A novel method for quantification of beam's-eye-view tumor tracking performance.

Yue-Houng Hu1, Marios Myronakis1, Joerg Rottmann1, Adam Wang2, Daniel Morf3, Daniel Shedlock2, Paul Baturin2, Josh Star-Lack2, Ross Berbeco1.   

Abstract

PURPOSE: In-treatment imaging using an electronic portal imaging device (EPID) can be used to confirm patient and tumor positioning. Real-time tumor tracking performance using current digital megavolt (MV) imagers is hindered by poor image quality. Novel EPID designs may help to improve quantum noise response, while also preserving the high spatial resolution of the current clinical detector. Recently investigated EPID design improvements include but are not limited to multi-layer imager (MLI) architecture, thick crystalline and amorphous scintillators, and phosphor pixilation and focusing. The goal of the present study was to provide a method of quantitating improvement in tracking performance as well as to reveal the physical underpinnings of detector design that impact tracking quality. The study employs a generalizable ideal observer methodology for the quantification of tumor tracking performance. The analysis is applied to study both the effect of increasing scintillator thickness on a standard, single-layer imager (SLI) design as well as the effect of MLI architecture on tracking performance.
METHODS: The present study uses the ideal observer signal-to-noise ratio (d') as a surrogate for tracking performance. We employ functions which model clinically relevant tasks and generalized frequency-domain imaging metrics to connect image quality with tumor tracking. A detection task for relevant Cartesian shapes (i.e., spheres and cylinders) was used to quantitate trackability of cases employing fiducial markers. Automated lung tumor tracking algorithms often leverage the differences in benign and malignant lung tissue textures. These types of algorithms (e.g., soft-tissue localization - STiL) were simulated by designing a discrimination task, which quantifies the differentiation of tissue textures, measured experimentally and fit as a power-law in trend (with exponent β) using a cohort of MV images of patient lungs. The modeled MTF and NPS were used to investigate the effect of scintillator thickness and MLI architecture on tumor tracking performance.
RESULTS: Quantification of MV images of lung tissue as an inverse power-law with respect to frequency yields exponent values of β = 3.11 and 3.29 for benign and malignant tissues, respectively. Tracking performance with and without fiducials was found to be generally limited by quantum noise, a factor dominated by quantum detective efficiency (QDE). For generic SLI construction, increasing the scintillator thickness (gadolinium oxysulfide - GOS) from a standard 290 μm to 1720 μm reduces noise to about 10%. However, 81% of this reduction is appreciated between 290 and 1000 μm. In comparing MLI and SLI detectors of equivalent individual GOS layer thickness, the improvement in noise is equal to the number of layers in the detector (i.e., 4) with almost no difference in MTF. Further, improvement in tracking performance was slightly less than the square-root of the reduction in noise, approximately 84-90%. In comparing an MLI detector with an SLI with a GOS scintillator of equivalent total thickness, improvement in object detectability is approximately 34-39%.
CONCLUSIONS: We have presented a novel method for quantification of tumor tracking quality and have applied this model to evaluate the performance of SLI and MLI EPID designs. We showed that improved tracking quality is primarily limited by improvements in NPS. When compared to very thick scintillator SLI, employing MLI architecture exhibits the same gains in QDE, but by mitigating the effect of optical Swank noise, results in more dramatic improvements in tracking performance.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  zzm321990EPIDzzm321990; zzm321990MTFzzm321990; zzm321990NPSzzm321990; ideal observer; multi-layer detector; portal imaging

Mesh:

Year:  2017        PMID: 28887836      PMCID: PMC5689096          DOI: 10.1002/mp.12572

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  37 in total

1.  Safety and efficacy of percutaneous fiducial marker implantation for image-guided radiation therapy.

Authors:  Nishita Kothary; Jeremy J Heit; John D Louie; William T Kuo; Billy W Loo; Albert Koong; Daniel T Chang; David Hovsepian; Daniel Y Sze; Lawrence V Hofmann
Journal:  J Vasc Interv Radiol       Date:  2008-11-18       Impact factor: 3.464

2.  Anatomical noise in contrast-enhanced digital mammography. Part II. Dual-energy imaging.

Authors:  Melissa L Hill; James G Mainprize; Ann-Katherine Carton; Sylvie Saab-Puong; Razvan Iordache; Serge Muller; Roberta A Jong; Clarisse Dromain; Martin J Yaffe
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

3.  Optimization of contrast-enhanced breast imaging: Analysis using a cascaded linear system model.

Authors:  Yue-Houng Hu; David A Scaduto; Wei Zhao
Journal:  Med Phys       Date:  2017-01-03       Impact factor: 4.071

4.  Characterisation of mammographic parenchymal pattern by fractal dimension.

Authors:  C B Caldwell; S J Stapleton; D W Holdsworth; R A Jong; W J Weiser; G Cooke; M J Yaffe
Journal:  Phys Med Biol       Date:  1990-02       Impact factor: 3.609

5.  Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms: FBP and SAFIRE.

Authors:  Justin Solomon; Ehsan Samei
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

6.  A novel multilayer MV imager computational model for component optimization.

Authors:  Marios Myronakis; Josh Star-Lack; Paul Baturin; Joerg Rottmann; Daniel Morf; Adam Wang; Yue-Houng Hu; Daniel Shedlock; Ross I Berbeco
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

7.  X-ray imaging performance of structured cesium iodide scintillators.

Authors:  Wei Zhao; Goran Ristic; J A Rowlands
Journal:  Med Phys       Date:  2004-09       Impact factor: 4.071

8.  Association between power law coefficients of the anatomical noise power spectrum and lesion detectability in breast imaging modalities.

Authors:  Lin Chen; Craig K Abbey; John M Boone
Journal:  Phys Med Biol       Date:  2013-02-19       Impact factor: 3.609

9.  Texture analysis of aggressive and nonaggressive lung tumor CE CT images.

Authors:  Omar S Al-Kadi; D Watson
Journal:  IEEE Trans Biomed Eng       Date:  2008-07       Impact factor: 4.538

10.  Clinical feasibility of using an EPID in CINE mode for image-guided verification of stereotactic body radiotherapy.

Authors:  Ross I Berbeco; Fred Hacker; Dan Ionascu; Harvey J Mamon
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-01       Impact factor: 7.038

View more
  6 in total

1.  Physics considerations in MV-CBCT multi-layer imager design.

Authors:  Yue-Houng Hu; Rony Fueglistaller; Marios Myronakis; Joerg Rottmann; Adam Wang; Daniel Shedlock; Daniel Morf; Paul Baturin; Pascal Huber; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-06-20       Impact factor: 3.609

2.  Low-dose megavoltage cone-beam computed tomography using a novel multi-layer imager (MLI).

Authors:  Marios Myronakis; Pascal Huber; Mathias Lehmann; Rony Fueglistaller; Matthew Jacobson; Yue-Houng Hu; Paul Baturin; Adam Wang; Mengying Shi; Thomas Harris; Daniel Morf; Ross Berbeco
Journal:  Med Phys       Date:  2020-01-28       Impact factor: 4.071

3.  Leveraging multi-layer imager detector design to improve low-dose performance for megavoltage cone-beam computed tomography.

Authors:  Yue-Houng Hu; Joerg Rottmann; Rony Fueglistaller; Marios Myronakis; Adam Wang; Pascal Huber; Daniel Shedlock; Daniel Morf; Paul Baturin; Josh Star-Lack; Ross Berbeco
Journal:  Phys Med Biol       Date:  2018-01-30       Impact factor: 3.609

4.  Frequency-dependent optimal weighting approach for megavoltage multilayer imagers.

Authors:  Ingrid Valencia Lozano; Mengying Shi; Marios Myronakis; Paul Baturin; Rony Fueglistaller; Pascal Huber; Mathias Lehmann; Daniel Morf; Dianne Ferguson; Matthew W Jacobson; Thomas Harris; Ross I Berbeco; Christopher L Williams
Journal:  Phys Med Biol       Date:  2021-04-16       Impact factor: 4.174

5.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174

6.  A new registration algorithm of electronic portal imaging devices images based on the automatic detection of bone edges during radiotherapy.

Authors:  Cheng Chen; Chaoyan Wu; Yahua Zhong; Conghua Xie; Yunfeng Zhou; Hui Liu; Jun Zhang; Jiuling Sheng; Dazheng Jiang; Hongli Zhao; Haijun Yu
Journal:  Sci Rep       Date:  2020-06-24       Impact factor: 4.379

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.