The effect of cleaning procedures on fracture properties and corrosion of NiTi files.

P Y Z O'Hoy1, H H Messer, J E A Palamara.   

Abstract

AIM: To evaluate the effect of repeated cleaning procedures on fracture properties and corrosion of nickel-titanium (NiTi) files.
METHODOLOGY: New NiTi instruments were subjected to 2, 5 and 10 cleaning cycles with the use of either diluted bleach (1% NaOCl) or Milton's solution (1% NaOCl plus 19% NaCl) as disinfectant. Each cleaning cycle consisted of scrubbing, rinsing and immersing in NaOCl for 10 min followed by 5 min of ultrasonication. Files were then tested for torsional failure and flexural fatigue, and observed for evidence of corrosion using scanning electron microscope (SEM). Four brands of NiTi files were immersed in either Milton's solution or diluted bleach overnight and evaluated for corrosion.
RESULTS: Up to 10 cleaning cycles did not significantly reduce the torque at fracture or number of revolutions to flexural fatigue (P > 0.05, two-way anova), although decreasing values were noted with increasing number of cleaning cycles using Milton's solution. No corrosion was detected on the surface of these files. Files immersed in 1% NaOCl overnight displayed a variety of corrosion patterns. The extent of corrosion was variable amongst different brands of files and amongst files in each brand. Overall, Milton's solution was much more corrosive than diluted bleach. Corrosion of file handles was often extreme.
CONCLUSIONS: Files can be cleaned up to 10 times without affecting fracture susceptibility or corrosion, but should not be immersed in NaOCl overnight. Milton's solution is much more corrosive than bleach with the same NaOCl concentration.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14641435     DOI: 10.1046/j.1365-2591.2003.00709.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  9 in total

1.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

2.  Electrochemical corrosion assessment of RaCe and Mtwo rotary nickle-titanium instruments after clinical use and sterilization.

Authors:  S Shahi; H Mokhtari; S Rahimi; V Shiezadeh; H Ashasi; M Abdolrahimi; M Foroughreyhani
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2012-03-01

3.  The effect of multiple autoclave cycles on the surface of rotary nickel-titanium endodontic files: An in vitro atomic force microscopy investigation.

Authors:  Ashish Shashikant Nair; Mahima Tilakchand; Balaram Damodar Naik
Journal:  J Conserv Dent       Date:  2015 May-Jun

4.  Effect of sodium hypochlorite on conventional and heat-treated nickel-titanium endodontic rotary instruments - An in vitro study.

Authors:  Jack Han-Hsing Lin; Bekir Karabucak; Su-Min Lee
Journal:  J Dent Sci       Date:  2020-09-14       Impact factor: 2.080

5.  Qualitative Assessment of the Surface Topographic Changes of XP-endo Shaper and TruNatomy files after exposure to Sodium Hypochlorite and Ethylenediaminetetraacetic Acid.

Authors:  Jerry Jose; Akshay Khandelwal; Riluwan Siddique
Journal:  Eur Endod J       Date:  2021-08

6.  A novel approach to determine the effect of irrigation on temperature and failure of Ni-Ti endodontic rotary files.

Authors:  Sayed Ali Mousavi; Vida Kargar-Dehnavi; Sayed Amir Mousavi
Journal:  Dent Res J (Isfahan)       Date:  2012-05

7.  Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization.

Authors:  Üreyen Kaya Bulem; Ayse Diljin Kececi; Hilmi Egemen Guldas
Journal:  J Appl Oral Sci       Date:  2013 Nov-Dec       Impact factor: 2.698

8.  Antibacterial and Toxic Effect of Hydrogen Peroxide Combined with Different Concentrations of Chlorhexidine in Comparison with Sodium Hypochlorite.

Authors:  Hosein Mirhadi; Abbas Abbaszadegan; Mohammad Ali Ranjbar; Mohammad Reza Azar; Bita Geramizadeh; Shima Torabi; Zeinab Sadat Aleyasin; Ahmad Gholami
Journal:  J Dent (Shiraz)       Date:  2015-12

9.  Comparison of Cleaning Efficacy and Instrumentation Time in Primary Molars: Mtwo Rotary Instruments vs. Hand K-Files.

Authors:  Fatemeh Ramezanali; Farzaneh Afkhami; Ali Soleimani; Mohammad Javad Kharrazifard; Farshid Rafiee
Journal:  Iran Endod J       Date:  2015
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.