Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization.

Üreyen Kaya Bulem, Ayse Diljin Kececi, Hilmi Egemen Guldas.   

Abstract

NiTi instruments have a high risk of separation due to torsional or flexural fatigue (cyclic fatigue). Chemomechanical preparation, cleaning procedures, chemical disinfection and sterilization cause the corrosion of endodontic instruments that may weaken the fracture resistance of the instruments. Objective To assess the effects of NaOCl immersion and autoclave sterilization on the cyclic fatigue resistance of ProFile, FlexMaster, Mtwo and TwistedFiles NiTi instruments (tip size 25, 0.06 taper, n=160). Material and Methods The instruments (n=10 for each subgroup) were dynamically immersed in NaOCl; immersed in NaOCl and sterilized in one autoclave cycle; 5 cycles immersed in NaOCl and sterilized in autoclave and not immersed in NaOCl and not sterilized (control group). Dynamic cyclic fatigue resistance was tested. The number of cycles to failure (NCF) were statistically analyzed (P<0.05). Results Kruskall-Wallis test indicated significant differences among the tested instruments in terms of NCF (P=0.000). The mean NCF of Mtwo (556.75) was higher than that of the Twisted File, Flexmaster and ProFile, 483.1, 376.12, 365.25, respectively. NaOCl immersion and autoclave sterilization have no effect on the NCF values of the tested instruments (P>.05). Conclusions Cyclic fatigue resistance of the tested NiTi instruments cannot be adversely affected by NaOCl immersion and autoclave sterilization. Production process (TwistedFiles) or design (Twisted Files, FlexMaster, Mtwo and ProFile) of the instruments can influence their cyclic fatigue resistance.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24473715      PMCID: PMC3891273          DOI: 10.1590/1679-775720130083

Source DB:  PubMed          Journal:  J Appl Oral Sci        ISSN: 1678-7757            Impact factor:   2.698


INTRODUCTION

Nickel-titanium (NiTi) instruments are very popular and provide many advantages during root canal treatment. However, NiTi instruments have a high risk of separation due to torsional or flexural fatigue (cyclic fatigue)[23]. Fractures due to cyclic fatigue occur because of metal fatigue and are more prevalent in curved root canals. The instrument rotates freely at the curvature without binding, so that the tensile and compressive stresses are concentrated at the point of maximum flexure. Resistance to fracture prior to an actual fracture is determined by the number of cycles in which an instrument operates under a specific loading condition[31]. When the tip or some part of the file binds in the root canal during rotation, torsional failure occurs[23]. An important factor that may weaken the fracture resistance is corrosion. Chemomechanical preparation, cleaning procedures, chemical disinfection, and sterilization cause the corrosion of endodontic instruments[17]. Sodium hypochlorite (NaOCl) is the most commonly used irrigant in endodontic treatment[2], and its active ClO- ions have the potential to increase corrosion[22]. NaOCl selectively removes nickel from the instrument surface and causes micropitting[22]. These are the areas where stress gets concentrated, forming cracks[22] and increasing the probability of fatigue failure[26]. Environmental conditions affect crack initiation and propagation processes[26]. The corrosive effect of NaOCl on endodontic NiTi instruments has been studied previously only in a static mode and for long contact times, which did not simulate actual clinical conditions[10]. The resistance to cyclic fatigue of NiTi rotary instruments can be increased via improvements in the manufacturing process or by the use of new alloys with superior mechanical properties. Twisted Files (SybronEndo, Orange, CA, U.S.A.), which were recently introduced, are reported to have better physical and mechanical properties than traditional NiTi instruments. They are produced with a special form of thermomechanical processing, which yields a new, heat-treated NiTi instrument that contains a martensitic phase stable superelastic alloy under clinical conditions[7,13]. It has been reported that the additional heating of NiTi instruments might increase their flexibility[30 ]and microhardness[5]. There have been many studies on the cyclic fatigue resistance of different NiTi rotary files having different designs or compositions[7,8,13]. There are also several reports related to the effect of autoclave sterilization, for either improving or degrading the surface, mechanical, and cutting properties of various NiTi rotary systems[11,24,28,29]. However, there is no information on the influence of both NaOCl and autoclave sterilization on the cyclic fatigue resistance of NiTi rotary instruments. Therefore, the aim of this study was to assess the effect NaOCl solution and autoclave sterilization has on the cyclic fatigue resistance of three traditional (ProFile, FlexMaster, and Mtwo) NiTi instruments and one heat-treated (Twisted Files) NiTi instrument in conditions similar to those used in clinical practice.

MATERIAL AND METHODS

Four NiTi rotary instruments - ProFile (Dentsply Maillefer, Ballaigues, Switzerland), FlexMaster (VDW, GmbH, Munich, Germany), Twisted Files (SybronEndo, Orange, CA, U.S.A.), and Mtwo (VDW, GmbH, Munich, Germany) - were selected for the fatigue resistance test; all the instruments were of the same size (#25, 0.06 taper). Each group included 40 unused instruments that were divided into four subgroups (n=10) to evaluate different protocols. In group 1, the instruments were dynamically immersed in 2.5% NaOCl for 5 min at room temperature. In group 2, the instruments were immersed in NaOCl, under the same conditions as in group 1, and then subjected to one cycle of autoclave sterilization (Statim 5000S, SciCan, Leutkirch im Allgäu, Germany) at a temperature of 135ºC for 32.5 min (17.5 min of sterilization and 15 min of drying). In group 3, the instruments were subjected to five cycles of the same test protocols as in group 2. In the control group, the instruments were not immersed in NaOCl and were not sterilized. For the dynamic immersion, the endodontic instruments were attached to an endodontic motor and rotated freely at a constant speed (300 rpm for ProFile, FlexMaster, and Mtwo; 500 rpm for Twisted Files) in a small glass container with an amount of the NaOCl solution that was sufficient to be in contact with the instruments but not the shaft[18]. The files were rinsed with distilled water to neutralize the effect of the NaOCl and then dried. A total of 160 instruments, 40 from each subgroup, were tested for dynamic cyclic fatigue resistance. Artificial canals were constructed initially by providing a copper duplicate of each instrument to obtain accurately-sized canals. The copper duplicate was milled by increasing the original size of the instrument by 0.1 mm, using a computer numeric control machining bench (Ajan CNC, İzmir, Turkey), according to the curvature properties (60º angle, 2.5 mm radius, curvature beginning 2 mm from the tip of the file) that were chosen for this study. It was then constructed in a stainless steel block using the negative mold, using a die-sinking electrical-discharge machining process (Hartford M 65 series, She Hong Industrial Co. Ltd., Taichung, Taiwan). The depth of the artificial canal was machined to the maximum diameter of the instrument plus 0.2 mm, thus allowing the instrument to rotate freely. The artificial canal was also covered with tempered glass to prevent the instruments from slipping out and to allow observation of the instruments. The dental hand-piece was fixed to permit precise placement of each instrument inside the artificial canal and to guarantee the same three-dimensional positions of the instruments (Figures 1A, B). All the instruments except the Twisted Files were rotated at a constant speed of 300 rpm using a 16:1 contra angle and powered by a torque-controlled electric motor (X-smart, Dentsply Maillefer, Ballaigues, Switzerland) with back-and-forth axial movements until fracture occurred. The Twisted Files were rotated at a constant speed of 500 rpm, as per the manufacturer's recommendation. The amplitude of the axial movements was 3 mm, with about 2 s for every displacement. High-flow synthetic oil (DME, Gloucester, UK) was poured into the simulated canal after each file change to reduce the friction between the instrument and the metal canal walls.
Figure 1

A) Cyclic fatigue testing device and the used canal type B) Trajectory of the size 25/.06 instrument and curvature properties of artificial root canal

A) Cyclic fatigue testing device and the used canal type B) Trajectory of the size 25/.06 instrument and curvature properties of artificial root canal The time-to-fracture was recorded visually with a 1/100-s chronometer. The number of cycles to failure (NCF) was determined by converting the time required to fracture the instrument into decimal units and then multiplying the time by the number of rotations per minute. The NCF values (n=40) obtained from four subgroups with different protocols were averaged to calculate the mean NCF value for each instrument (ProFile, FlexMaster, Twisted Files and Mtwo). The length of each fractured tip was also recorded.

Statistical analysis

Bartlett's test showed that the data did not provide preconditions of homogeneity of variance (P<0.05) and were not normally distributed according to Anderson-Darling test (P<0.05). Thus, the significances among the tested instruments were statistically analyzed with Kruskall-Wallis test (P=0.05).

RESULTS

Kruskall-Wallis test indicated significant differences among the tested instruments in terms of NCF, P=0.000. The mean NCF of Mtwo (556.75) was higher than that of the Twisted File, Flexmaster and ProFile, 483.1, 376.12, 365.25, respectively (Table 1).
Table 1

The mean (SD) number of cycles to failure (NCF)* and fragment lengths in mm (SD) of tested NiTi instruments

Tested NiTi instruments NCF (SD) Fragment length (mm) (SD)
ProFile (n=40)365.25 (75.64)c 2.5 (0.4)
FlexMaster (n=40)376.12 (63.38)c 2.6 (0.4)
Twisted File (n=40)483.1 (133.85)b 2.5 (0.5)
Mtwo (n=40)556.75 (78.58)a 2.7 (0.3)

SD=Standard deviation;

different letters in the same column indicate statistically significant difference

The mean (SD) number of cycles to failure (NCF)* and fragment lengths in mm (SD) of tested NiTi instruments SD=Standard deviation; different letters in the same column indicate statistically significant difference Kruskal-Wallis test showed that NaOCl immersion and autoclave sterilization have no effect on the NCF values of the tested instruments (P>.05), when the effects of NaOCl immersion and autoclave sterilization on each instrument were searched (Table 2).
Table 2

Mean (SD) of the number of cycles to failure (NCF) and fragment lengths in mm (SD) of tested NiTi instruments according to the groups

NiTi instruments Groups NCF (SD) Fragment length mm (SD)
ProFile (n=40)Control396.5 (77.46)2.6 (0.4)
5 min %2.5 NaOCl385.5 (76.03)2.4 (0.5)
5 min %2.5 NaOCl + sterilization359 (73.25)2.5 (0.6)
5X (5 min %2.5 NaOCl + sterilization)320 (61.55)2.6 (0.2)
FlexMaster (n=40)Control379 (70.54)3 (0.4)
5 min %2.5 NaOCl378 (53.39)2.5 (0.4)
5 min %2.5 NaOCl + sterilization374 (70.82)2.7 (0.3)
5X (5 min %2.5 NaOCl + sterilization)373.5 (67.37)2.5 (0.4)
Twisted File (n=40)Control551.2 (118.78)2.4 (0.4)
5 min %2.5 NaOCl526.5 (105.8)2.4 (0.2)
5 min %2.5 NaOCl + sterilization439 (148.16)2.6 (0.4)
5X (5 min %2.5 NaOCl + sterilization)415.7 (126.21)2.5 (0.2)
Mtwo (n=40)Control599.5 (79.46)2.8 (0.2)
5 min %2.5 NaOCl560 (56.81)2.7 (0.2)
5 min %2.5 NaOCl + sterilization553 (87.94)2.6 (0.2)
5X (5 min %2.5 NaOCl + sterilization)514.5 (73.76)2.6 (0.2)

SD=Standard deviation

Mean (SD) of the number of cycles to failure (NCF) and fragment lengths in mm (SD) of tested NiTi instruments according to the groups SD=Standard deviation There were no statistically significant differences among the tested instruments in terms of the length of fractured segments, (P>.05) (Table 1). The mean lengths of the separated segments as measured from the instruments' tips were 2.7, 2.6, 2.5 and, 2.5 mm for Mtwo, FlexMaster, Twisted File and Profile, respectively (Table 1).

DISCUSSION

In the literature, there is still a debate regarding the impact of torsional stress and metal fatigue on the fracturing of NiTi rotary instruments. A number of studies have stated that fatigue is the predominant mechanism in material failure[6,25]. On the other hand, some authors have hypothesized that torsional failure occurs more frequently than flexural fatigue[1,23]. The fatigue life of an instrument is determined by the radius and angle of curvature, which are associated with the diameter of the instrument in the region of maximum bending[3]. Thus, as the diameter of the instrument increases and the curvature radius of the canal decreases, the fatigue resistance of the rotary instruments decreases[29]. Other factors related to cyclic fatigue are the design and cross section of the instrument[27,31], the working speed, torque and sterilization procedures[4,5] and clinical usage[3,8]. To simulate clinical conditions, the test protocol chosen for this study included dynamic immersion in 2.5% NaOCl, a maximum of five cycles of autoclave sterilization, and dynamic cyclic fatigue resistance testing. To remain within a realistic time-frame of clinical practice and considering the mean life of the control instruments (between 1 min, 51.36 s for Mtwo; and 1 min, 13.06 s for ProFile), 5 min was selected as the contact time of the solution with the instrument. The shaft of the instrument was not immersed in the irrigation solution, as is the case in root canal instrumentation procedures. This also served the purpose of avoiding galvanic corrosion phenomena[4]. A dynamic model that incorporates cyclic axial movement not only provides better simulation of the clinical environment but also lengthens the lifespan of the rotary files, because the compressive and tensile stresses are distributed along the instruments[14,16]. However, the NCF values obtained in the present study were lower than those obtained in other studies that used a dynamic model[16,31]. Those studies used an artificial canal that was not constructed based on the specific dimensions of each file group to ensure a suitable file trajectory. Therefore, the canal design they used may have affected the fatigue life and resulted in a wide variation in the fatigue life of the instruments tested, as was claimed by Plotino, et al.[21] (2009). The number of times that rotary instruments can be reused is suggested to vary between 1 and 10[9,12], depending on the canal conditions and the type of instrument usage. Clinicians often recycle NiTi files owing to economic considerations[11]. In the present study, reuse for five times was simulated by five cycles of NaOCl immersion and autoclave sterilization. In a recent study, Pedullà, et al.[18] (2011) reported that static or dynamic immersion in 5% NaOCl did not significantly reduce the cyclic fatigue resistance of Twisted Files, Mtwo, or Revo S SU. The authors attributed this to the location of the corrosion. If the corrosion is not in the maximum stress area, then, the instrument's resistance to cyclic fatigue probably cannot be reduced[18]. In another study, ProFile and RaCe instruments showed a reduction in fatigue resistance after immersion in a heated NaOCl solution for 1 h[19]. In the present study, the results showed no significant difference for the instruments made of traditional NiTi alloy (ProFile, FlexMaster and Mtwo) before and after NaOCl immersion and autoclave sterilization. The findings of other studies that investigated autoclave sterilization effects on NiTi instruments were in agreement with the present results[11,20]. On the other hand, Viana, et al.[29] (2006) reported that the NCF of ProFile rotary instruments (25, 0.06) increased by 16% after five autoclave cycles in a temperature range 122-128ºC. Furthermore, it has been demonstrated that the five cycles of autoclave sterilization increased the microhardness of the NiTi endodontic instruments[5]. NaOCl immersion and autoclave sterilization also had no significant influence on the cyclic fatigue life of the heat-treated instruments (Twisted Files). In contrary, Hilfer, et al.[11] (2011) found that repeated autoclaving significantly reduced the mean NCF of 25/.06 Twisted Files. As they explained it, 135ºC is too low a temperature to produce a heat treatment effect, because it is known that such changes require a temperature of at least 170ºC, and that the maximum fatigue resistance can be obtained at 430-440ºC[15]. Furthermore, multiple autoclave cycles have been reported to increase the depth of NiTi file surface irregularities, causing fatigue propagation[1,28]. On the other hand, Plotino, et al.[20] (2012) suggested that autoclave sterilization might improve the mechanical properties of instruments (K3 XF protypes, SybronEndo) that are manufactured using a type of heat treatment. The quality of manufacturing processes and different thermal treatments influence the performance of NiTi rotary instruments because NiTi is an alloy that is very sensitive to both the thermal and mechanical (machining) stresses that can occur during the manufacturing process[20]. A better arrangement of the crystal structure and changes in the percentage of phases of the alloy obtained via thermal treatment provide improvements in superelastic and plastic behaviors[20]. In the production process of Twisted Files, R-phase heat treatment technology modifies the crystalline structure, making it finer than that of traditionally processed materials, and it maximizes the flexibility and resistance to breakage without affecting the superelasticity[16]. Although Plotino, et al.[20] (2012) reported that this new manufacturing technology enhanced the mechanical properties of instruments, the results of the present study did not confirm this finding. The traditional NiTi instrument, Mtwo, showed better cyclic fatigue resistance than the other traditional instruments, ProFile and FlexMaster, and the heat-treated instrument, Twisted Files. Twisted Files showed better cyclic fatigue resistance than ProFile and FlexMaster, as reported previously[16]. Oh, et al.[16] (2010) demonstrated that Twisted Files have fewer surface defects (microfractures) because of twisting, and a smaller cross-sectional area than ProFile instruments, characteristics that might result in their better fatigue resistance. However, scanning electron micrography (SEM) micrographs revealed wrinkle-like grooves because of twisting on the flute faces of Twisted Files, which may increase their level of stress during instrumentation[31]. All tested instruments fractured at the point of the maximum flexure within the curved part of the artificial canal where the stress concentrates. As expected, the lengths of the fractured segments of the instruments were not influenced by the NaOCl immersion and autoclave sterilization.

CONCLUSION

The comparison within the conditions of this study showed that cyclic fatigue resistance of the tested NiTi instruments cannot be adversely affected by NaOCl immersion and autoclave sterilization. Production process (TwistedFiles) or design (Twisted Files, FlexMaster, Mtwo and ProFile) of the instruments can influence their cyclic fatigue resistance. However, further studies investigating the different factors which can affect the instruments cyclic fatigue resistance, fracture modes and different apparatus design with more similar features to root dentine are necessary.
  31 in total

1.  Defects in nickel-titanium instruments after clinical use. Part 2: Fractographic analysis of fractured surface in a cohort study.

Authors:  Ya Shen; Gary Shun-Pan Cheung; Bin Peng; Markus Haapasalo
Journal:  J Endod       Date:  2009-01       Impact factor: 4.171

2.  Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods.

Authors:  Gianluca Gambarini; Nicola Maria Grande; Gianluca Plotino; Francesco Somma; Manish Garala; Massimo De Luca; Luca Testarelli
Journal:  J Endod       Date:  2008-06-25       Impact factor: 4.171

3.  Multiple autoclave cycles affect the surface of rotary nickel-titanium files: an atomic force microscopy study.

Authors:  Caroline R A Valois; Luciano P Silva; Ricardo B Azevedo
Journal:  J Endod       Date:  2008-07       Impact factor: 4.171

4.  Defects in rotary nickel-titanium files after clinical use.

Authors:  B Sattapan; G J Nervo; J E Palamara; H H Messer
Journal:  J Endod       Date:  2000-03       Impact factor: 4.171

5.  Effect of immersion in sodium hypochlorite on torque and fatigue resistance of nickel-titanium instruments.

Authors:  Ove A Peters; Jan O Roehlike; Michael A Baumann
Journal:  J Endod       Date:  2007-03-12       Impact factor: 4.171

Review 6.  A review of cyclic fatigue testing of nickel-titanium rotary instruments.

Authors:  Gianluca Plotino; Nicola M Grande; Massimo Cordaro; Luca Testarelli; Gianluca Gambarini
Journal:  J Endod       Date:  2009-09-01       Impact factor: 4.171

7.  Cleaning of flattened root canals with different irrigating solutions and nickel-titanium rotary instrumentation.

Authors:  Marcos Pôrto de Arruda; Jacy Ribeiro de Carvalho Junior; Carlos Eduardo Saraiva Miranda; Cristina Paschoalato; Silvio Rocha C Silva
Journal:  Braz Dent J       Date:  2009

8.  Deformation and fracture of Mtwo rotary nickel-titanium instruments after clinical use.

Authors:  Ugur Inan; Nihan Gonulol
Journal:  J Endod       Date:  2009-08-15       Impact factor: 4.171

9.  Cyclic fatigue analysis of a new generation of nickel titanium rotary instruments.

Authors:  C Michael Larsen; Ikuya Watanabe; Gerald N Glickman; Jianing He
Journal:  J Endod       Date:  2009-03       Impact factor: 4.171

10.  Effect of heat treatment on transformation temperatures and bending properties of nickel-titanium endodontic instruments.

Authors:  Y Yahata; T Yoneyama; Y Hayashi; A Ebihara; H Doi; T Hanawa; H Suda
Journal:  Int Endod J       Date:  2009-05-08       Impact factor: 5.264

View more
  8 in total

1.  Effect of autoclave sterilization on cyclic fatigue and torsional fracture resistance of NiTi rotary instruments.

Authors:  Wooyoung Kim; Soram Oh; Gil-Joo Ryu; Tae-Hwan Kim; Sung-Jae Kim; Dong-Hyung Kim; Bin-Na Lee; Kee-Yeon Kum; Seok Woo Chang; Ji-Hyun Jang
Journal:  Odontology       Date:  2019-09-11       Impact factor: 2.634

2.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

3.  Cyclic fatigue resistance of D-RaCe, ProTaper, and Mtwo nickel-titanium retreatment instruments after immersion in sodium hypochlorite.

Authors:  Hüseyin Sinan Topçuoğlu; Kanşad Pala; Ahmet Aktı; Salih Düzgün; Gamze Topçuoğlu
Journal:  Clin Oral Investig       Date:  2015-09-30       Impact factor: 3.573

Review 4.  Nickel-Titanium Rotary File Systems: What's New?

Authors:  Sadia Tabassum; Kamil Zafar; Fahad Umer
Journal:  Eur Endod J       Date:  2019-10-18

5.  The Effect of Multiple Autoclave Sterilization on the Cyclic Fatigue of Three Heat-Treated Nickel-Titanium Rotary Files: EdgeFile X7, Vortex Blue, and TRUShape.

Authors:  Reham Sharroufna; Mohammed Mashyakhy
Journal:  Biomed Res Int       Date:  2020-12-14       Impact factor: 3.411

6.  Cyclic Fatigue Resistance of Heat-Treated Nickel-Titanium Instruments.

Authors:  Mário Tanomaru-Filho; Camila Galletti Espir; Ana Carolina Venção; Nathaly Macedo-Serrano; Jader Camilo-Pinto; JulianeMaria Guerreiro-Tanomaru
Journal:  Iran Endod J       Date:  2018

Review 7.  Management of Instrument Sterilization Workflow in Endodontics: A Systematic Review and Meta-Analysis.

Authors:  Mario Dioguardi; Diego Sovereto; Gaetano Illuzzi; Enrica Laneve; Bruna Raddato; Claudia Arena; Vito Carlo Alberto Caponio; Giorgia Apollonia Caloro; Khrystyna Zhurakivska; Giuseppe Troiano; Lorenzo Lo Muzio
Journal:  Int J Dent       Date:  2020-02-08

8.  Influence of autoclave sterilization procedures on the cyclic fatigue resistance of heat-treated nickel-titanium instruments: a systematic review.

Authors:  Emmanuel João Nogueira Leal Silva; Mayara Zanon; Fernanda Hecksher; Felipe Gonçalves Belladonna; Rafaela Andrade de Vasconcelos; Tatiana Kelly da Silva Fidalgo
Journal:  Restor Dent Endod       Date:  2020-03-31
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.