3.   

Abstract

PURPOSE: Electronic portal imagers (EPIDs) with high detective quantum efficiencies (DQEs) are sought to facilitate the use of the megavoltage (MV) radiotherapy treatment beam for image guidance. Potential advantages include high quality (treatment) beam's eye view imaging, and improved cone-beam computed tomography (CBCT) generating images with more accurate electron density maps with immunity to metal artifacts. One approach to increasing detector sensitivity is to couple a thick pixelated scintillator array to an active matrix flat panel imager (AMFPI) incorporating amorphous silicon thin film electronics. Cadmium tungstate (CWO) has many desirable scintillation properties including good light output, a high index of refraction, high optical transparency, and reasonable cost. However, due to the 0 1 0 cleave plane inherent in its crystalline structure, the difficulty of cutting and polishing CWO has, in part, limited its study relative to other scintillators such as cesium iodide and bismuth germanate (BGO). The goal of this work was to build and test a focused large-area pixelated "strip" CWO detector.
METHODS: A 361 × 52 mm scintillator assembly that contained a total of 28 072 pixels was constructed. The assembly comprised seven subarrays, each 15 mm thick. Six of the subarrays were fabricated from CWO with a pixel pitch of 0.784 mm, while one array was constructed from BGO for comparison. Focusing was achieved by coupling the arrays to the Varian AS1000 AMFPI through a piecewise linear arc-shaped fiber optic plate. Simulation and experimental studies of modulation transfer function (MTF) and DQE were undertaken using a 6 MV beam, and comparisons were made between the performance of the pixelated strip assembly and the most common EPID configuration comprising a 1 mm-thick copper build-up plate attached to a 133 mg/cm(2) gadolinium oxysulfide scintillator screen (Cu-GOS). Projection radiographs and CBCT images of phantoms were acquired. The work also introduces the use of a lightweight edge phantom to generate MTF measurements at MV energies and shows its functional equivalence to the more cumbersome slit-based method.
RESULTS: Measured and simulated DQE(0)'s of the pixelated CWO detector were 22% and 26%, respectively. The average measured and simulated ratios of CWO DQE(f) to Cu-GOS DQE(f) across the frequency range of 0.0-0.62 mm(-1) were 23 and 29, respectively. 2D and 3D imaging studies confirmed the large dose efficiency improvement and that focus was maintained across the field of view. In the CWO CBCT images, the measured spatial resolution was 7 lp/cm. The contrast-to-noise ratio was dramatically improved reflecting a 22 × sensitivity increase relative to Cu-GOS. The CWO scintillator material showed significantly higher stability and light yield than the BGO material.
CONCLUSIONS: An efficient piecewise-focused pixelated strip scintillator for MV imaging is described that offers more than a 20-fold dose efficiency improvement over Cu-GOS.

Entities:  

Mesh:

Year:  2015        PMID: 26328960      PMCID: PMC4529442          DOI: 10.1118/1.4927786

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  36 in total

1.  Monte Carlo simulations of the imaging performance of metal plate/phosphor screens used in radiotherapy.

Authors:  C Kausch; B Schreiber; F Kreuder; R Schmidt; O Dössel
Journal:  Med Phys       Date:  1999-10       Impact factor: 4.071

2.  Megavoltage cone-beam computed tomography using a high-efficiency image receptor.

Authors:  Ed J Seppi; Peter Munro; Stan W Johnsen; Ed G Shapiro; Carlo Tognina; Dan Jones; John M Pavkovich; Chris Webb; Ivan Mollov; Larry D Partain; Rick E Colbeth
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-03-01       Impact factor: 7.038

3.  Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit.

Authors:  Sanford L Meeks; Joseph F Harmon; Katja M Langen; Twyla R Willoughby; Thomas H Wagner; Patrick A Kupelian
Journal:  Med Phys       Date:  2005-08       Impact factor: 4.071

4.  A bench-top megavoltage fan-beam CT using CdWO4-photodiode detectors. II. Image performance evaluation.

Authors:  T T Monajemi; D Tu; B G Fallone; S Rathee
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

5.  Slit design for efficient and accurate MTF measurement at megavoltage x-ray energies.

Authors:  Amit Sawant; Larry Antonuk; Youcef El-Mohri
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

6.  A simple method for determining the modulation transfer function in digital radiography.

Authors:  H Fujita; D Y Tsai; T Itoh; K Doi; J Morishita; K Ueda; A Ohtsuka
Journal:  IEEE Trans Med Imaging       Date:  1992       Impact factor: 10.048

7.  A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

8.  A method for measuring the presampled MTF of digital radiographic systems using an edge test device.

Authors:  E Samei; M J Flynn; D A Reimann
Journal:  Med Phys       Date:  1998-01       Impact factor: 4.071

9.  X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays.

Authors:  P Munro; D C Bouius
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

10.  Optimal short scan convolution reconstruction for fanbeam CT.

Authors:  D L Parker
Journal:  Med Phys       Date:  1982 Mar-Apr       Impact factor: 4.071

View more
  11 in total

1.  Super-resolution imaging in a multiple layer EPID.

Authors:  Haijian Chen; Joerg Rottmann; Stephen Sf Yip; Daniel Morf; Rony Füglistaller; Josh Star-Lack; George Zentai; Ross Berbeco
Journal:  Biomed Phys Eng Express       Date:  2017-02-21

2.  Multi-layer imager design for mega-voltage spectral imaging.

Authors:  Marios Myronakis; Yue-Houng Hu; Rony Fueglistaller; Adam Wang; Paul Baturin; Pascal Huber; Daniel Morf; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-05-10       Impact factor: 3.609

3.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

4.  A novel EPID design for enhanced contrast and detective quantum efficiency.

Authors:  Joerg Rottmann; Daniel Morf; Rony Fueglistaller; George Zentai; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

5.  A Monte Carlo study of the impact of phosphor optical properties on EPID imaging performance.

Authors:  Mengying Shi; Marios Myronakis; Yue-Houng Hu; Daniel Morf; Joerg Rottmann; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-08-20       Impact factor: 3.609

6.  A novel multilayer MV imager computational model for component optimization.

Authors:  Marios Myronakis; Josh Star-Lack; Paul Baturin; Joerg Rottmann; Daniel Morf; Adam Wang; Yue-Houng Hu; Daniel Shedlock; Ross I Berbeco
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

7.  A novel method for quantification of beam's-eye-view tumor tracking performance.

Authors:  Yue-Houng Hu; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Morf; Daniel Shedlock; Paul Baturin; Josh Star-Lack; Ross Berbeco
Journal:  Med Phys       Date:  2017-10-13       Impact factor: 4.071

8.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174

9.  A rapid, accurate image simulation strategy for mega-voltage cone-beam computed tomography.

Authors:  Mengying Shi; Marios Myronakis; Matthew Jacobson; Mathias Lehmann; Dianne Ferguson; Paul Baturin; Pascal Huber; Rony Fueglistaller; Thomas Harris; Ingrid Valencia Lozano; Christopher Williams; Daniel Morf; Ross I Berbeco
Journal:  Phys Med Biol       Date:  2020-07-06       Impact factor: 4.174

10.  Simulated performances of pixelated CsI(Tl) scintillation screens with different micro-column shapes and array structures in X-ray imaging.

Authors:  Hui Chen; Mu Gu; Xiaolin Liu; Juannan Zhang; Bo Liu; Shiming Huang; Chen Ni
Journal:  Sci Rep       Date:  2018-11-14       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.