Cyclic fatigue resistance of three different nickel-titanium instruments after immersion in sodium hypochlorite.

Eugenio Pedullà1, Nicola M Grande, Gianluca Plotino, Alfio Pappalardo, Ernesto Rapisarda.   

Abstract

INTRODUCTION: The purpose of this study was to assess the resistance to cyclic fatigue of three nickel-titanium (NiTi) files after the immersion in sodium hypochlorite (NaOCl) solution in conditions similar to those used in clinical practice.
METHODS: A total of 150 new Twisted Files (SybronEndo, Orange, CA), Revo S SU files (Micro Mega, Besancon, France), and Mtwo files (Sweden and Martina, Padova, Italy), size 25.06, were tested. Fifty files of the same brand were randomly assigned to five groups (n = 10) and submitted to the following immersion protocol in 5% NaOCl at 37°C for 16 mm: no immersion (control), 5 minutes statically, 1 minute statically, 5 minutes dynamically (300 rpm/min), and 1 minute dynamically. Resistance to cyclic fatigue was determined by counting the numbers of cycles to failure in a 60° curve with a 5-mm radius. Data were analyzed by two-way analyses of variance.
RESULTS: Resistance to cyclic fatigue of the same NiTi file was not significantly affected by immersion in NaOCl (P > .05). The Twisted File showed a higher resistance in all groups than Revo S SU (P < .001). The comparison between the same groups of Twisted Files and Mtwo files or between Mtwo and Revo S files did not show significant differences (P > .05) except for two cases: group 2 of the Twisted Files and Mtwo files and group 5 of the Mtwo and Revo S SU files (P < .05).
CONCLUSIONS: Static or dynamic immersion in NaOCl for 1 minute or 5 minutes did not reduce the cyclic fatigue resistance of NiTi significantly. However, the type of instrument influences cyclic fatigue resistance. In our study, Twisted Files were more resistant followed by Mtwo and Revo S SU files.
Copyright © 2011 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21763909     DOI: 10.1016/j.joen.2011.04.008

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  8 in total

1.  Resistance to Cyclic Fatigue of Nickel-Titanium Files Immersed in Sodium Hypochlorite at Body Temperature.

Authors:  Hind F Abuhulaibah; Ammar AbuMostafa
Journal:  Int J Dent       Date:  2020-09-19

2.  Effect of NaOCl and EDTA irrigating solutions on the cyclic fatigue resistance of EdgeTaper Platinum instruments.

Authors:  Hussam Alfawaz; Abdullah Alqedairi; Maha Alhamdan; Najla Alkhzim; Shatha Alfarraj; Ahmed Jamleh
Journal:  BMC Oral Health       Date:  2022-05-22       Impact factor: 3.747

3.  Cyclic fatigue resistance of D-RaCe, ProTaper, and Mtwo nickel-titanium retreatment instruments after immersion in sodium hypochlorite.

Authors:  Hüseyin Sinan Topçuoğlu; Kanşad Pala; Ahmet Aktı; Salih Düzgün; Gamze Topçuoğlu
Journal:  Clin Oral Investig       Date:  2015-09-30       Impact factor: 3.573

4.  Cyclic fatigue resistance of Mtwo NiTi rotary instruments used by experienced and novice operators--an in vivo and in vitro study.

Authors:  Gianluca Plotino; Dina Al-Sudani; Stefano Pulino; Nicola M Grande; Piero Alessandro Marcoli; Silvia Pizzi; Luca Testarelli; Gianluca Gambarini
Journal:  Med Sci Monit       Date:  2012-06

5.  Effect of immersion in sodium hypochlorite on cyclic fatigue resistance of RaCe, HyFlex CM and XP-endo shaper files.

Authors:  Esmaeel Dadgar; Masoud Khabiri; Amin Mortaheb; Saeid Hoseini
Journal:  Dent Res J (Isfahan)       Date:  2022-07-18

6.  Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization.

Authors:  Üreyen Kaya Bulem; Ayse Diljin Kececi; Hilmi Egemen Guldas
Journal:  J Appl Oral Sci       Date:  2013 Nov-Dec       Impact factor: 2.698

7.  Comparison of the shaping ability of GT® Series X, Twisted Files and AlphaKite rotary nickel-titanium systems in simulated canals.

Authors:  Raidan Ba-Hattab; Anne-Kathrin Pröhl; Hermann Lang; Dieter Pahncke
Journal:  BMC Oral Health       Date:  2013-12-17       Impact factor: 2.757

8.  An analysis of cyclic fatigue resistance of reciprocating instruments in different canal curvatures after immersion in sodium hypochlorite and autoclaving: An in vitro study.

Authors:  C Champa; V Divya; A Srirekha; Rupali Karale; Ashwija Shetty; Pushpalatha Sadashiva
Journal:  J Conserv Dent       Date:  2017 May-Jun
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.