Literature DB >> 8908856

Effects of fabrication, finishing, and polishing procedures on preload in prostheses using conventional "gold' and plastic cylinders.

A B Carr1, J B Brunski, E Hurley.   

Abstract

This study reviews fundamental concepts related to the use of screws and presents data describing the effect of fabrication, finishing, and polishing procedures on as-received preload for implant cylinders. Specifically, this study measured and compared preload produced when using as-received gold cylinders (the reference or gold standard), and cast cylinders produced from premade gold and plastic cylinders in the as-cast condition and following postcast finishing and polishing manipulations. The results reveal that preload in the gold screw-gold cylinder-abutment joint can be affected by the casting process, and that the choice of cylinder type, casting alloy, investment, and finishing/polishing technique may affect the resultant preload as compared to as-received joint conditions. The data from this study indicate that when plastic patterns are used as part of the framework, finishing and polishing of implant cylinder components should provide an increased preload compared to no such manipulations. Also, if maximum preload is desired, the use of premade metal cylinders offers an advantage over plastic patterns in both preload magnitude and precision.

Mesh:

Substances:

Year:  1996        PMID: 8908856

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

Review 1.  Nanoparticles used in dentistry: A review.

Authors:  Subhashree Priyadarsini; Sumit Mukherjee; Monalisa Mishra
Journal:  J Oral Biol Craniofac Res       Date:  2017-12-07

2.  Preload and torque removal evaluation of three different abutment screws for single standing implant restorations.

Authors:  Rafael Augusto Stüker; Eduardo Rolim Teixeira; João Carlos Pinheiro Beck; Nilza Pereira da Costa
Journal:  J Appl Oral Sci       Date:  2008 Jan-Feb       Impact factor: 2.698

3.  Effect of axial loads on implant-supported partial fixed prostheses by strain gauge analysis.

Authors:  Luis Gustavo Oliveira de Vasconcellos; Renato Sussumu Nishioka; Luana Marotta Reis de Vasconcellos; Lea Nogueira Braulino de Melo Nishioka
Journal:  J Appl Oral Sci       Date:  2011 Nov-Dec       Impact factor: 2.698

4.  Leakage evaluation of original and compatible implant-abutment connections: In vitro study using Rhodamine B.

Authors:  Antoine Berberi; George Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Bassam Badran; Haidar Akl
Journal:  J Dent Biomech       Date:  2014-08-11

5.  Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment.

Authors:  Rudys Rodolfo de Jesus Tavarez; Wellington Cardoso Bonachela; Anuar Antônio Xible
Journal:  J Appl Oral Sci       Date:  2011 Jan-Feb       Impact factor: 2.698

6.  Machined and plastic copings in three-element prostheses with different types of implant-abutment joints: a strain gauge comparative analysis.

Authors:  Renato Sussumu Nishioka; Lea Nogueira Braulino de Melo Nishioka; Celina Wanderley Abreu; Luis Gustavo Oliveira de Vasconcellos; Ivan Balducci
Journal:  J Appl Oral Sci       Date:  2010 May-Jun       Impact factor: 2.698

7.  Effect of dynamic cyclic loading on screw loosening of retightened versus new abutment screw in both narrow and standard implants (in-vitro study).

Authors:  Eman Mohammed Nasr Attiah; Attiah Ali AlGendy; Tamer Mohamed Nasr Mostafa
Journal:  Int J Implant Dent       Date:  2020-07-28

8.  In vitro evaluation of leakage at implant-abutment connection of three implant systems having the same prosthetic interface using rhodamine B.

Authors:  Antoine Berberi; Georges Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Nabil El Zein; Bassam Badran; Haidar Akl
Journal:  Int J Dent       Date:  2014-05-11
  8 in total