Cyclic fatigue of nickel-titanium rotary instruments after clinical use with low- and high-torque endodontic motors.

G Gambarini1.   

Abstract

This study investigated the cyclic fatigue resistance of used nickel-titanium rotary instruments, clinically operated by a traditional high-torque motor and a new low-torque electric motor. Fifty instruments (ProFile, Maillefer, Baillagues, Switzerland) for each of the following sizes were used: .06-25, .06-20, and .04-25. They were divided into three groups: A = 10 new instruments (control), B = 20 used instruments operated by a high-torque motor, and C = 20 used instruments operated by a low-torque motor. Each instrument was used in 10 clinical cases (at least six molars). Cyclic fatigue testing of new and used instruments was performed with a device that allowed the instruments to rotate freely inside a stainless-steel artificial canal at 350 rpm. Data were recorded and statistically analyzed. Mean (and SD) numbers of rotation-to-breakage cycles for .06-25, .06-20, and .04-25 instruments were, respectively: group A 463 (85), 535 (101), and 613 (105); group B 229 (64), 343 (92), and 233 (84); and group C 352 (77), 454 (89), and 466 (86). Results showed that group A instruments exhibited significantly higher values of rotation-to-breakage cycles (two-sample t test, p < 0.01), when compared with groups B and C, respectively. Statistically higher resistance to cyclic fatigue was noted for group C in all sizes compared with group B (two-sample t test, p < 0.01), showing that the use of endodontic motor with lower torque values reduced cyclic fatigue of nickel-titanium rotary instruments.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11771588     DOI: 10.1097/00004770-200112000-00015

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  20 in total

1.  Influence of file motion on cyclic fatigue of new nickel titanium instruments.

Authors:  Alessio Giansiracusa Rubini; Gianpaolo Sannino; Giancarlo Pongione; Luca Testarelli; Dina Al Sudani; Jeeraphat Jantarat; Massimo De Luca; Gianluca Gambarini
Journal:  Ann Stomatol (Roma)       Date:  2013-03-20

2.  Flexibility and resistance to cyclic fatigue of endodontic instruments made with different nickel-titanium alloys: a comparative test.

Authors:  Giancarlo Pongione; Giorgio Pompa; Valerio Milana; Stefano Di Carlo; Alessio Giansiracusa; Emanuele Nicolini; Francesca De Angelis
Journal:  Ann Stomatol (Roma)       Date:  2012-01-14

3.  Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals.

Authors:  Jung-Hong Ha; Sang-Shin Park
Journal:  Restor Dent Endod       Date:  2012-11-21

Review 4.  Endodontic Rotary Files, What Should an Endodontist Know?

Authors:  Ana-Belén Dablanca-Blanco; Pablo Castelo-Baz; Ramón Miguéns-Vila; Pablo Álvarez-Novoa; Benjamín Martín-Biedma
Journal:  Medicina (Kaunas)       Date:  2022-05-27       Impact factor: 2.948

5.  Cyclic fatigue resistance of four nickel-titanium rotary instruments: a comparative study.

Authors:  Eugenio Pedullà; Gianluca Plotino; Nicola Maria Grande; Alfio Pappalardo; Ernesto Rapisarda
Journal:  Ann Stomatol (Roma)       Date:  2012-08-09

6.  In Vivo Evaluation of Operative Torque Generated by Two Nickel-Titanium Rotary Instruments during Root Canal Preparation.

Authors:  Gianluca Gambarini; Massimo Galli; Marco Seracchiani; Dario Di Nardo; Marco A Versiani; Lucila Piasecki; Luca Testarelli
Journal:  Eur J Dent       Date:  2019-12-31

7.  An in vitro evaluation of the effect of deep dry cryotreatment on the cutting efficiency of three rotary nickel titanium instruments.

Authors:  Gingu Koshy George; Kavitha Sanjeev; Mahalaxmi Sekar
Journal:  J Conserv Dent       Date:  2011-04

8.  Cyclic Fatigue Resistance and Fractographic Analysis of Race and Protaper Rotary NiTi Instruments.

Authors:  Shahram Azimi; Parisa Delvari; Hamid Cyrus Hajarian; Mohammad Ali Saghiri; Kasra Karamifar; Mehrdad Lotfi
Journal:  Iran Endod J       Date:  2011-05-15

9.  Cyclic fatigue resistance tests of Nickel-Titanium rotary files using simulated canal and weight loading conditions.

Authors:  Ok-In Cho; Antheunis Versluis; Gary Sp Cheung; Jung-Hong Ha; Bock Hur; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2013-02-26

10.  [Not Available].

Authors:  L Testarelli; N M Grande; G Plotino; M Lendini; G Pongione; G De Paolis; F Rizzo; V Milana; G Gambarini
Journal:  Open Dent J       Date:  2009-04-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.