Comparison of fit accuracy between Procera custom abutments and three implant systems.

Tiago de Morais Alves da Cunha1, Roberto Paulo Correia de Araújo, Paulo Vicente Barbosa da Rocha, Rosa Maria Pazos Amoedo.   

Abstract

BACKGROUND: Although increase of misfit has been reported when associating implant and abutment from different manufacturers, Procera® (Nobel Biocare™, Göteborg, Sweden) custom abutment has been universally used in clinical practice.
PURPOSE: The purpose of this investigation was to compare the vertical gap of zirconia Procera abutment associated with implants from the same manufacturer (Nobel Biocare) and two other implant systems.
MATERIALS AND METHODS: Twenty-four zirconia Procera abutments were produced using computer-assisted design and manufacture (CAD/CAM) and paired with (1) eight MK Iii RP 4.1 × 10 mm implants (Nobel Biocare) - GNB group; (2) eight Try on, 4.1 × 10 mm implants (Sistema de Implantes, São Paulo, Brazil) - ES group; and (3) eight Master screw, 4.1 × 10 mm implants (Conexão Sistema de Prótese, São Paulo, Brazil) - EC group. A comparison of the vertical misfit at the implant-abutment interface was taken at six measuring sites on each sample using scanning electron microscopy with a magnification of 408×. One-way analysis of variance was used to test for differences, and Tukey's test was used for pairwise comparison of groups (α = 0.05).
RESULTS: Significant differences relative to average misfit were found when Procera abutments were associated with other implant manufacturers. The ES group and EC group did not differ significantly, but both demonstrated significantly larger average misfit than the GNB group (p = .001). The average misfit was 5.7 µm ± 0.39, 9.53 µm ± 0.52 and 10.62 µm ± 2.16, respectively, for groups GNB, ES, and EC.
CONCLUSION: The association of Procera zirconia abutment with other implant systems different from its manufacturer demonstrated significant alteration of vertical misfit at implant-abutment interface.
© 2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20977612     DOI: 10.1111/j.1708-8208.2010.00311.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  4 in total

1.  Leakage evaluation of original and compatible implant-abutment connections: In vitro study using Rhodamine B.

Authors:  Antoine Berberi; George Tehini; Khaldoun Rifai; Farah Bou Nasser Eddine; Bassam Badran; Haidar Akl
Journal:  J Dent Biomech       Date:  2014-08-11

Review 2.  Abutment misfit in implant-supported prostheses manufactured by casting technique: An integrative review.

Authors:  Lorena M S Pereira; Mariane B Sordi; Ricardo S Magini; Antônio R Calazans Duarte; Júlio C M Souza
Journal:  Eur J Dent       Date:  2017 Oct-Dec

3.  Comparative Evaluation of Retentive Properties of Two Compatible Ball Attachments in Mandibular Implant-Retained Overdentures: An In Vitro Study.

Authors:  Maryam Memarian; Simindokht Zarrati; Sedigheh Karimi; Mehran Bahrami
Journal:  J Dent (Tehran)       Date:  2018-03

4.  Accuracy combining different brands of implants and abutments.

Authors:  María-Fernanda Solá-Ruíz; Eduardo Selva-Otaolaurruchi; Gisela Senent-Vicente; Inés González-de-Cossio; Vicente Amigó-Borrás
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-03-01