| Literature DB >> 25342984 |
Antoine Berberi1, George Tehini1, Khaldoun Rifai1, Farah Bou Nasser Eddine2, Bassam Badran2, Haidar Akl2.
Abstract
Leakage has been addressed as a major contributing factor to inflammatory reactions at the implant-abutment connection, leading to problems such as oral malodor, inflammation, and marginal bone loss. The aim of this study was to investigate in vitro the leakage at implant-abutment interface of OsseoSpeed™ implants connected to original and compatible abutments. A total of 28 OsseoSpeed implants were divided into four groups (n = 7). Each group was connected to four different abutments according to manufacturers' recommendations: group A (TiDesign™); group B (Natea™); group C (Dual™); and group D (Implanet™) abutments. The inner volume of each implant-abutment combination was calculated and leakage was detected for each group with spectrophotometric analysis at 1 h (D0) and 48 h (D1) of incubation time using Rhodamine B. At 1 h, leakage volume was significantly lower in TiDesign and Dual than in Natea and Implanet (P < 0.001). At 48 h, however, leakage was significantly lower between TiDesign and all other systems (P < 0.005). Compatible abutments do not fit internal connection of OsseoSpeed implants perfectly, which increases the leakage of the final assembly.Entities:
Keywords: Dental implants; Rhodamine B; abutment connection; compatible abutment; implant–abutment interface; leakage
Year: 2014 PMID: 25342984 PMCID: PMC4206669 DOI: 10.1177/1758736014547143
Source DB: PubMed Journal: J Dent Biomech ISSN: 1758-7360
Group combination with implants and abutments size, type.
| Group | Size (mm) | Type | Implant | Abutment | ||
|---|---|---|---|---|---|---|
| Manufacturer | Size (mm) | Type | Manufacturer | |||
| A | 5 × 11 | OsseoSpeed™ | Astra Tech Implant System™, Mölndal, Sweden | 5.5 × 1.5 | TiDesign™ | Astra Tech Implant System |
| B | 5 × 11 | OsseoSpeed | Astra Tech Implant System | 5.8 × 2 | Natea™ | Euroteknika™ Groupe, Sallanches, France |
| C | 5 × 11 | OsseoSpeed | Astra Tech Implant System | 5.5 × 1.5 | Dual™ | Implantium, Dentium Implant System, Seoul, South Korea |
| D | 5 × 11 | OsseoSpeed | Astra Tech Implant System | 5.5 × 1.5 | Implanet™ | Derig LTDA, Sao Paulo, Brazil |
Figure 1.(a–d) Scanning electron microscope pictures showing the marginal gap of each combination.
Figure 2.Placement of Rhodamine B solution inside the implant.
Figure 3.Calibration curve was determined by linear regression considering the fluorescence (R2 = 0.9948).
Mean and standard deviation (SD) of the leakage volumes for each system at D0 and D1 (P < 0.05).
| OsseoSpeed™ + TiDesign™ | OsseoSpeed + Natea™ | OsseoSpeed + Dual™ | OsseoSpeed + Implanet™ | |||||
|---|---|---|---|---|---|---|---|---|
| Mean (µL) | SD (µL) | Mean (µL) | SD (µL) | Mean (µL) | SD (µL) | Mean (µL) | SD (µL) | |
| D0 | 0.0432 | 0.0022 | 0.2768 | 0.0382 | 0.1237 | 0.0116 | 0.4810 | 0.0625 |
| D1 | 0.0792 | 0.0074 | 0.3818 | 0.0192 | 0.2008 | 0.0193 | 0.6196 | 0.0725 |
SD: standard deviation.
Percentage of leakage of each group at D0 and D1 with percentages of fold increase.
| OsseoSpeed + TiDesign™ | OsseoSpeed + Natea™ | OsseoSpeed + Dual™ | OsseoSpeed + Implanet™ | ||||
|---|---|---|---|---|---|---|---|
| Percentage | Percentage | % Fold increase | Percentage | % Fold increase | Percentage | % Fold increase | |
| D0 | 1.484631 | 27.92083 | 640.5166 | 10.59397 | 286.2503 | 51.03048 | 1113.0185 |
| D1 | 5.558788 | 39.80301 | 481.9738 | 19.3194 | 253.4914 | 66.71397 | 782.1495 |
Figure 4.Measurements of the inner volumes of each combination.
Figure 5.At D0 and D1, the leakage volume of each system is shown. According to the Prism convention of significance, results were considered highly significant: ***P ≤ 0.001; ****P ≤ 0.0001.