Megavoltage cone-beam computed tomography using a high-efficiency image receptor.

Ed J Seppi1, Peter Munro, Stan W Johnsen, Ed G Shapiro, Carlo Tognina, Dan Jones, John M Pavkovich, Chris Webb, Ivan Mollov, Larry D Partain, Rick E Colbeth.   

Abstract

PURPOSE: To develop an image receptor capable of forming high-quality megavoltage CT images using modest radiation doses. METHODS AND MATERIALS: A flat-panel imaging system consisting of a conventional flat-panel sensor attached to a thick CsI scintillator has been fabricated. The scintillator consists of individual CsI crystals 8 mm thick by 0.38 mm x 0.38-mm pitch. Five sides of each crystal are coated with a reflecting powder/epoxy mixture, and the sixth side is in contact with the flat-panel sensor. A timing interface coordinates acquisition by the imaging system and pulsing of the linear accelerator. With this interface, as little as one accelerator pulse (0.023 cGy at the isocenter) can be used to form projection images. Different CT phantoms irradiated by a 6-MV X-ray beam have been imaged to evaluate the performance of the imaging system. The phantoms have been mounted on a rotating stage and rotated while 360 projection images are acquired in 48 s. These projections have been reconstructed using the Feldkamp cone-beam CT reconstruction algorithm. RESULTS AND DISCUSSION: Using an irradiation of 16 cGy (360 projections x 0.046 cGy/projection), the contrast resolution is approximately 1% for large objects. High-contrast structures as small as 1.2 mm are clearly visible. The reconstructed CT values are linear (R(2) = 0.98) for electron densities between 0.001 and 2.16 g/cm(3), and the reconstruction time for a 512 x 512 x 512 data set is 6 min. Images of an anthropomorphic phantom show that soft-tissue structures such as the heart, lung, kidneys, and liver are visible in the reconstructed images (16 cGy, 5-mm-thick slices).
CONCLUSIONS: The acquisition of megavoltage CT images with soft-tissue contrast is possible with irradiations as small as 16 cGy.

Mesh:

Year:  2003        PMID: 12573767     DOI: 10.1016/s0360-3016(02)04155-x

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  21 in total

1.  Performance evaluation of polycrystalline HgI2 photoconductors for radiation therapy imaging.

Authors:  Qihua Zhao; Larry E Antonuk; Youcef El-Mohri; Yi Wang; Hong Du; Amit Sawant; Zhong Su; Jin Yamamoto
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

2.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

3.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  A Monte Carlo investigation of Swank noise for thick, segmented, crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

5.  Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Authors:  Meng Wu; Andreas Keil; Dragos Constantin; Josh Star-Lack; Lei Zhu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

6.  Monte Carlo investigations of the effect of beam divergence on thick, segmented crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Youcef El-Mohri; Larry E Antonuk; Qihua Zhao
Journal:  Phys Med Biol       Date:  2010-06-04       Impact factor: 3.609

7.  Optimization of the performance of segmented scintillators for radiotherapy imaging through novel binning techniques.

Authors:  Youcef El-Mohri; Larry E Antonuk; Richard B Choroszucha; Qihua Zhao; Hao Jiang; Langechuan Liu
Journal:  Phys Med Biol       Date:  2014-02-03       Impact factor: 3.609

8.  Characterization of linear accelerator X-ray source size using a laminated beam-spot camera.

Authors:  Collins Yeboah
Journal:  J Appl Clin Med Phys       Date:  2011-05-10       Impact factor: 2.102

9.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

10.  Investigation of sagittal image acquisition for 4D-MRI with body area as respiratory surrogate.

Authors:  Yilin Liu; Fang-Fang Yin; Zheng Chang; Brian G Czito; Manisha Palta; Mustafa R Bashir; Yujiao Qin; Jing Cai
Journal:  Med Phys       Date:  2014-10       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.