| Literature DB >> 18634935 |
Gianluca Gambarini1, Nicola Maria Grande, Gianluca Plotino, Francesco Somma, Manish Garala, Massimo De Luca, Luca Testarelli.
Abstract
The aim of the present study was to investigate whether cyclic fatigue resistance is increased for nickel-titanium instruments manufactured by using new processes. This was evaluated by comparing instruments produced by using the twisted method (TF; SybronEndo, Orange, CA) and those using the M-wire alloy (GTX; Dentsply Tulsa-Dental Specialties, Tulsa, OK) with instruments produced by a traditional NiTi grinding process (K3, SybronEndo). Tests were performed with a specific cyclic fatigue device that evaluated cycles to failure of rotary instruments inside curved artificial canals. Results indicated that size 06-25 TF instruments showed a significant increase (p < 0.05) in the mean number of cycles to failure when compared with size 06-25 K3 files. Size 06-20 K3 instruments showed no significant increase (p > 0.05) in the mean number of cycles to failure when compared with size 06-20 GT series X instruments. The new manufacturing process produced nickel-titanium rotary files (TF) significantly more resistant to fatigue than instruments produced with the traditional NiTi grinding process. Instruments produced with M-wire (GTX) were not found to be more resistant to fatigue than instruments produced with the traditional NiTi grinding process.Entities:
Mesh:
Substances:
Year: 2008 PMID: 18634935 DOI: 10.1016/j.joen.2008.05.007
Source DB: PubMed Journal: J Endod ISSN: 0099-2399 Impact factor: 4.171