AIM: To examine the topographic features of the fracture surface of a NiTi instrument after fatigue failure, and to correlate the measurements of some features with the cyclic load.
METHODOLOGY: A total of 212 ProFile rotary instruments were subjected to a rotational-bending test at various curvatures until broken. The fracture surface of all fragments was examined by SEM to identify the crack origins. The crack radius, i.e. extent of the fatigue-crack growth towards the centroid of the cross-section, was also measured, and correlated with the strain amplitude for each instrument.
RESULTS: All fracture surfaces revealed the presence of one or more crack origins, a region occupied by microscopic striations, and an area with microscopic dimples. The number of specimens showing multiple crack origins was significantly greater in the group fatigued under water than in air (P < 0.05). A linear relationship between the reciprocal of the square root of the crack radius and the strain amplitude was discernible (P < 0.001), the slopes of which were not significantly different for instruments fatigued in air and water.
CONCLUSIONS: The fractographic appearance of NiTi engine-files that had failed because of fatigue is typical of that for other metals. The fatigue behaviour of NiTi instruments is adversely affected by water, not only for the low-cycle fatigue life, but also the number of crack origins. There appears to be a critical extent of crack propagation for various strain amplitudes leading to final rupture (akin to the Griffith's criterion for brittle materials).

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17511786     DOI: 10.1111/j.1365-2591.2007.01256.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  6 in total

1.  Fatigue resistance of rotary instruments manufactured using different nickel-titanium alloys: a comparative study.

Authors:  Gianluca Plotino; Luca Testarelli; Dina Al-Sudani; Gianlcarlo Pongione; Nicola M Grande; Gianluca Gambarini
Journal:  Odontology       Date:  2012-10-18       Impact factor: 2.634

2.  Metallurgical analysis and fatigue resistance of WaveOne and ProTaper nickel-titanium instruments.

Authors:  Chiara Pirani; Oddone Ruggeri; Pier Paolo Cirulli; Gian Andrea Pelliccioni; Maria Giovanna Gandolfi; Carlo Prati
Journal:  Odontology       Date:  2013-04-09       Impact factor: 2.634

3.  Incidence of Deformation and Fracture of Twisted File Adaptive Instruments after Repeated Clinical Use.

Authors:  Gianluca Gambarini; Lucila Piasecki; Dario Di Nardo; Gabriele Miccoli; Gianni Di Giorgio; Everdan Carneiro; Dina Al-Sudani; Luca Testarelli
Journal:  J Oral Maxillofac Res       Date:  2016-12-28

4.  Experimental evaluation of cyclic fatigue resistance of four different nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization.

Authors:  Üreyen Kaya Bulem; Ayse Diljin Kececi; Hilmi Egemen Guldas
Journal:  J Appl Oral Sci       Date:  2013 Nov-Dec       Impact factor: 2.698

5.  The Effect of Autoclave Sterilization on Resistance to Cyclic Fatigue of Hero Endodontic File #642 (6%) at Two Artificial Curvature.

Authors:  Masoud Khabiri; Maziar Ebrahimi; Mohammad Reza Saei
Journal:  J Dent (Shiraz)       Date:  2017-12

6.  Cyclic fatigue resistance of M-Pro and RaCe Ni-Ti rotary endodontic instruments in artificial curved canals: a comparative in vitro study.

Authors:  Hadeer Mostafa El Feky; Khalid Mohammed Ezzat; Marwa Mahmoud Ali Bedier
Journal:  Restor Dent Endod       Date:  2019-11-07
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.