Literature DB >> 27494207

A novel EPID design for enhanced contrast and detective quantum efficiency.

Joerg Rottmann1, Daniel Morf, Rony Fueglistaller, George Zentai, Josh Star-Lack, Ross Berbeco.   

Abstract

Beams-eye-view imaging applications such as real-time soft-tissue motion estimation are hindered by the inherently low image contrast of electronic portal imaging devices (EPID) currently available for clinical use. We introduce and characterize a novel EPID design that provides substantially increased detective quantum efficiency (DQE), contrast-to-noise ratio (CNR) and sensitivity without degradation in spatial resolution. The prototype design features a stack of four conventional EPID layers combined with low noise integrated readout electronics. Each layer consists of a copper plate, a scintillator ([Formula: see text]) and a photodiode/TFT-switch (aSi:H). We characterize the prototype's signal response to a 6 MV photon beam in terms of modulation transfer function (MTF), DQE and CNR. The presampled MTF is estimated using a slanted slit technique, the DQE is calculated from measured normalized noise power spectra (nNPS) and the MTF and CNR is estimated using a Las Vegas contrast phantom. The prototype has been designed and built to be interchangeable with the current clinical EPID on the Varian TrueBeam platform (AS-1200) in terms of size and data output specifications. Performance evaluation is conducted in absolute values as well as in relative terms using the Varian AS-1200 EPID as a reference detector. A fivefold increase of DQE(0) to about 6.7% was observed by using the four-layered design versus the AS-1200 reference detector. No substantial differences are observed between each layer's individual MTF and the one for all four layers operating combined indicating that defocusing due to beam divergence is negligible. Also, using four layers instead of one increases the signal to noise ratio by a factor of 1.7.

Entities:  

Mesh:

Year:  2016        PMID: 27494207      PMCID: PMC5525153          DOI: 10.1088/0031-9155/61/17/6297

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  17 in total

1.  Clinical use of electronic portal imaging: report of AAPM Radiation Therapy Committee Task Group 58.

Authors:  M G Herman; J M Balter; D A Jaffray; K P McGee; P Munro; S Shalev; M Van Herk; J W Wong
Journal:  Med Phys       Date:  2001-05       Impact factor: 4.071

2.  Segmented phosphors: MEMS-based high quantum efficiency detectors for megavoltage x-ray imaging.

Authors:  Amit Sawant; Larry E Antonuk; Youcef El-Mohri; Yixin Li; Zhong Su; Yi Wang; Jin Yamamoto; Qihua Zhao; Hong Du; Jurgen Daniel; Robert Street
Journal:  Med Phys       Date:  2005-02       Impact factor: 4.071

3.  Segmented crystalline scintillators: an initial investigation of high quantum efficiency detectors for megavoltage x-ray imaging.

Authors:  Amit Sawant; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Yixin Li; Zhong Su; Yi Wang; Jin Yamamoto; Hong Du; Ian Cunningham; Misha Klugerman; Kanai Shah
Journal:  Med Phys       Date:  2005-10       Impact factor: 4.071

4.  Slit design for efficient and accurate MTF measurement at megavoltage x-ray energies.

Authors:  Amit Sawant; Larry Antonuk; Youcef El-Mohri
Journal:  Med Phys       Date:  2007-05       Impact factor: 4.071

5.  A simple method for determining the modulation transfer function in digital radiography.

Authors:  H Fujita; D Y Tsai; T Itoh; K Doi; J Morishita; K Ueda; A Ohtsuka
Journal:  IEEE Trans Med Imaging       Date:  1992       Impact factor: 10.048

6.  X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays.

Authors:  P Munro; D C Bouius
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

7.  Low-dose megavoltage cone-beam CT for radiation therapy.

Authors:  Jean Pouliot; Ali Bani-Hashemi; Josephine Chen; Michelle Svatos; Farhad Ghelmansarai; Matthias Mitschke; Michele Aubin; Ping Xia; Olivier Morin; Kara Bucci; Mack Roach; Paco Hernandez; Zirao Zheng; Dimitre Hristov; Lynn Verhey
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-02-01       Impact factor: 7.038

8.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

9.  The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking.

Authors:  Stephen Yip; Joerg Rottmann; Ross Berbeco
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

10.  Countering beam divergence effects with focused segmented scintillators for high DQE megavoltage active matrix imagers.

Authors:  Langechuan Liu; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Hao Jiang
Journal:  Phys Med Biol       Date:  2012-08-01       Impact factor: 3.609

View more
  21 in total

1.  Super-resolution imaging in a multiple layer EPID.

Authors:  Haijian Chen; Joerg Rottmann; Stephen Sf Yip; Daniel Morf; Rony Füglistaller; Josh Star-Lack; George Zentai; Ross Berbeco
Journal:  Biomed Phys Eng Express       Date:  2017-02-21

2.  Multi-layer imager design for mega-voltage spectral imaging.

Authors:  Marios Myronakis; Yue-Houng Hu; Rony Fueglistaller; Adam Wang; Paul Baturin; Pascal Huber; Daniel Morf; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-05-10       Impact factor: 3.609

3.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

4.  Evaluation of a hybrid direct-indirect active matrix flat-panel imager using Monte Carlo simulation.

Authors:  Scott Dow; Adrian Howansky; Anthony R Lubinsky; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2020-05-12

5.  Low-dose megavoltage cone-beam computed tomography using a novel multi-layer imager (MLI).

Authors:  Marios Myronakis; Pascal Huber; Mathias Lehmann; Rony Fueglistaller; Matthew Jacobson; Yue-Houng Hu; Paul Baturin; Adam Wang; Mengying Shi; Thomas Harris; Daniel Morf; Ross Berbeco
Journal:  Med Phys       Date:  2020-01-28       Impact factor: 4.071

6.  Leveraging multi-layer imager detector design to improve low-dose performance for megavoltage cone-beam computed tomography.

Authors:  Yue-Houng Hu; Joerg Rottmann; Rony Fueglistaller; Marios Myronakis; Adam Wang; Pascal Huber; Daniel Shedlock; Daniel Morf; Paul Baturin; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-01-30       Impact factor: 3.609

7.  A Monte Carlo study of the impact of phosphor optical properties on EPID imaging performance.

Authors:  Mengying Shi; Marios Myronakis; Yue-Houng Hu; Daniel Morf; Joerg Rottmann; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-08-20       Impact factor: 3.609

8.  A novel multilayer MV imager computational model for component optimization.

Authors:  Marios Myronakis; Josh Star-Lack; Paul Baturin; Joerg Rottmann; Daniel Morf; Adam Wang; Yue-Houng Hu; Daniel Shedlock; Ross I Berbeco
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

9.  A novel method for quantification of beam's-eye-view tumor tracking performance.

Authors:  Yue-Houng Hu; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Morf; Daniel Shedlock; Paul Baturin; Josh Star-Lack; Ross Berbeco
Journal:  Med Phys       Date:  2017-10-13       Impact factor: 4.071

10.  Spectral imaging using clinical megavoltage beams and a novel multi-layer imager.

Authors:  Marios Myronakis; Rony Fueglistaller; Joerg Rottmann; Yue-Houng Hu; Adam Wang; Paul Baturin; Pascal Huber; Daniel Morf; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2017-11-14       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.