Implants with original and non-original abutment connections.

Michel Gigandet1, Gianni Bigolin, Francisco Faoro, Walter Bürgin, Urs Brägger.   

Abstract

AIM: To test in vitro the mechanical resistance, rotational misfit and failure mode of three original implant-abutment connections and to compare them to two connections between non-original abutments connected to one of the original implants.
MATERIAL AND METHODS: Three different implants with small diameters (3.3 mm for Straumann Roxolid, 3.5 mm for Nobel Biocare Replace and Astra Tech Osseospeed TX) were connected with individualized titanium abutments. Twelve implants from each system were connected to their original abutments (Straumann CARES, Nobel Biocare Procera, Astra Tech Atlantis). Twenty-four Roxolid implants were connected with non-original abutments using CAD/CAM procedures from the other two manufacturers (12 Nobel Biocare Procera and 12 Astra Tech Atlantis). For the critical bending test, a Zwick/Roell 1475 machine and the Xpert Zwick/Roell software were used.
RESULTS: The rotational misfit varied when comparing the different interfaces. The use of non-original grade V titanium abutments on Roxolid implants increased the force needed for deformation. The fracture mode was different with one of the original connections.
CONCLUSIONS: Non-original abutments differ in design of the connecting surfaces and material and demonstrate higher rotational misfit. These differences may result in unexpected failure modes.
© 2012 Wiley Periodicals, Inc.

Entities:  

Keywords:  CAD/CAM; ISO 14801; abutment connection; abutments; dental implants; non-original; original; titanium grade IV; titanium grade V

Mesh:

Substances:

Year:  2012        PMID: 22804848     DOI: 10.1111/j.1708-8208.2012.00479.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  8 in total

1.  A comparative biomechanical study of original and compatible titanium bases: evaluation of screw loosening and 3D-crown displacement following cyclic loading analysis.

Authors:  Rimantas Ožiūnas; Jurgina Sakalauskienė; Darius Jegelevičius; Gintaras Janužis
Journal:  J Adv Prosthodont       Date:  2022-04-27       Impact factor: 1.989

2.  Leakage evaluation of original and compatible implant-abutment connections: In vitro study using Rhodamine B.

Authors:  Antoine Berberi; George Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Bassam Badran; Haidar Akl
Journal:  J Dent Biomech       Date:  2014-08-11

3.  Mechanical Outcomes, Microleakage, and Marginal Accuracy at the Implant-Abutment Interface of Original versus Nonoriginal Implant Abutments: A Systematic Review of In Vitro Studies.

Authors:  Marco Tallarico; Joseph Fiorellini; Yasushi Nakajima; Yuki Omori; Iida Takahisa; Luigi Canullo
Journal:  Biomed Res Int       Date:  2018-12-30       Impact factor: 3.411

4.  Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit.

Authors:  Jorge Vélez; Jesús Peláez; Carlos López-Suárez; Rubén Agustín-Panadero; Celia Tobar; María J Suárez
Journal:  J Clin Med       Date:  2020-07-24       Impact factor: 4.241

5.  An In-Vitro Analysis of Peri-Implant Mucosal Seal Following Photofunctionalization of Zirconia Abutment Materials.

Authors:  Masfueh Razali; Wei Cheong Ngeow; Ros Anita Omar; Wen Lin Chai
Journal:  Biomedicines       Date:  2021-01-15

6.  Comparing the Fracture Resistance and Modes of Failure in Different Types of CAD/CAM Zirconia Abutments with Internal Hexagonal Implants: An In Vitro Study.

Authors:  Yu-Tsen Chang; Yu-Ling Wu; Hung-Shyong Chen; Ming-Hsu Tsai; Chia-Chen Chang; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-04-04       Impact factor: 3.623

7.  Effect of Marginal Bone Integrity and Aftermarket Abutment Screws on Dental Implant Systems-A Preliminary Study with Finite Element Method.

Authors:  Yu-Ling Wu; Ming-Hsu Tsai; Hung-Shyong Chen; Ching-Ping Lin; Aaron Yu-Jen Wu
Journal:  Materials (Basel)       Date:  2022-08-28       Impact factor: 3.748

8.  Biomechanical effects of original equipment manufacturer and aftermarket abutment screws in zirconia abutment on dental implant assembly.

Authors:  Yu-Ling Wu; Ming-Hsu Tsai; Hung-Shyong Chen; Yu-Tsen Chang; Tsai-Te Lin; Aaron Yu-Jen Wu
Journal:  Sci Rep       Date:  2020-10-27       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.